Are component positioning and prosthesis size associated with hip resurfacing failure?
Tóm tắt
Recent studies suggest that there is a learning curve for metal-on-metal hip resurfacing. The purpose of this study was to assess whether implant positioning changed with surgeon experience and whether positioning and component sizing were associated with implant longevity. We evaluated the first 361 consecutive hip resurfacings performed by a single surgeon, which had a mean follow-up of 59 months (range, 28 to 87 months). Pre and post-operative radiographs were assessed to determine the inclination of the acetabular component, as well as the sagittal and coronal femoral stem-neck angles. Changes in the precision of component placement were determined by assessing changes in the standard deviation of each measurement using variance ratio and linear regression analysis. Additionally, the cup and stem-shaft angles as well as component sizes were compared between the 31 hips that failed over the follow-up period and the surviving components to assess for any differences that might have been associated with an increased risk for failure. Surgeon experience was correlated with improved precision of the antero-posterior and lateral positioning of the femoral component. However, femoral and acetabular radiographic implant positioning angles were not different between the surviving hips and failures. The failures had smaller mean femoral component diameters as compared to the non-failure group (44 versus 47 millimeters). These results suggest that there may be differences in implant positioning in early versus late learning curve procedures, but that in the absence of recognized risk factors such as intra-operative notching of the femoral neck and cup inclination in excess of 50 degrees, component positioning does not appear to be associated with failure. Nevertheless, surgeons should exercise caution in operating patients with small femoral necks, especially when they are early in the learning curve.
Tài liệu tham khảo
Amstutz HC, Beaule PE, Dorey FJ, Le Duff MJ, Campbell PA, Gruen TA: Metal-on-metal hybrid surface arthroplasty: two to six-year follow-up study. J Bone Joint Surg Am. 2004, 86-A: 28-39.
Bergeron SG, Desy NM, Nikolaou VS, Debiparshad K, Antoniou J: The early results of metal-on-metal hip resurfacing - a prospective study at a minimum two-year follow-up. Bull NYU Hosp Jt Dis. 2009, 67: 132-134.
O'Neill M, Beaule PE, Bin Nasser A, Garbuz D, Lavigne M, Duncan C, Kim PR, Schemitsch E: Canadian academic experience with metal-on-metal hip resurfacing. Bull NYU Hosp Jt Dis. 2009, 67: 128-131.
Amstutz HC, Le Duff MJ: Eleven years of experience with metal-on-metal hybrid hip resurfacing: a review of 1000 conserve plus. J Arthroplasty. 2008, 23: 36-43. 10.1016/j.arth.2008.04.017.
Mont MA, Schmalzried TP: Modern metal-on-metal hip resurfacing: important observations from the first ten years. J Bone Joint Surg Am. 2008, 90 (Suppl 3): 3-11. 10.2106/JBJS.H.00750.
Shimmin A, Beaule PE, Campbell P: Metal-on-metal hip resurfacing arthroplasty. J Bone Joint Surg Am. 2008, 90: 637-654. 10.2106/JBJS.G.01012.
Siebel T, Maubach S, Morlock MM: Lessons learned from early clinical experience and results of 300 ASR hip resurfacing implantations. Proc Inst Mech Eng H. 2006, 220: 345-353.
Stulberg BN, Trier KK, Naughton M, Zadzilka JD: Results and lessons learned from a United States hip resurfacing investigational device exemption trial. J Bone Joint Surg Am. 2008, 90 (Suppl 3): 21-26. 10.2106/JBJS.H.00718.
Marker DR, Seyler TM, Jinnah RH, Delanois RE, Ulrich SD, Mont MA: Femoral neck fractures after metal-on-metal total hip resurfacing: a prospective cohort study. J Arthroplasty. 2007, 22: 66-71. 10.1016/j.arth.2007.05.017.
Schmalzried TP, Fowble VA, Ure KJ, Amstutz HC: Metal on metal surface replacement of the hip. Technique, fixation, and early results. Clin Orthop Relat Res. 1996, 329: S106-114. 10.1097/00003086-199608001-00011.
Beaule PE, Lee JL, Le Duff MJ, Amstutz HC, Ebramzadeh E: Orientation of the femoral component in surface arthroplasty of the hip. A biomechanical and clinical analysis. J Bone Joint Surg Am. 2004, 86: 2015-2021.
De Haan R, Campbell PA, Su EP, De Smet KA: Revision of metal-on-metal resurfacing arthroplasty of the hip: the influence of malpositioning of the components. J Bone Joint Surg Br. 2008, 90: 1158-1163. 10.1302/0301-620X.90B9.19891.
Langton DJ, Jameson SS, Joyce TJ, Webb J, Nargol AV: The effect of component size and orientation on the concentrations of metal ions after resurfacing arthroplasty of the hip. J Bone Joint Surg Br. 2008, 90: 1143-1151.
Liu F, Jin Z, Roberts P, Grigoris P: Importance of head diameter, clearance, and cup wall thickness in elastohydrodynamic lubrication analysis of metal-on-metal hip resurfacing prostheses. Proc Inst Mech Eng H. 2006, 220: 695-704.
Radcliffe IA, Taylor M: Investigation into the effect of varus-valgus orientation on load transfer in the resurfaced femoral head: a multi-femur finite element analysis. Clin Biomech (Bristol, Avon). 2007, 22: 780-786. 10.1016/j.clinbiomech.2007.03.011.
Vail TP, Glisson RR, Dominguez DE, Kitaoka K, Ottaviano D: Position of hip resurfacing component affects strain and resistance to fracture in the femoral neck. J Bone Joint Surg Am. 2008, 90: 1951-1960. 10.2106/JBJS.F.00788.
Lewinnek GE, Lewis JL, Tarr R, Compere CL, Zimmerman JR: Dislocations after total hip-replacement arthroplasties. J Bone Joint Surg Am. 1978, 60: 217-220.
Freeman MA: Some anatomical and mechanical considerations relevant to the surface replacement of the femoral head. Clin Orthop Relat Res. 1978, 19-24.
Hart AJ, Sabah S, Henckel J, Lewis A, Cobb J, Sampson B, Mitchell A, Skinner JA: The painful metal-on-metal hip resurfacing. J Bone Joint Surg Br. 2009, 91: 738-744. 10.1302/0301-620X.91B6.21682.
Cobb JP, Kannan V, Dandachli W, Iranpour F, Brust KU, Hart AJ: Learning how to resurface cam-type femoral heads with acceptable accuracy and precision: the role of computed tomography-based navigation. J Bone Joint Surg Am. 2008, 90 (Suppl 3): 57-64. 10.2106/JBJS.H.00606.
Cobb JP, Kannan V, Brust K, Thevendran G: Navigation reduces the learning curve in resurfacing total hip arthroplasty. Clin Orthop Relat Res. 2007, 463: 90-97.
Seyler TM, Lai LP, Sprinkle DI, Ward WG, Jinnah RH: Does computer-assisted surgery improve accuracy and decrease the learning curve in hip resurfacing? A radiographic analysis. J Bone Joint Surg Am. 2008, 90 (Suppl 3): 71-80. 10.2106/JBJS.H.00697.
Mont MA, Seyler TM, Ulrich SD, Beaule PE, Boyd HS, Grecula MJ, Goldberg VM, Kennedy WR, Marker DR, Schmalzried TP: Effect of changing indications and techniques on total hip resurfacing. Clin Orthop Relat Res. 2007, 465: 63-70.
Romanowski JR, Swank ML: Imageless navigation in hip resurfacing: avoiding component malposition during the surgeon learning curve. J Bone Joint Surg Am. 2008, 90 (Suppl 3): 65-70. 10.2106/JBJS.H.00462.
Hart AJ, Buddhdev P, Winship P, Faria N, Powell JJ, Skinner JA: Cup inclination angle of greater than 50 degrees increases whole blood concentrations of cobalt and chromium ions after metal-on-metal hip resurfacing. Hip Int. 2008, 18: 212-219.
Smith SL, Dowson D, Goldsmith AA: The effect of femoral head diameter upon lubrication and wear of metal-on-metal total hip replacements. Proc Inst Mech Eng H. 2001, 215: 161-170.
The pre-publication history for this paper can be accessed here:http://www.biomedcentral.com/1471-2474/11/227/prepub