Liệu pháp bức xạ định vị cơ thể: Báo cáo của Nhóm Nhiệm vụ 101 AAPM

Medical Physics - Tập 37 Số 8 - Trang 4078-4101 - 2010
Stanley Benedict1,2, Kamil M. Yenice3, D Followill4, James M. Galvin5, W. Hinson6, Brian D. Kavanagh7, Paul Keall8, Michael Lovelock9, Sanford L. Meeks10, Lech Papież11, Thomas G. Purdie12, R Sadagopan4, Michael C. Schell13, Bill J. Salter14, David Schlesinger2, Almon S. Shiu4, Timothy D. Solberg11, Danny Y. Song15, Volker W. Stieber16, Robert Timmerman11, Wolfgang A. Tomé17, Dirk Verellen18, Lu Wang19, F Yin20
1Electronic mail: [email protected]
2University of Virginia Health System, Charlottesville, Virginia 22908
3University of Chicago, Chicago, Illinois 60637
4University of Texas MD Anderson Cancer Center, Houston, Texas 77030
5Thomas Jefferson University Hospital, Philadelphia, Pennsylvania 19107
6Wake Forest University, Winston Salem, North Carolina 27157
7University of Colorado School of Medicine, Aurora, Colorado 80045
8Stanford University Palo Alto, California 94305
9Memorial Sloan–Kettering Cancer Center, New York, New York 10021
10M.D. Anderson Cancer Center Orlando, Orlando, Florida 32806
11University of Texas Southwestern Medical Center, Dallas, Texas 75390
12University of Toronto, Princess Margaret Hospital, Toronto, Ontario M5G 2M9, Canada
13University of Rochester Medical Center, Rochester, New York 14642
14University of Utah, Salt Lake City, Utah 84112
15Johns Hopkins University, Baltimore, Maryland 21231
16Forsyth Regional Cancer Center, Winston Salem, North Carolina 27103
17University of Wisconsin, Madison, Wisconsin 53792
18UV Brussel, Vrije Universiteit Brussel (VUB), Brussels B-1090, Belgium
19Fox Chase Cancer Center, Philadelphia, Pennsylvania 19111
20Duke University Medical Center, Durham, North Carolina, 27710

Tóm tắt

Nhóm Nhiệm vụ 101 của AAPM đã chuẩn bị báo cáo này dành cho các nhà vật lý y tế, bác sĩ lâm sàng và các nhà trị liệu nhằm phác thảo các hướng dẫn thực hành tốt nhất cho kỹ thuật liệu pháp bức xạ bên ngoài được gọi là liệu pháp bức xạ định vị cơ thể (SBRT). Báo cáo của nhóm nhiệm vụ bao gồm một đánh giá tài liệu để xác định các phát hiện lâm sàng đã được báo cáo và các kết quả dự kiến cho phương thức điều trị này. Thông tin được cung cấp để thiết lập một chương trình SBRT, bao gồm các quy trình, thiết bị, tài nguyên và quy trình đảm bảo chất lượng. Ngoài ra, các đề xuất cho việc phát triển tài liệu nhất quán để kê đơn, báo cáo và ghi chép việc thực hiện điều trị SBRT cũng được cung cấp.

Từ khóa

#Liệu pháp bức xạ định vị cơ thể #SBRT #hướng dẫn thực hành tốt #vật lý y tế #đảm bảo chất lượng

Tài liệu tham khảo

10.1016/S0140-6736(04)16250-8

10.3171/jns.2001.94.1.0001

10.1016/S0360-3016(00)80308-9

Izawa M., 2000, Gamma knife radiosurgery for pituitary adenomas, J. Neurosurg., 93, 19, 10.3171/jns.2000.93.supplement_3.0019

10.1097/00006123-200111000-00001

10.3171/jns.2002.97.2.0347

10.3171/jns.1998.89.2.0183

10.1016/j.semradonc.2008.04.001

10.1200/JCO.2009.25.0928

10.3322/caac.20013

10.1200/JCO.1997.15.3.938

10.1056/NEJM199002223220802

10.1080/02841860802662722

10.1200/JCO.1995.13.1.8

10.1038/ncponc0075

10.1002/cncr.23209

10.1016/S0022-5223(97)70397-0

10.1016/j.radonc.2005.03.022

10.1158/1078-0432.CCR-08-0358

10.1016/j.semradonc.2005.12.008

10.1038/ncponc0560

Chang J. E., 2007, Radiotherapy and radiosensitizers in the treatment of glioblastoma multiforme, Clin Adv Hematol & Oncol, 5, 894

10.1016/j.critrevonc.2006.06.007

10.3109/07853890008995907

10.1080/028418699432806

10.3109/02841869509127197

10.1016/S0360-3016(99)00413-7

10.1200/JCO.2001.19.1.164

10.1007/s001170050928

10.1159/000078271

10.1007/PL00002379

10.1378/chest.124.5.1946

10.1016/S0003-4975(02)04681-7

10.1002/cncr.10853

10.1016/S0167-8140(02)00063-4

10.1016/S0169-5002(03)00040-0

10.1016/S0360-3016(01)02731-6

10.1002/cncr.20539

10.1016/S0360-3016(03)01132-5

10.3171/sup.2004.101.supplement3.0413

10.1227/01.NEU.0000114263.01917.1E

10.1016/j.ijrobp.2004.04.025

10.1002/cncr.11296

Ryu S., 2004, Patterns of failure after single‐dose radiosurgery for spinal metastasis, J. Neurosurg., 101, 402, 10.3171/sup.2004.101.supplement3.0402

10.1016/j.ijrobp.2004.06.260

10.1159/000100177

10.1097/01.brs.0000251863.76595.a2

10.1016/S0360-3016(02)02846-8

10.1016/j.ijrobp.2005.03.073

10.1016/j.ijrobp.2005.01.002

10.1016/j.radonc.2009.02.003

10.1016/j.radonc.2004.12.022

10.1080/02841860600904854

10.1016/j.ijrobp.2005.07.002

10.1007/s001170050846

10.1002/(SICI)1098-2388(200003)18:2<152::AID-SSU9>3.0.CO;2-H

10.1097/00130404-200403000-00006

10.1016/j.bpg.2004.10.005

10.1016/S0360-3016(99)00288-6

10.1016/S0360-3016(98)00351-4

10.1088/0031-9155/45/8/305

10.1259/bjr.74.877.740015

10.1016/S0360-3016(00)01472-3

10.1148/radiology.178.3.1847239

10.1016/0360-3016(90)90130-C

10.1055/s-2004-821042

Gambhir S. S., 2001, A tabulated summary of the FDG PET literature, J. Nucl. Med., 42, 1S

10.3816/CLC.2001.n.019

10.1056/NEJMoa022136

10.1053/j.gastro.2004.09.029

10.1097/00006231-200504000-00003

10.1007/978-3-662-03107-0_7

10.1097/00004424-199408000-00002

10.1177/030089169608200512

10.1148/radiol.2291020859

10.1118/1.1398037

10.1016/S0360-3016(02)02803-1

10.1016/S0360-3016(01)01621-2

10.1016/S0360-3016(02)04609-6

10.1053/j.semradonc.2003.10.004

10.1016/S0360-3016(01)01716-3

10.1016/j.ijrobp.2004.12.066

10.1002/(SICI)1097-0142(19980315)82:6<1062::AID-CNCR8>3.0.CO;2-G

10.1053/j.semradonc.2003.10.009

10.1016/S0360-3016(01)01592-9

10.1016/S0360-3016(99)00154-6

10.1118/1.1570372

10.1016/j.ijrobp.2006.05.032

10.1016/j.ijrobp.2003.07.005

10.1016/S0360-3016(03)00010-5

10.1016/S0360-3016(03)00710-7

10.1016/S0360-3016(02)04143-3

10.1016/S0360-3016(97)80617-7

10.1118/1.1494830

10.1080/02841860600902817

10.1016/j.ijrobp.2005.05.045

10.1016/j.ijrobp.2005.10.006

10.1088/0031-9155/54/3/018

Barish M. A., 1999, Motion artifact control in body MR imaging, Magn. Reson Imaging Clin. N. Am., 7, 289, 10.1016/S1064-9689(21)00023-4

10.1016/S0033-8389(02)00070-2

10.1016/S0009-9260(97)80250-4

10.1016/S0009-9260(02)00540-8

10.1002/jmri.20144

10.1118/1.1448824

10.1118/1.1809778

10.1118/1.1739671

Nehmeh S. A., 2003, Reduction of respiratory motion artifacts in PET imaging of lung cancer by respiratory correlated dynamic PET: Methodology and comparison with respiratory gated PET, J. Nucl. Med., 44, 1644

Cardenes H., 2002, Extracranial stereotactic radioablation: Review of biological basis, technique and preliminary clinical experience, Oncologica, 25, 193

10.3109/02841869409121782

10.1016/S0895-7177(03)00081-5

10.1016/S0360-3016(99)00203-5

ICRU “Prescribing recording and reporting photon beam therapy ” ICRU Report No. 50 1993.

ICRU “Prescribing recording and reporting photon beam therapy (supplement to ICRU Report No. 50) ” ICRU Report No. 62 1999.

10.1016/j.ijrobp.2007.03.023

Papiez L., 2004, Stereotactic Body Radiation Therapy, 160

10.1016/S0939-3889(15)70476-9

10.1016/S0360-3016(02)04504-2

10.1016/j.ijrobp.2004.02.060

10.1016/j.ijrobp.2008.08.004

10.1016/j.ijrobp.2004.07.691

10.1016/j.ijrobp.2007.09.050

10.1016/j.ijrobp.2008.03.062

10.1080/02841860310013490

10.1186/1748-717X-4-3

10.1016/j.meddos.2007.04.001

10.1118/1.1842911

10.1016/S0360-3016(03)01579-7

10.1142/S0218202500000501

10.1023/B:JORA.0000010880.40483.c4

Matuszak M. M., Clinical applications of volumetric modulated arc therapy, Int. J. Radiat. Oncol., Biol., Phys., 77, 608, 10.1016/j.ijrobp.2009.08.032

10.1016/S0360-3016(02)03942-1

Yenice K., 1993, A Practical Guide to Intensity‐Modulated Radiation Therapy, 450

10.1118/1.1843354

10.1259/bjr/42085182

10.1088/0031-9155/51/19/008

10.2307/3574407

10.1016/0167-8140(91)90068-R

10.1016/S0167-8140(83)80021-8

10.1118/1.598063

10.1118/1.1543571

10.1016/0167-8140(92)90244-O

Tome W. A., 2005, How can tumor effects and normal tissue effects be balanced in stereotactic body radiotherapy?, Radiosurgery, 6, 86

10.1016/S0169-5002(99)00019-7

10.1016/j.ijrobp.2008.06.1939

10.1016/j.ijrobp.2007.06.043

10.1097/JTO.0b013e318074de34

10.1016/j.semradonc.2008.04.005

10.1007/BF02982830

10.1016/j.ijrobp.2007.10.059

10.1016/j.ijrobp.2008.03.016

10.1016/j.ijrobp.2009.02.027

10.1200/JCO.2006.07.5937

10.1200/JCO.2006.09.7469

10.1016/j.ijrobp.2009.07.090

10.1016/j.ijrobp.2008.04.026

10.1118/1.1521722

10.1007/s00066-007-1695-0

10.1016/j.radonc.2006.07.001

10.1097/00006123-200110000-00011

10.1177/153303460400300107

10.1120/jacmp.v5i2.1947

10.1118/1.1951042

10.1016/S0360-3016(03)00792-2

10.1016/j.ijrobp.2004.10.011

10.1016/j.semradonc.2005.01.003

10.1159/000106031

10.1016/S0360-3016(02)02884-5

10.1016/S0360-3016(03)00090-7

10.1097/00006123-200301000-00018

10.1016/S0167-8140(02)00385-7

10.1016/S0360-3016(00)00748-3

10.1016/0360-3016(94)00382-U

10.1016/j.ijrobp.2004.03.038

10.1016/S0360-3016(96)00341-0

10.1080/02841860600919233

10.1097/01.RVI.0000142599.48497.6B

10.1016/j.ijrobp.2005.04.024

10.1016/j.ijrobp.2005.11.016

10.1016/j.ijrobp.2006.05.029

10.1016/S0360-3016(01)02762-6

10.3171/foc.2001.11.6.7

10.1016/S0022-5223(19)35243-2

10.1016/S0360-3016(01)01703-5

10.1016/S0360-3016(02)04406-1

10.1016/j.meddos.2007.01.007

10.1016/j.semradonc.2005.01.009

10.1016/j.ijrobp.2004.11.009

10.1118/1.2349696

10.1118/1.2106448

10.1016/j.ijrobp.2006.04.031

10.1080/02841860600908962

10.1016/j.ijrobp.2007.01.014

10.1016/j.ijrobp.2006.12.022

10.1080/02841860600904839

10.1016/j.ijrobp.2007.08.023

10.1016/j.ijrobp.2007.08.030

10.1118/1.1869074

10.1016/j.semradonc.2005.01.004

10.1007/BF02518871

10.1016/S0360-3016(97)00055-2

10.1118/1.598837

10.1016/0360-3016(94)90412-X

10.1118/1.598578

10.1016/S0010-4825(00)00026-3

10.1118/1.1984263

10.1177/153303460700600409

10.1002/1097-0150(2000)5:4<263::AID-IGS5>3.0.CO;2-2

10.1118/1.1774132

10.1118/1.2779941

10.1016/j.radonc.2008.08.010

10.1088/0031-9155/41/1/007

10.1016/j.ijrobp.2005.01.032

10.1016/S0360-3016(01)02681-5

10.1118/1.1448826

10.1120/jacmp.v2i4.2596

10.1016/j.radonc.2007.10.038

10.1118/1.596475

10.1088/0031-9155/44/12/401

10.1118/1.1544678

10.1118/1.1543150

10.1088/0031-9155/45/9/306

10.1118/1.1536290

10.1016/S0958-3947(98)00031-4

10.1088/0031-9155/42/3/011

10.1118/1.1501138

10.1118/1.598693

10.1118/1.597256

10.1118/1.598915

10.1118/1.1760191

10.1118/1.597427

10.1016/0360-3016(77)90040-2

10.1118/1.1487421

10.1016/S0958-3947(97)00001-0

10.1118/1.596497

10.1016/j.ijrobp.2005.08.023

N.Papanikolaou J.Battista A.Boyer C.Kappas E.Klein T.Mackie M.Sharpe andJ.Van Dyke AAPM Report No. 85: Tissue Inhomogeneity Corrections for Megavoltage Photon Beams American Association of Physicists in Medicine (2004).

10.1118/1.3006353

10.1016/j.ijrobp.2004.07.701

10.1016/j.jacr.2005.02.009

I. ABT Associates 2008.

10.1016/j.ijrobp.2007.05.092

10.1118/1.3190392

10.1118/1.597316

10.1118/1.597398

P.Linet al. “Specification and acceptance testing for computed tomography scanners ” AAPM Report No. 39 Vol.95(American Institute of Physics Inc. 1993).

10.1118/1.596903

10.1118/1.598373

10.1118/1.1591194

2008, Quality assurance for radiation therapy, quality assurance of radiation therapy: The challenges of advanced technologies symposium, Int. J. Radiat. Oncol., Biol., Phys., 71, S1

10.1016/0360-3016(88)90446-4

10.1016/j.semradonc.2007.07.002

10.1016/S0360-3016(99)00190-X

10.1227/01.NEU.0000141080.54647.11

10.1016/j.ijrobp.2007.07.2385

10.1016/j.lungcan.2005.08.011

10.1080/02841860600907329

10.1227/00006123-199502000-00010

10.1118/1.598005

10.1016/S0360-3016(00)00617-9

10.1016/S0167-8140(00)00226-7

10.1118/1.2143141

10.1016/j.ijrobp.2007.10.002

10.1016/j.ijrobp.2007.05.097

10.1016/j.ijrobp.2007.07.2386

10.1016/j.ijrobp.2007.06.086

R.Timmerman personal communication (26 October2009).