Improved interstitial and intravascular MR imaging coils
Proceedings IEEE International Symposium on Biomedical Imaging - Trang 963-965
Tóm tắt
An improved intravascular (IV) and interstitial (IS) prototype magnetic resonance (MR) imaging coil design offers dramatic gains in signal-to-noise ratio (SNR) and imaging volume. A multi-turn rectangular solenoid was wound inside catheters with 1.7 mm (IV) and 3.0 mm (IS) diameters. A factor of 200 in peak SNR gain over the standard birdcage head coil was observed, with the region of improved SNR persisting for a radial distance of 10 and 15 mm respectively. Very high resolution ex-vivo intravascular images and very high sensitivity MR spectroscopy (MRS) are demonstrated. This coil design offers interesting performance options in SNR and imaging volume compared to existing FDA-approved catheter coil designs.
Từ khóa
#Coils #Magnetic resonance imaging #Catheters #Prototypes #Magnetic resonance #Signal design #Signal to noise ratio #Solenoids #Wounds #HeadTài liệu tham khảo
10.1002/jmri.1880020411
10.1002/mrm.1910360415
10.1002/(SICI)1522-2594(199904)41:4<722::AID-MRM10>3.0.CO;2-E
viswanathan, 2001, A microcoil device for local, wide field of view and large gain MRI
konyer, 2001, Interstitial RF coil incorporating multiple micro-catheters, Proc ISMRM-ESMRMB, 2170