FID modulus: a simple and efficient technique to phase and align MR spectra

Yann Le Fur1, Patrick J. Cozzone2,3
1Aix-Marseille Université, CNRS, CRMBM UMR 7339, Marseille, France
2Aix Marseille Université, CNRS, CRMBM UMR 7339, Marseille, France
3APHM, Hôpitaux de la Timone, CEMEREM, Marseille, France

Tóm tắt

The post-processing of MR spectroscopic data requires several steps more or less easy to automate, including the phase correction and the chemical shift assignment. First, since the absolute phase is unknown, one of the difficulties the MR spectroscopist has to face is the determination of the correct phase correction. When only a few spectra have to be processed, this is usually performed manually. However, this correction needs to be automated as soon as a large number of spectra is involved, like in the case of phase coherent averaging or when the signals collected with phased array coils have to be combined. A second post-processing requirement is the frequency axis assignment. In standard mono-voxel MR spectroscopy, this can also be easily performed manually, by simply assigning a frequency value to a well-known resonance (e.g. the water or NAA resonance in the case of brain spectroscopy). However, when the correction of a frequency shift is required before averaging a large amount of spectra (due to B 0 spatial inhomogeneities in chemical shift imaging, or resulting from motion for example), this post-processing definitely needs to be performed automatically. Zero-order phase and frequency shift of a MR spectrum are linked respectively to zero-order and first-order phase variations in the corresponding free induction decay (FID) signal. One of the simplest ways to remove the phase component of a signal is to calculate the modulus of this signal: this approach is the basis of the correction technique presented here. We show that selecting the modulus of the FID allows, under certain conditions that are detailed, to automatically phase correct and frequency align the spectra. This correction technique can be for example applied to the summation of signals acquired from combined phased array coils, to phase coherent averaging and to B 0 shift correction. We demonstrate that working on the modulus of the FID signal is a simple and efficient way to both phase correct and frequency align MR spectra automatically. This approach is particularly well suited to brain proton MR spectroscopy.

Tài liệu tham khảo

Higinbotham J, Marshall I (2000) NMR lineshapes and lineshape fitting procedures. Annu Rep NMR Spectrosc 43:59–120 Thiel T, Czisch M, Elbel GK, Hennig J (2002) Phase coherent averaging in magnetic resonance spectroscopy using interleaved navigator scans: compensation of motion artifacts and magnetic field instabilities. Magn Reson Med 47(6):1077–1082 Dydak U, Weiger M, Pruessmann KP, Meier D, Boesiger P (2001) Sensitivity-encoded spectroscopic imaging. Magn Reson Med 46(4):713–722 Maril N, Lenkinski RE (2005) An automated algorithm for combining multivoxel MRS data acquired with phased-array coils. J Magn Reson Imaging 21(3):317–322 Ratiney H, Sdika M, Coenradie Y, Cavassila S, Ormondt D, Graveron-Demilly D (2005) Time-domain semi-parametric estimation based on a metabolite basis set. NMR Biomed 18(1):1–13 Vanhamme L, van den Boogaart A, Van Huffel S (1997) Improved method for accurate and efficient quantification of MRS data with use of prior knowledge. J Magn Reson 129(1):35–43 Brown MA (2004) Time-domain combination of MR spectroscopy data acquired using phased-array coils. Magn Reson Med 52(5):1207–1213 Maudsley AA, Darkazanli A, Alger JR, Hall LO, Schuff N, Studholme C, Yu Y, Ebel A, Frew A, Goldgof D, Gu Y, Pagare R, Rousseau F, Sivasankaran K, Soher BJ, Weber P, Young K, Zhu X (2006) Comprehensive processing, display and analysis for in vivo MR spectroscopic imaging. NMR Biomed 19(4):492–503 Dong Z, Peterson B (2007) The rapid and automatic combination of proton MRSI data using multi-channel coils without water suppression. Magn Reson Imaging 25(8):1148–1154 Natt O, Bezkorovaynyy V, Michaelis T, Frahm J (2005) Use of phased array coils for a determination of absolute metabolite concentrations. Magn Reson Med 53(1):3–8 Prock T, Collins DJ, Dzik-Jurasz AS, Leach MO (2002) An algorithm for the optimum combination of data from arbitrary magnetic resonance phased array probes. Phys Med Biol 47(2):N39–N46 Schaffter T, Bornert P, Leussler C, Carlsen IC, Leibfritz D (1998) Fast 1H spectroscopic imaging using a multi-element head-coil array. Magn Reson Med 40(2):185–193 Antoine JP, Chauvin C, Coron A (2001) Wavelets and related time-frequency techniques in magnetic resonance spectroscopy. NMR Biomed 14(4):265–270 Ebel A, Soher BJ, Maudsley AA (2001) Assessment of 3D proton MR echo-planar spectroscopic imaging using automated spectral analysis. Magn Reson Med 46(6):1072–1078 Bhattacharyya PK, Lowe MJ, Phillips MD (2007) Spectral quality control in motion-corrupted single-voxel J-difference editing scans: an interleaved navigator approach. Magn Reson Med 58(4):808–812 Star-Lack J, Spielman D, Adalsteinsson E, Kurhanewicz J, Terris DJ, Vigneron DB (1998) In vivo lactate editing with simultaneous detection of choline, creatine, NAA, and lipid singlets at 1.5 T using PRESS excitation with applications to the study of brain and head and neck tumors. J Magn Reson 133(2):243–254 Evans CJ, Puts NA, Robson SE, Boy F, McGonigle DJ, Sumner P, Singh KD, Edden RA (2012) Subtraction artifacts and frequency (Mis-)alignment in J-difference GABA editing. J Magn Reson Imaging. doi:10.1002/jmri.23923 Le Fur Y, Nicoli F, Guye M, Confort-Gouny S, Cozzone PJ, Kober F (2010) Grid-free interactive and automated data processing for MR chemical shift imaging data. Magn Reson Mater Phy 23(1):23–30 Serrai H, Clayton DB, Senhadji L, Zuo C, Lenkinski RE (2002) Localized proton spectroscopy without water suppression: removal of gradient induced frequency modulations by modulus signal selection. J Magn Reson 154(1):53–59 Sijbers J, den Dekker AJ, Van Audekerke J, Verhoye M, Van Dyck D (1998) Estimation of the noise in magnitude MR images. Magn Reson Imaging 16(1):87–90 Gudbjartsson H, Patz S (1995) The Rician distribution of noisy MRI data. Magn Reson Med 34(6):910–914 Pijnappel W, Van den Boogaart A, De Beer R, Van Ormondt D (1992) SVD-based quantification of magnetic resonance signals. J Magn Reson 97:97–122 Sdika M, Le Fur Y, Cozzone PJ (2011) Optimal recombination of multi-coils CSI data using image based sensitivity map. In: Proceeding of the 19th annual meeting, international society of magnetic resonance in medicine, Montral, Canada, p 3478 Sdika M, Le Fur Y, Cozzone PJ (2011) Optimal recombination of multi-coils CSI data using image based sensitivity map. In: Proceedings of the European Society of magnetic resonance in medicine and biology, Leipzig, DE, p 583 Maudsley AA, Domenig C, Govind V, Darkazanli A, Studholme C, Arheart K, Bloomer C (2009) Mapping of brain metabolite distributions by volumetric proton MR spectroscopic imaging (MRSI). Magn Reson Med 61(3):548–559 Klose U (1990) In vivo proton spectroscopy in presence of eddy currents. Magn Reson Med 14(1):26–30 Ebel A, Maudsley AA (2005) Detection and correction of frequency instabilities for volumetric 1H echo-planar spectroscopic imaging. Magn Reson Med 53(2):465–469 Chadzynski GL, Klose U (2013) Proton CSI without solvent suppression with strongly reduced field gradient related sideband artifacts. Magn Reson Mater Phy 26(2):183–192 Chadzynski GL, Klose U (2010) Chemical shift imaging without water suppression at 3 T. Magn Reson Imaging 28(5):669–675 Dong Z, Dreher W, Leibfritz D (2004) Experimental method to eliminate frequency modulation sidebands in localized in vivo 1H MR spectra acquired without water suppression. Magn Reson Med 51(3):602–606 Dong Z, Dreher W, Leibfritz D (2006) Toward quantitative short-echo-time in vivo proton MR spectroscopy without water suppression. Magn Reson Med 55(6):1441–1446 Nixon TW, McIntyre S, Rothman DL, de Graaf RA (2008) Compensation of gradient-induced magnetic field perturbations. J Magn Reson 192(2):209–217