Correction of parallel transmission using concurrent RF and gradient field monitoring
Tóm tắt
The accuracy and precision of the parallel RF excitations are highly dependent on the spatial and temporal fidelity of the magnetic fields involved in spin excitation. The consistency between the nominal and effective fields is typically limited by the imperfections of the employed hardware existing both in the gradient system and the RF chain. In this work, we experimentally presented highly improved spatially tailored parallel excitations by turning the native hardware accuracy challenge into a measurement and control problem using an advanced field camera technology to fully correct parallel RF transmission experiment. An array of NMR field probes is used to measure the multiple channel RF pulses and gradient waveforms recording the high power RF pulses simultaneously with low frequency gradient fields on equal time basis. The recorded waveforms were integrated in RF pulse design for gradient trajectory correction, time imperfection compensation and introduction of iterative RF pre-emphasis. Superior excitation accuracy was achieved. Two major applications were presented at 7 Tesla including multi-dimensional tailored RF pulses for spatially selective excitation and slice-selective spoke pulses for
$$B_{1}^{ + }$$
mitigation. Comprehensive field monitoring is a highly effective means of correcting for the field deviations during parallel transmit pulse design.
Tài liệu tham khảo
Pruessmann KP, Weiger M, Scheidegger MB, Boesiger P (1999) SENSE: sensitivity encoding for fast MRI. Magn Reson Med 42:952–962
Sodickson DK, Manning WJ (1997) Simultaneous acquisition of spatial harmonics (SMASH): fast imaging with radiofrequency coil arrays. Magn Reson Med 38:591–603
Griswold MA, Jakob PM, Heidemann RM, Nittka M, Jellus V, Wang JM, Kiefer B, Haase A (2002) Generalized Autocalibrating Partially Parallel Acquisitions (GRAPPA). Magn Reson Med 47:1202–1210
Zhu YD (2004) Parallel excitation with an array of transmit coils. Magn Reson Med 51:775–784
Katscher U, Bornert P, Leussler C, van den Brink JS (2003) Transmit SENSE. Magn Reson Med 49:144–150
Pauly J, Nishimura D, Macovski A (1989) A k-space analysis of small-tip-angle excitation. J Magn Reson 81:43–56
Grissom W, Yip CY, Zhang ZH, Stenger VA, Fessler JA, Noll DC (2006) Spatial domain method for the design of RF pulses in multicoil parallel excitation. Magn Reson Med 56:620–629
Setsompop K, Alagappan V, Gagoski B, Witzel T, Polimeni J, Potthast A, Hebrank F, Fontius U, Schmitt F, Wald LL, Adalsteinsson E (2008) Slice-selective RF pulses for in vivo B(1)(+) inhomogeneity mitigation at 7 tesla using parallel RF excitation with a 16-element coil. Magn Reson Med 60:1422–1432
Brunner DO, Dietrich BE, Cavusoglu M, Wilm BJ, Schmid T, Gross S, Barmet C, Pruessmann KP (2016) Concurrent recording of RF pulses and gradient fields–comprehensive field monitoring for MRI. NMR Biomed 29:1162–1172
Boesch C, Gruetter R, Martin E (1991) Temporal and spatial-analysis of fields generated by eddy currents in superconducting magnets—optimization of corrections and quantitative characterization of magnet gradient systems. Magn Reson Med 20:268–284
Liu Q, Hughes DG, Allen PS (1994) Quantitative characterization of the eddy-current fields in a 40-cm bore superconducting magnet. Magn Reson Med 31:73–76
Wu YH, Chronik BA, Bowen C, Mechefske CK, Rutt BK (2000) Gradient-induced acoustic and magnetic field fluctuations in a 4T whole-body MR imager. Magn Reson Med 44:532–536
Foerster BU, Tomasi D, Caparelli EC (2005) Magnetic field shift due to mechanical vibration in functional magnetic resonance imaging. Magn Reson Med 54:1261–1267
Jehenson P, Westphal M, Schuff N (1990) Analytical method for the compensation of eddy-current effects induced by pulsed magnetic-field gradients in NMR systems. J Magn Reson 90:264–278
Vanvaals JJ, Bergman AH (1990) Optimization of eddy-current compensation. J Magn Reson 90:52–70
Wysong RE, Madio DP, Lowe IJ (1994) A novel eddy-current compensation scheme for pulsed gradient systems. Magn Reson Med 31:572–575
Schneider JT, Haas M, Ruhm W, Hennig J, Ullmann P (2011) Robust spatially selective excitation using radiofrequency pulses adapted to the effective spatially encoding magnetic fields. Magn Reson Med 65:409–421
Wu XP, Vaughan JT, Ugurbil K, Van de Moortele PF (2010) Parallel excitation in the human brain at 9.4 T counteracting k-space errors with RF pulse design. Magn Reson Med 63:524–529
Harkins KD, Does MD, Grissom WA (2014) Iterative method for predistortion of MRI gradient waveforms. IEEE Trans Med Imaging 33:1641–1647
Aldefeld B, Bornert P (1998) Effects of gradient anisotropy in MRI. Magn Reson Med 39:606–614
Grissom WA, Kerr AB, Stang P, Scott GC, Pauly JM (2010) Minimum envelope roughness pulse design for reduced amplifier distortion in parallel excitation. Magn Reson Med 64:1433–1440
Chan F, Pauly J, Macovski A (1992) Effects of RF amplifier distortion on selective excitation and their correction by prewarping. Magn Reson Med 23:224–238
Stang PP, Kerr A, Grissom W, Pauly J, Scott GC (2009) Vector iterative predistortion: An auto-calibration method for transmit arrays. In: Proc Int Soc Mag Reson Med, Honolulu, Hawai, USA
Davies NP, Jezzard P (2005) Calibration of gradient propagation delays for accurate two-dimensional radiofrequency pulses. Magn Reson Med 53:231–236
Scott GC (2007) Signal vector decoupling for transmit arrays. In: Int Soc Magn Reson Med, Berlin, Germany
Zelinski AC, Wald LL, Setsompop K, Goyal VK, Adalsteinsson E (2008) Sparsity-enforced slice-selective MRI RF excitation pulse design. IEEE Trans Med Imaging 27:1213–1229
Tse DHY, Wiggins CJ, Poser BA (2016) Estimating and eliminating the excitation errors in bipolar gradient composite excitations caused by radiofrequency-gradient delay: example of bipolar spokes pulses in parallel transmission. Magn Reson Med. doi:10.1002/mrm.26586
Gras V, Vignaud A, Amadon A, Mauconduit F, Bihan D, Boulant N (2017) New method to characterize and correct with sub-us precision gradient delays in bipolar multispoke RF pulses. Magn Reson Med. doi:10.1002/mrm.26614
Barmet C, De Zanche N, Pruessmann KP (2008) Spatiotemporal magnetic field monitoring for MR. Magn Reson Med 60:187–197
De Zanche N, Barmet C, Nordmeyer-Massner JA, Pruessmann KP (2008) NMR probes for measuring magnetic fields and field dynamics in MR systems. Magn Reson Med 60:176–186
Dietrich BE, Bruneer DO, Barmet C, Wilm BJ, Pruessmann KP (2012) A standalone system for concurrent gradient and RF sequence monitoring Proc Int Soc Magn Reson Med
Zhang J, Huang XJ, Suzuki H (2012) Phase-shifted interpolation for complex signals. IEEE Commun Lett 16:1466–1469
Yarnykh VL (2007) Actual flip-angle imaging in the pulsed steady state: a method for rapid three-dimensional mapping of the transmitted radiofrequency field. Magn Reson Med 57:192–200
Malik SJ, Kenny GD, Hajnal JV (2011) slice profile correction for transmit sensitivity mapping using actual flip angle imaging. Magn Reson Med 65:1393–1399
Bornert P, Schomberg H, Aldefeld B, Groen J (1999) Improvements in spiral MR imaging. Magn Reson Mater Phy 9:29–41
Bornert P, Aldefeld B (1998) On spatially selective RF excitation and its analogy with spiral MR image acquisition. Magn Reson Mater Phy 7:166–178
Yip CY, Fessler JA, Douglas CN (2005) Iterative RF pulse design for multidimensional, small-tip-angle selective excitation. Magn Reson Med 54:908–917
Pauly J, Leroux P, Nishimura D, Macovski A (1991) Parameter relations for the Shinnar-Le Roux selective excitation pulse design algorithm. IEEE Trans Med Imaging 10:53–65
Ma C, Xu D, King KF, Liang ZP (2011) Joint design of spoke trajectories and RF pulses for parallel excitation. Magn Reson Med 65:973–985
Barmet C, De Zanche N, Wilm BJ, Pruessmann KP (2009) A transmit/receive system for magnetic field monitoring of in vivo MRI. Magn Reson Med 62:269–276
Grissom WA, Khalighi MM, Sacolick LI, Rutt BK, Vogel MW (2012) Small-tip-angle spokes pulse design using interleaved greedy and local optimization methods. Magn Reson Med 68:1553–1562
Yip CY, Grissom WA, Fessler JA, Noll DC (2007) Joint design of trajectory and RF pulses for parallel excitation. Magn Reson Med 58:598–604
Vannesjo SJ, Haeberlin M, Kasper L, Pavan M, Wilm BJ, Barmet C, Pruessmann KP (2013) Gradient system characterization by impulse response measurements with a dynamic field camera. Magn Reson Med 69:583–593
Zanchi MG, Stang P, Kerr A, Pauly JM, Scott GC (2011) Frequency-offset cartesian feedback for MRI power amplifier linearization. IEEE Trans Med Imaging 30:512–522
Hoult DI, Foreman D, Kolansky G, Kripiakevich D (2008) Overcoming high-field RF problems with non-magnetic Cartesian feedback transceivers. Magn Reson Mater Phy 21:15–29
Lebsack ET, Wright SM (2002) Iterative RF pulse refinement for magnetic resonance imaging. Ieee T Bio Med Eng 49:41–48
Brunner DO, Pruessmann KP (2009) B-1(+) interferometry for the calibration of RF transmitter arrays. Magn Reson Med 61:1480–1488