Ultrasound transmission gel as a bolus device for skin irradiation of irregular surfaces: technical note

La radiologia medica - Tập 115 - Trang 975-982 - 2010
G. Catalano1, P. Canino1, M. Cassinotti1, S. Pagella1, V. Piazzi1, S. Re1, G. Wizemann2, E. Bucci3
1Unit of Radiotherapy, MultiMedica Clinical Institute, Castellanza, Italy
2Unit of General Surgery, MultiMedica Clinical Institute, Castellanza, Italy
3Unit of Medical Oncology, MultiMedica Clinical Institute, Castellanza, Italy

Tóm tắt

This paper describes an uncommon radiation treatment of the external ear, with ultrasound (US) transmission gel used as bolus device to compensate for the irregularity of the target surface. Postoperative radiotherapy for cutaneous carcinoma was performed with a single highenergy electron beam directed over the ear auricle. Due to the irregular surface of the target, a “missing tissue” compensator was employed. Daily, after patient positioning, the concha was filled and flattened with US gel, and a dose of 54 Gy in 27 fractions was delivered. Water equivalence of the gel was verified by comparing the gel’s computed tomography (CT) number [Hounsfield units (HU)] and density with the corresponding values for water and another commercial bolus device. Whereas ultrasound gel and water had comparable values (HU: 0; density 1 g/cm3 for both), the corresponding values for the commercial device were slightly higher (HU: 80; density 1.02 g/cm3). Ultrasound gel proved to be an easy, fast and cheap compensating tool. Its water equivalence allows it to be used as an alternative to water, though easier to position and with lower risk of displacement. Thus, it is recommendable as a practical tool for most irregular sites. Further investigations are warranted to validate this solution in more complex irradiation techniques.

Tài liệu tham khảo

Faw FL, Johnson RE, Warren CA et al (1971) A standard set of “individualized” compensating filters for mantle field radiotherapy of Hodgkin’s disease. AJR Am J Roentgentol 111:376–381 Ansbacher W, Robinson DM, Scrimger JW (1992) Missing tissue compensators: evaluation and optimization of a commercial device. Med Phys 19:1267–1273 Kutcher GJ, Burman C, Mohan R (1991) Compensation in three-dimensional non-coplanar treatment planning. Int J Radiat Oncol Biol Phys 20:127–135 Mageras GS, Mohan R, Burman C et al (2001) Compensators for three-dimensional treatment planning. Med Phys 18:133–139 Sharma SC, Derbel FC, Khan RM (1983) Tissue equivalence of bolus materials for electron beams. Radiology 146:854–859 Morrison WH, Wong P, Starkschall G et al (1995) Water bolus for electron irradiation of the ear canal. Int J Radiat Oncol Biol Phys 33:479–483 Hogstrom KR (1991) Clinical electron beam dosimetry: basic dosimetry data. In: Purdy J (ed) Advances in radiation oncology physics: 1990 Proceedings of the summer school of the AAPM. American Institute of Physics, New York, pp 390–429 Perkins GH, McNeese MD, Antolak JA et al (2001) A custom three-dimensional electron bolus technique for optimization of postmastectomy irradiation. J Radiat Oncol Biol Phys 51:1142–1151 Kudchadker RJ, Antolak JA, Morrison WH et al (2003) Utilization of electron bolus in head and neck radiotherapy. J Appl Clin Med Phys 4:321–333 Kong M, Holloway L (2007) An investigation of central axis depth dose distribution perturbation due to air gaps between patient and bolus for electron beams. Australas Phys Eng Med 30:111–119