Comparison of Two Automated Multi‐Channel Blood Cell Counting Systems for Analysis of Blood of Common Domestic Animals

Veterinary Clinical Pathology - Tập 12 Số 2 - Trang 25-32 - 1983
M. G. Weiser1,2
1Department of Pathology College of Veterinary Medicine Colorado State University Ft. Collins, Colorado 80523
2Department of Veterinary Pathobiology College of Veterinary Medicine The Ohio State University Columbus, Ohio 43210

Tóm tắt

Summary

An automated, multi‐channel blood cell counting system (S‐Plus) was compared to a reference counting system using blood samples from 187 animals of four species. The standard red cell bath aperture current of 150 volts (V) was used during analysis of 75% of the samples. At this setting, all samples with a Mean Corpuscular Volume (MCV) greater than 50 fl had accurate erythrocyte counts. As the MCV decreased below 50 fl, the severity of false low erythrocyte counts and false high MCV values increased. The remaining 25% of samples were analyzed with the red cell bath aperture current increased to 200 V. At this setting, only 5% or less of erythrocytes from animals with normal MCV values (>36 fl) were below the erythrocyte threshold. The red cell distribution width values provided by the S‐Plus indicated that equine and bovine erythrocytes have greater anisocytosis than canine and feline erythrocytes. Leukocyte counts were significantly lower on the S‐Plus (p < 0.01). Canine and equine samples most frequently had platelet size distribution within the S‐Plus platelet counting threshold window. Electronic whole blood platelet counting appeared unsatisfactory in cats due to large platelet size and erythrocyte‐platelet size overlap. Small platelet size in cattle indicated that further modifications of the red cell bath aperture current would be required to count and size platelets in this species. Following electronic modifications, this state‐of‐the‐art system appears adaptable to hematologic profiling in most species.

Từ khóa


Tài liệu tham khảo

Koepke JA, 1982, Innovations in Hematology Instrumentation, Laboratory Perspectives, 1, 2

10.1093/ajcp/52.6.679

10.1136/jcp.23.1.68

10.1093/ajcp/74.2.157

Weiser MG, 1982, Erythrocyte Volume Distribution in Clinically Normal Dogs, Cats, Horses and Dairy Cows, Am J Vet Res, 43, 163

10.1177/030098588302000405

10.1177/030098588302000101

10.7326/0003-4819-89-2-200

10.1056/NEJM197203092861001

10.1093/ajcp/68.4.449

10.1093/ajcp/76.5.685

10.1203/00006450-198002000-00013