Validity and Between-Day Reliability of the Cervical Range of Motion (CROM) Device

Journal of Orthopaedic and Sports Physical Therapy - Tập 40 Số 5 - Trang 318-323 - 2010
Isabel Audette1, Jean-Pierre Dumas, Julie N. Côté, Sophie J. De Serres
1School of Physical and Occupational Therapy, McGill University, Montreal, Canada

Tóm tắt

Study Design

Clinical measurement, validity and intrarater reliability study.

Objectives

(1) To confirm the validity and assess between-day test-retest reliability of cervical spine motion measurements made with the cervical range of motion (CROM) device in flexion, extension, bilateral rotation, and bilateral side flexion; (2) to provide meaningful information to clinicians about the standard error of measurement and the minimal detectable change for the CROM device.

Background

Range of motion is a common outcome measure used in the assessment of the cervical spine. The CROM device is one of the tools used to measure cervical range of motion in the clinical setting. However, its psychometric properties are not well established, especially for measurements taken on separate days.

Methods

Quasi-experimental design with 1 group comparison. Twenty healthy adults (9 men and 11 women) participated in this study. Cervical range of motion was simultaneously recorded with the CROM device and the Fastrak motion analysis system for all 6 cervical movements mentioned above. The CROM device was placed on the participant's head consistent with standard clinical procedures. Two Fastrak sensors were positioned with 1 on the forehead and 1 over the spinous process of the T6 vertebra. Test-retest reliability of measurements made with the CROM device was assessed, as well as its standard error of measurement and minimal detectable change, with measures taken on 2 separate days spaced 48 hours apart.

Results

Values obtained by the 2 measuring devices yielded Pearson correlation coefficients ranging between 0.93 and 0.98. Test-retest reliability of measurements of cervical range of motion using the CROM was found to be good, with ICCs ranging between 0.89 and 0.98. The standard errors of measurement across the 6 movements ranged from 1.6° to 2.8° and the minimal detectable changes across the 6 movements ranged from 3.6° to 6.5°.

Conclusions

The measurements made with the CROM were shown to be reliable in all movement directions. J Orthop Sports Phys Ther 2010;40(5):318–323.doi:10.2519/jospt.2010.3180

Từ khóa


Tài liệu tham khảo

10.1016/S1356-689X(03)00009-2

10.1016/S0140-6736(86)90837-8

10.1097/00007632-199908010-00011

10.2519/jospt.2008.0303

10.1016/j.pain.2004.09.004

10.1007/s00586-008-0656-3

10.1186/1477-7525-4-54

10.2519/jospt.2009.2826

10.1097/00007632-199606010-00012

10.2519/jospt.2008.2680

10.2519/jospt.2009.2914

10.2519/jospt.2009.2831

10.1054/math.1995.0248

10.1093/rheumatology/39.4.382

Jordan K, 2004, J Rheumatol, 31, 2207

10.1007/s005860000187

10.1243/PIME_PROC_1991_205_295_02

10.1054/math.2000.0232

Portney LG, 2000, Upper Saddle River, 2

10.1097/00007632-199501000-00001

10.1016/j.pain.2004.12.005

10.2310/6640.2004.15377

10.2519/jospt.2006.36.4.242

10.1097/00007632-200301010-00014

Yang Z, 2005, Conf Proc IEEE Eng Med Biol Soc., 7, 6871

10.1093/ptj/71.2.98

10.1093/ptj/72.11.770

10.1046/j.1526-4610.1997.3701006.x