Validation and Clinical Application of a First Order Step Response Equation for Nitrogen Clearance During FRC Measurement

Journal of Clinical Monitoring and Computing - Tập 22 - Trang 1-9 - 2007
Gary Choncholas1, Soren Sondergaard2, Erkki Heinonen3
1Department of Research & Development, GE Healthcare, Madison, USA
2Department of Anaesthesia, Royal Adelaide Hospital, Adelaide, Australia
3Department of Research & Development, GE Healthcare, Helsinki, Finland

Tóm tắt

To derive a difference equation based on mass conservation and on alveolar tidal volumes for the calculation of Functional Residual Capacity. Derive an equation for the FRC from the difference equation. Furthermore, to derive and validate a step response equation as a solution of the difference equation within the framework of digital signal processing where the FRC is known a priori. A difference equation for the calculation of Functional Residual Capacity is derived and solved as step response of a first order system. The step response equation calculates endtidal fractions of nitrogen during multiple breath nitrogen clearance. The step response equation contains the eigenvalue defined as the ratio of FRC to the sum of FRC and alveolar tidal ventilation. Agreement of calculated nitrogen fractions with measured fractions is demonstrated with data from a metabolic lung model, measurements from patients in positive pressure ventilation and volunteers breathing spontaneously. Examples of eigenvalue are given and compared between diseased and healthy lungs and between ventilatory settings. Comparison of calculated and measured fractions of endtidal nitrogen demonstrates a high degree of agreement in terms of regression and bias and limits of agreement (precision) in Bland & Altman analysis. Examples illustrate the use of the eigenvalue as a possible discriminator between disease states. The first order step response equation reliably calculates endtidal fractions of nitrogen during washout based on a Functional Residual Capacity. The eigenvalue may be a clinically valuable index alone or in conjunction with other indices in the analysis of respiratory states and may aid in the setting of the ventilator.

Tài liệu tham khảo

Darling RC, Cournand A, Richards DW. Studies on the intrapulmonary mixture of gases. III. An open circuit method for measuring residual air. J Clin Invest 1940; 19: 609–618 Lilly JC. Mixing of gases within respiratory system with a new type nitrogen meter. Am J Physiol 1950; 161: 342–351 Robertson JS, Siri WE, Jones HB. Lung ventilation patterns determined by analysis of nitrogen elimination rates; use of mass spectrometer as a continuous gas analyzer. J Clin Invest 1950; 29: 577–590 Wolfe WA, Carlson LD. Studies of pulmonary capacity and mixing with the nitrogen meter. J Clin Invest 1950; 29: 1568–1575 Wrigge H, Sydow M, Zinserling J, Neumann P, Hinz J, Burchardi H. Determination of functional residual capacity (FRC) by multibreath nitrogen washout in a lung model and in mechanically ventilated patients. Accuracy depends on continuous dynamic compensation for changes of gas sampling delay time. Intensive Care Med 1998; 24: 487–493 Darling RC, Cournand A, Richards DW, Domanski B. Studies on intrapulmonary mixture of gases. V. Forms of inadequate ventilation in normal and emphysematous lungs, analyzed by means of breathing pure oxygen. J Clin Invest 1944; 23: 55–67 Fowler WS, Cornish ER Jr., Kety SS. Lung function studies. VIII. Analysis of alveolar ventilation by pulmonary N2 clearance curves. J Clin Invest 1952; 31: 40–50 Wagner PD. Information content of the multibreath nitrogen washout. J Appl Physiol 1979; 46: 579–587 Wauer HJ, Lorenz BA, Kox WJ. FRC measurement in intensive care patients. A definition of standards. Anaesthesist 1998; 47: 844–855 Olegard C, Sondergaard S, Houltz E, Lundin S, Stenqvist O. Estimation of functional residual capacity at the bedside using standard monitoring equipment: a modified nitrogen washout/washin technique requiring a small change of the inspired oxygen fraction. Anesth Analg 2005; 101: 206–212 Oppenheim AV, Schafer RW. Discrete-time signal processing. Englewood Cliffs, N.J.: Prentice Hall, 1989: p. xv, 879 p. Frayn KN. Calculation of substrate oxidation rates in vivo from gaseous exchange. J Appl Physiol 1983; 55: 628–634 Stenqvist O, Sonander H, Lofstrom B, Nilsson K. An oxygen-consuming model lung for evaluation of anaesthetic circuits. Acta Anaesthesiol Scand 1982; 26: 322–326 Bland JM, Altman DG. Measuring agreement in method comparison studies. Stat Methods Med Res 1999; 8: 135–160 Bendjelid K, Schutz N, Suter PM, Romand JA. Continuous cardiac output monitoring after cardiopulmonary bypass: a comparison with bolus thermodilution measurement. Intensive Care Med 2006; 32: 919–922 Rossing RG, Danford MB. A comparison of continuous distributions of parameters of exponential decay curves. Biometrics 1968; 24: 117–134 Gattinoni L, Pesenti A. The concept of “baby lung” Intensive Care Med 2005; 31: 776–784 Anderson D, Nagasawa G, Norfleet W, Olszowka A, Lundgren C. O2 pressures between 0.12 and 2.5 atm abs, circulatory function, and N2 elimination. Undersea Biomed Res 1991; 18: 279–292 Ruppel G. Manual of pulmonary function testing. St. Louis: Mosby, 1998: p. xii, 365 p. Bartels J, Severinghaus JW, Forster RE, Briscoe WA, Bates DV. The respiratory dead space measured by single breath analysis of oxygen, carbon dioxide, nitrogen or helium. J Clin Invest 1954; 33: 41–48 Saidel GM, Lin JS. Transport abnormalities from single-breath dynamics of Ar, CO2 and O2. Respir Physiol 1986; 64: 253–266 Giannella-Neto A, Paoletti P, Fornai E, Giuntini C. Estimates of mean alveolar gas in patients with chronic airways obstruction. Eur Respir J 1989; 2: 451–460 Koulouris NG, Latsi P, Dimitroulis J, Jordanoglou B, Gaga M, Jordanoglou J. Noninvasive measurement of mean alveolar carbon dioxide tension and Bohr’s dead space during tidal breathing. Eur Respir J 2001; 17: 1167–1174