Improved centric phase encoding orders for three‐dimensional magnetization‐prepared mr angiography

Magnetic Resonance in Medicine - Tập 36 Số 3 - Trang 384-392 - 1996
Alan H. Wilman1, Stephen J. Riederer2
1MR Research Laboratory, Mayo Clinic, Rochester, Minnesota 55905, USA.
2MR Research Laboratory, Mayo Clinic, Rochester, Minnesota

Tóm tắt

Abstract

New centric phase encoding orders are introduced for transient state three‐dimensional MR imaging. Unlike previous centric orders, these variants are not fixed with respect to differences in the relative y and z fields of view, but are tailored to account for the differential degrees of gradient dephasing in the ky and kz phase encoding directions and thus maintain the highest weighting for the most central k‐space values. The new centric orders are compared with previous methods in both simulation and experimental magnetization‐prepared MR angiography of the carotid and renal arteries. In comparison with other tested centric orders, the new methods provided better retention of true vessel shape caused by isotropic blurring across the phase encoding plane. In addition, the new methods provided increased contrast, with the degree of improvement over previous methods depending on vessel orientation, size, and degree of inflow.

Từ khóa


Tài liệu tham khảo

10.1002/jmri.1880010509

10.1002/jmri.1880020313

10.1148/radiology.183.2.1561351

10.1148/radiology.187.2.8475281

10.1002/mrm.1910340104

10.1002/mrm.1910310410

10.1002/mrm.1910350322

10.1148/radiology.165.1.3628776

10.1148/radiology.173.2.2798885

10.1002/mrm.1910170215

10.1148/radiology.179.3.2028008

10.1002/jmri.1880020609

10.1002/mrm.1910160315

10.1148/radiology.191.1.8134563

10.1002/mrm.1910050411

10.1016/0730-725X(90)90132-L

10.1002/mrm.1910130109

10.1002/jmri.1880020613

10.1002/mrm.1910300220

10.1002/mrm.1910300416

10.1002/mrm.1910250217

10.1002/mrm.1910310213