An artefact of PET attenuation correction caused by iron overload of the liver in clinical PET-MRI

European Journal of Hybrid Imaging - Tập 1 - Trang 1-6 - 2017
Florian Büther1,2, Benjamin Noto1, Katharina Auf der Springe3, Thomas Allkemper3, Lars Stegger1
1Department of Nuclear Medicine, University Hospital Münster, Münster, Germany
2European Institute for Molecular Imaging, University of Münster, Münster, Germany
3Institute for Clinical Radiology, University Hospital Münster, Münster, Germany

Tóm tắt

Attenuation correction is one of the most important steps in producing quantitative PET image data. In hybrid PET-MRI systems, this correction is far from trivial, as MRI data are not correlated to PET attenuation properties of the scanned object. Commercially available systems often employ correction schemes based on segmenting the body into different tissue classes (air, lung tissue, fat-, and water-like soft tissue), e.g. by using a dual time-point Dixon sequence. However, several pitfalls are known for this approach. Here a specific artefact of MR-based PET attenuation correction is reported, caused by misidentifying the liver as lung tissue due to iron overload. A patient with a history of hematopoietic stem cell transplantation underwent a whole-body [18F]FDG PET-MRI scan. Markedly low liver uptake values were noted in the PET images, seemingly caused by an erroneous assignment of lung tissue attenuation values to the liver. A closer investigation demonstrated markedly low MRI intensity values of the liver, indicative of secondary hemochromatosis (iron overload) most probably due to a history of multiple blood transfusions. Manual assignment of adequate liver attenuation values resulted in more realistic PET images. Iron overload of the liver was identified as a cause of a specific attenuation correction artefact. It remains to be seen how frequent this artefact will be encountered; however, this case highlights that attenuation maps should always be checked during PET image interpretation in hybrid PET-MRI.

Tài liệu tham khảo

Berker Y, Franke J, Salomon A et al (2012) MRI-based attenuation correction for hybrid PET/MRI systems: a 4-class tissue segmentation technique using a combined ultrashort-echo-time/Dixon MRI sequence. J Nucl Med 53:796–804 Boktor RR, Walker G, Stacey R, Gledhill S, Pitman AG (2013) Reference range for Intrapatient variability in blood-pool and liver SUV for 18F-FDG PET. J Nucl Med 54:677–682 Borra RJH, Cho HS, Boewn SL et al (2015) Effects of ferumoxytol on quantitative PET measurements in simultaneous PET/MR whole-body imaging: a pilot study in a baboon model. EJNMMI Physics 2:6 Büther F, Schober O. Positron emission tomography (PET)/computer tomography (CT). In: Comprehensive biomedical physics, volume 1: nuclear medicine and molecular imaging. Brahme a, Ed. Elsevier; 2014:157–180 Hofmann M, Bezrukof I, Mantlik F et al (2011) MRI-based attenuation correction for whole-body PET/MRI: quantitative evaluation of segmentation- and atlas-based methods. J Nucl Med 52:1392–1399 Keller SH, Holm S, Hansen AE et al (2013) Image artifacts from MR-based attenuation correction in clinical, whole-body PET/MRI. Magn Reson Mater Phy. 26:173–181 Martinez-Möller A, Souvatzoglou M, Delso G et al (2009) Tissue classification as a potential approach for attenuation correction in whole-body PET/MRI: evaluation with PET/CT data. J Nucl Med 50:520–526 Queiroz-Andrade M, Blasbalg R, Ortega CD et al (2009) MR imaging findings of iron overload. Radiographics 29:1575–1589 Quick HH (2014) Integrated PET/MR. J Magn Reson Imaging 39:243–258 Rausch I, Rust P, DiFranco MD et al (2016) Reproducibility of MRI Dixon-based attenuation correction in combined PET/MR with applications for lean body mass estimation. J Nucl Med 57:1096–1101 Stelljes M, Hermann S, Albring J et al (2008) Clinical molecular imaging in intestinal graft-versus-host disease: mapping of disease activity, prediction, and monitoring of treatment efficiency by positron emission tomography. Blood 111:2909–2918 Wagenknecht G, Kaiser H-J, Mottaghy FM, Herzog H (2013) MRI for attenuation correction in PET: methods and challenges. Magn Reson Mater Phy 26:99–113