Real-time processing of spectroscopic optical coherence tomography

Jian Ye1, A.W. Schaefer1, D.L. Marks1, J.J. Reynolds1, S.A. Boppart1
1Beckman Institute for Advanced Science and Technology, University of Illinois, Urbana-Champaign, Urbana, IL, USA

Tóm tắt

Spectroscopic OCT improves upon the conventional OCT system by processing the entire detected interference signal and extracting spectral information. We implement it as a real-time process using FPGA/DSP based processing hardware. We propose an algorithm based on the Morlet wavelet transform that can be approximated using our existing OCT system and hardware. The spectral information that can be obtained by this system allows for better distinguishing of scattering properties when imaging biological tissue. We also present some of our experimental results to demonstrate the concept. The processing speed achieved by our system allows image generation at approximately the same rate as data acquisition.

Từ khóa

#Spectroscopy #Tomography #Optical scattering #Hardware #Biomedical optical imaging #Coherence #Interference #Signal processing #Signal detection #Data mining

Tài liệu tham khảo

10.1109/2944.796348 10.1364/OL.25.000820 2001, Xilinx Core Generator System 10.1364/OL.22.001811 10.1073/pnas.94.9.4256 1999, TMS320C6000 CPU and Instruction Set Reference Guide 10.1364/OL.25.000111 10.1038/nm0798-861 boppart, 1998, Optical coherence tomography imaging in developmental biology, Methods in Molecular Biology, 135, 217