An intensity ratio calibration method for the accurate determination of Mg/Ca and Sr/Ca of marine carbonates by ICP‐AES

American Geophysical Union (AGU) - Tập 3 Số 1 - 2002
Stephanie de Villiers1, Mervyn Greaves1, Henry Elderfield1
1Department of Earth Sciences University of Cambridge Downing Street, Cambridge CB2 3EQ England, U.K.

Tóm tắt

An inductively coupled plasma‐atomic emission spectroscopy (ICP‐AES) method for the accurate and precise simultaneous measurement of the Mg/Ca and Sr/Ca content of carbonates was established. While a precision of <0.3% (1σ standard deviation (SD)) is easily obtainable for both Mg/Ca and Sr/Ca analysis, a Ca matrix effect complicates achieving similar levels of accuracy with conventional calibration procedures. An alternative ratio calibration procedure is proposed which overcomes the Ca matrix effects and ensures the accuracy of the Mg/Ca and Sr/Ca analysis of marine carbonates to <0.3%, almost an order of magnitude better than conventional calibration methods. The longer‐term precision is <0.1% if the batch run average values are corrected for longer‐term drift. The method is suitable for analysis of foraminiferal calcite and coral aragonite and can easily be adjusted for the analysis of other carbonates or microsamples.

Từ khóa


Tài liệu tham khảo

10.1126/science.257.5070.644

10.1038/35013033

10.1029/1999GC000031

10.1126/science.263.5147.663

10.1016/S0016-7037(99)00197-0

10.1126/science.289.5485.1719

10.1029/1999GC000006

Miller J. C., 1988, Statistics for Analytical Chemistry

Moody J. R., 1988, Recommended inorganic chemicals for calibration, Anal. Chem., 60, 1203A, 10.1021/ac00172a724

10.1016/0016-7037(95)00446-7

10.1039/ja9870200033

10.1016/S0377-8398(97)00028-5

10.1016/S0016-7037(97)00181-6

10.1021/ac981410x

10.1029/1998PA900025

10.1016/S0016-7037(98)00042-8

10.1029/1999PA000460