Comparison of 1-d and 2-d surface coil arrays for accelerated volume MR imaging using sensitivity encoding

P. Kellman1, M.J. Schnermann, E.R. McVeigh1
1National Institutes of Health, National Heart, Lung, and Blood Institute, Bethesda, MD, USA

Tóm tắt

The sensitivity encoding (SENSE) method of parallel MR accelerated imaging is evaluated and compared for 1-d and 2-d surface coil array geometries. Accelerated MR imaging using SENSE may be applied to volume imaging using either 3-d or 2-d multi-slice acquisition strategies. For higher accelerations such as rate R=4, 3-d SENSE may be applied along either or both phase encode directions. Image quality (SENSE g-factor) is compared for R=4 acceleration implemented with a reduced number of phase encodes in the y-direction, and (assuming a 3-d acquisition) with a reduced number of phase encodes in both yand z-directions. Simulations show that the performance for 1-d and 2-d array geometries depends highly on the slice orientation.

Từ khóa

#Acceleration #Coils #Geometry #Encoding #Solid modeling #Phased arrays #Heart #Lungs #Blood #Educational institutions

Tài liệu tham khảo

10.1002/1522-2586(200102)13:2<313::AID-JMRI1045>3.0.CO;2-W dezwart, 2001, Design of a Sensitivity Optimized SENSE MRI Receive Coil for Brain Imaging, International Society for Magnetic Resonance in Medicine Workshop on Minimum MR Data Acquisition Methods nayfeh, 1985, Electricity and Magnetism, 244 10.1002/mrm.1910160203 10.1002/1522-2594(200103)45:3<495::AID-MRM1065>3.0.CO;2-V 10.1002/(SICI)1522-2594(199911)42:5<952::AID-MRM16>3.0.CO;2-S