Optimization of a high sensitivity MRI receive coil for parallel human brain imaging

J.A. de Zwart1, P.J. Ledden2, P. van Gelderen1, P. Kellman3, Yuxi Pang1, Renxin Chu1, J.H. Duyn1
1Advanced MRI, LFMI, NINDS, Bethesda, MD, USA
2Nova Medical, Inc., Wakefield, MA, USA
3National Institutes of Health, LCE, NHLBI, Bethesda, MD, USA

Tóm tắt

Two eight-channel MRI receive-only coils were developed to provide whole-brain coverage at 1.5 T and 3.0 T field strength, respectively. Objectives were an image signal-to-noise ratio superior to standard designs throughout the human brain, as well as high parallel imaging performance. Electromagnetic field simulations were used to determine array diameter and inter-element coil gap. Low mutual inductive coupling was achieved at 1.5 and 3.0 T using high-impedance pre-amplifiers. Coils show an average SNR improvement over commercial birdcage coils of 2.4 and 2.3 for the 1.5 T and 3.0 T design, respectively. The mean of the noise-amplification factor related to reconstruction of under-sampled data (g-factor) was 1.03 for 2-fold under-sampled data (rate-2) and 1.22 for rate-3 at 1.5 T. For data acquired with the 3.0 T coil array, these values were respectively 1.06 for rate-2 and 1.37 for rate-3.

Từ khóa

#Magnetic resonance imaging #Humans #Coils #Signal to noise ratio #Magnetic noise #Brain modeling #Acceleration #Magnetic heads #Impedance #Noise measurement

Tài liệu tham khảo

ledden, 2001, Four channel preamplifier decoupled phased array for brain imaging at 1.5 T, Proceedings of the 9th scientific meeting and exhibition ISMRM, 1117 10.1002/mrm.1910160203 10.1002/mrm.1910400204 10.1002/mrm.1910400213 10.1002/1522-2594(200103)45:3<495::AID-MRM1065>3.0.CO;2-V 10.1002/(SICI)1522-2594(199911)42:5<952::AID-MRM16>3.0.CO;2-S 10.1002/mrm.1910380414