Remote Magnetic Versus Manual Catheter Navigation for Ablation of Supraventricular Tachycardias: A Randomized, Multicenter Trial

PACE - Pacing and Clinical Electrophysiology - Tập 31 Số 10 - Trang 1313-1321 - 2008
Mark A. Wood1, Michael V. Orlov2, Karthik Ramaswamy3, Charles I. Haffajee2, Kenneth A. Ellenbogen1
1Virginia Commonwealth University Medical Center, Richmond, Virginia,
2St. Elizabeth's Medical Center, Boston, Massachusetts
3Northeast Georgia Heart Center, Gainesville, Georgia

Tóm tắt

Introduction: The potential benefits of remote robotic navigation for catheter ablation procedures have not been demonstrated in controlled clinical trials. The purpose of this study was to compare remote magnetic catheter navigation to manual navigation for the ablation of common supraventricular arrhythmias.

Methods and Results: Patients with supraventricular arrhythmias due to atrioventricular (AV) nodal reentry, accessory pathways, or undergoing AV junctional ablation for complete heart block were randomized in a 3:1 ratio between magnetic (Niobe system and Helios II catheter, Stereotaxis, Inc., St. Louis, MO) and manual navigation for radiofrequency ablation at 13 centers. The primary endpoint of the study was total fluoroscopic time. Fifty‐six patients were randomized to magnetic navigation and 15 to manual navigation. AV nodal reentry was the most common arrhythmia in both groups. Total fluoroscopy time was reduced in the magnetic navigation group (median 17.8 minutes, interquartile (IQ) range 9.9,27.8 minutes) compared to manual navigation (27.1, IQ 19.0,48.0, P < 0.05). The acute success rates (91% for magnetic and 87% for manual navigation, P > 0.05) did not differ between groups. The number of lesions delivered was less for magnetic navigation (6, IQ 4,9 vs 10, IQ 7, 26, P < 0.05). Total procedure time (median 151, IQ 111, 221 minutes magnetic and 151, IQ 110, 221 minutes manual) and complication rates (5.4% patients magnetic and 6.7% patients manual) were similar between the groups (both P > 0.05).

Conclusions: Remote magnetic catheter navigation reduces fluoroscopic time and radiofrequency lesion deliveries for the ablation of common supraventricular arrhythmias compared to manual catheter navigation.

Từ khóa


Tài liệu tham khảo

10.1161/01.CIR.0000038704.84304.6F

10.1161/01.CIR.0000125126.83579.1B

10.1016/j.jacc.2005.11.058

10.1007/s10840-006-9040-6

10.1016/j.hrthm.2006.01.028

10.1093/europace/euj026

10.1016/j.hrthm.2005.11.019

10.1056/NEJM199207303270504

10.1056/NEJM199106063242301

10.1007/s10840-006-9040-6

10.1093/eurheartj/ehl447

10.1111/j.1540-8167.1993.tb01277.x