The determination of total and unbound midazolam in human plasma. A comparison of high performance liquid chromatography, gas chromatography and gas chromatography/mass spectrometry
Tóm tắt
Midazolam concentrations in patients' plasma was determined after extraction with high performance liquid chromatography (HPLC), gas chromatography (GC) and gas chromatography/mass spectrometry (GC/MS). GC was selected for routine plasma assays in terms of selectivity, simplicity, precision, accuracy and sensitivity (0.02 μg/ml); HPLC analysis was less sensitive (0.1 μg/mL) than GC; GC/MS was used for analysis validation. Plasma protein binding of midazolam was determined by GC in patients' plasma after in vitro incubation with midazolam, ultrafiltration and extraction; 5% of the drug was unbound to plasma proteins. Midazolam distribution in lipoprotein fractions separated by ultracentrifugation of plasma obtained from patients on prolonged midazolam treatment was also assayed by GC.
Từ khóa
Tài liệu tham khảo
Amrein R., 1981, Arzneim. Forsch./ Drug Res., 31, 2202
de Vries J. X., 1987, 17th Steenbock Symposium Current advances in Vitamin K research. University of Wisconsin‐Madison, 31
Dixon R., 1982, Res. Commun. Chem. Pathol. Pharmacol., 37, 11
Heizmann P., 1981, Arzneim. Forsch./ Drug Res., 31, 2220
Kohlmeier M., 1986, Arztl. Lab., 32, 46
Vree T. B., 1981, Arzneim. Forsch./ Drug Res., 31, 2215