Determination of Metals and Metalloids in Wine Using Inductively Coupled Plasma Optical Emission Spectrometry and Mini-torch

Food Analytical Methods - Tập 10 - Trang 180-190 - 2016
Anetta Zioła-Frankowska1, Marcin Frankowski2
1Faculty of Chemistry, Department of Analytical Chemistry, Adam Mickiewicz University in Poznań, Poznan, Poland
2Faculty of Chemistry, Department of Water and Soil Analysis, Adam Mickiewicz University in Poznań, Poznan, Poland

Tóm tắt

In the present study, an inductively coupled plasma optical emission spectrometer ICPE-9820 with the mini-torch technology that significantly reduces the consumption of plasma gas to 7.0 L min−1 was applied for the multielemental analysis of wine samples. Three sample preparation procedures (microwave digestion in closed vessels, acid digestion in open vessels in a water bath, both with HNO3, and 1:10 dilution with deionized water) were compared for the determination of 15 elements in wines. In all wines analyzed, the lowest concentration was obtained for Li (0.06–0.09 mg L−1) and the highest for K (475–995 mg L−1). The concentration >1 mg L−1 of B, Fe, and Na was determined in all wines analyzed by the three sample preparation procedures. The average concentrations found for the acid extracts for Al (1.31 mg L−1) and Fe (3.40 mg L−1) were higher in comparison to results of sample dilution (0.73 and 1.56 mg L−1 respectively). The recoveries were between 88.2 and 110 %, 65.9 and 106 %, and 62.8 and 109 %, whereas the relative standard deviations (RSDs) varied between 0.04 and 4.62 %, 0.74 and 5.57 %, and 0.70 and 5.80 %, respectively, for the procedure of sample dilution, microwave digestion, and water bath digestion.

Tài liệu tham khảo

Chambre d’Agriculture de la Gironde-Service Vigne et Vin, France (2013) [email protected]

Moreno IM, González-Weller D, Gutierrez V, Marino M, Cameán AM, González AG, Hardisson A (2008) Determination of Al, Ba, Ca, Cu, Fe, K, Mg, Mn, Na, Sr and Zn in red wine samples by inductively coupled plasma optical emission spectroscopy: evaluation of preliminary sample treatments. Microchem J 88:56–61

The Food and Environment Research Agency, UK (2015) www.fapas.com