The Thermal Effects of Kirschner Wire Fixation on Small Bones

A. KHANNA1, S.J. PLESSAS1, P. BARRETT1, L.C. BAINBRIDGE1
1Derbyshire Royal Infirmary and the University of Derby, Derby, UK

Tóm tắt

The aim of this study was to assess bone heating caused by the passage of fine (<2 mm) K-wires of different types. Stainless steel K-wires of trocar and diamond point configurations (0.8–2.0 mm) were drilled into the metatarsal bones from a freshly amputated lower limb at a constant force. Temperature measurements were made by miniature thermocouples inserted into the bones, at 3 second intervals over a period of 3 minutes while each K-wire was drilled three times. The temperature reached varied with the tip configurations and the diameter of the K-wires. Regardless of point configuration, thinner wires generated more heat than thicker ones.


Tài liệu tham khảo

10.1016/0021-9290(68)90026-2 Botte MJ, 1992, Clinical Orthopaedics and Related Research, 276, 194 10.1016/0363-5023(94)90275-5 Gerwig WH, 1996, Surgery, 59, 1065 10.2106/00004623-197254020-00008 10.2106/00004623-198466070-00015 Namba RS, 1987, Clinical Orthopaedics and Related Research, 214, 19, 10.1097/00003086-198701000-00005 10.1016/S0363-5023(89)80006-1 10.1016/0021-9290(76)90095-6 10.3109/17453679308993700