Fluorescence energy transfer measurements on cell surfaces: A critical comparison of steady‐state fluorimetric and flow cytometric methods

Wiley - Tập 5 Số 2 - Trang 210-216 - 1984
János Szöllősi1, Lajos Trón1, Sándor Damjanovich1, Stephen H. Helliwell2, Donna J. Arndt‐Jovin2, Thomas M. Jovin3
1Department of Biophysics, Medical University School of Debrecen, H‐4012, Debrecen, Hungary
2Max-Planck-Institut für Biophysikalische Chemie, Abteilung Molekulare Biologie, 3400 Göttingen, Federal Republic of Germany
3Max‐Planck‐Institut für Biophysikalische Chemie, Abteilung Molekulare Biologie, 3400 Göttingen, FRG

Tóm tắt

Abstract

The energy transfer efficiency E was measured between fluorescein‐conjugated concanavalin A (Con A) and rhodamine‐conjugated Con A bound to homogeneous tissue culture cells, the HK22 murine lymphoma cell line. Results from a flow cytometric energy transfer method (FCET) and two different steady‐state fluorimeter methods were compared. The data were found to be in close agreement after careful correction of the steady‐state fluorimetric measurements for contributions from dissociating ligand. The biological variability of the individual cells with respect to E was calculated using an error propagation analysis and were found to be less than the variability in the absolute amount of ligand binding per cell. FCET has a number of advantages over the fluorimetric measurements using suspensions of cells: (1) relatively labile receptor‐ligand complexes can be measured; (2) the analysis can be restricted to undamaged cells by gating the data collection on the light‐scattering signals; (3) heterogeneous populations of cells with respect to donor and acceptor topology can be distinguished by the correlation of E with other cellular parameters derived from additional signals or combinations thereof; and (4) the dynamics of donor‐acceptor redistribution on subpopulations can be measured.

Từ khóa


Tài liệu tham khảo

10.1038/270695a0

Cantor CR, 1980, Biophysical Chemistry, 433

10.1177/27.1.374620

10.1073/pnas.80.19.5985

Dale RE, 1981, Fluorescent Probes, 159

10.1038/264411a0

10.1007/BF01570434

10.1007/978-1-4615-9203-7_6

10.1073/pnas.77.6.3469

10.1038/277661a0

Schlessinger J, 1979, Physical Chemical Aspects of Cell Surface Events in Cellular Regulation, 89

10.1073/pnas.75.6.2659

10.1083/jcb.94.2.406

TrónL SzöllösiJ DamjanovichS HelliwellSH Arndt‐JovinDJ JovinTM:Flow cytometric measurement of fluorescence responance energy transfer on cell surfaces. Quantitative evaluation of the transfer efficiency on a cell‐by‐cell basis. Biophy J1984(in press)

Bevington PR, 1969, Data Reduction and Error Analysis for the Physical Sciences