The reliability and measurement error of protractor-based goniometry of the fingers: A systematic review

Journal of Hand Therapy - Tập 30 - Trang 457-467 - 2017
Yara E. van Kooij1, Alexandra Fink1, Maria W. Nijhuis-van der Sanden2, Caroline M. Speksnijder2,3,4
1Handtherapie Nederland, Utrecht, The Netherlands
2Radboud University Medical Center, Radboud Institute for Health Sciences, Scientific Institute for Quality of Healthcare (IQ Healthcare), Nijmegen, The Netherlands
3Physical Therapy Science, University Medical Center Utrecht, Program in Clinical Health Sciences, Utrecht, The Netherlands
4Department of Oral-Maxillofacial Surgery, University Medical Center Utrecht, Prosthodontics and Special Dental Care, Utrecht, The Netherlands

Tài liệu tham khảo

Pratt, 2001, A review of goniometry use within current hand therapy practice, Br J Hand Therapy, 6, 45, 10.1177/175899830100600202 Groth, 2001, Goniometry of the proximal and distal interphalangeal joints, part II: placement preferences, interrater reliability, and concurrent validity, J Hand Ther, 14, 23, 10.1016/S0894-1130(01)80021-1 Ellis, 1997, Joint angle measurement: a comparative study of the reliability of goniometry and wire tracing for the hand, Clin Rehabil, 11, 314, 10.1177/026921559701100408 Bruton, 1999, Comparison of visual estimation and goniometry for assessment of metacarpophalangeal joint angle, Physiotherapy, 85, 201, 10.1016/S0031-9406(05)65665-9 Ellis, 2002, A study to compare the reliability of composite finger flexion with goniometry for measurement of range of motion in the hand, Clin Rehabil, 16, 562, 10.1191/0269215502cr513oa Groth, 2001, Goniometry of the proximal and distal interphalangeal joints, part I: a survey of instrumentation and placement preferences, J Hand Ther, 14, 18, 10.1016/S0894-1130(01)80020-X Gibson, 2015, Clinical Assessment Recommendations, Vol. 3, 78 Ferenci, 2011, How reliable are measurements of limb movement joints? A comparison of estimates and goniometry [German], Z Physiother, 63, 6 Maissan, 2006, The role of training physiotherapists in the reliability of 'active range of motion' and muscle strength measurements: a pragmatic study [Dutch], Nederlands Tijdschrift Voor Fysiotherapie, 116, 50 Pratt, 2004, An investigation into the degree of precision achieved by a team of hand therapists and surgeons using hand goniometry with a standardised protocol, Br J Hand Therapy, 9, 116, 10.1177/175899830400900401 Mokkink, 2010, The COSMIN study reached international consensus on taxonomy, terminology, and definitions of measurement properties for health-related patient-reported outcomes, J Clin Epidemiol, 63, 737, 10.1016/j.jclinepi.2010.02.006 Terwee, 2009, Development of a methodological PubMed search filter for finding studies on measurement properties of measurement instruments, Qual Life Res, 18, 1115, 10.1007/s11136-009-9528-5 Cambridge-Keeling, 2002, Range-of-Motion Measurement of the Hand, 169 Mokkink, 2010, The COSMIN checklist for evaluating the methodological quality of studies on measurement properties: a clarification of its content, BMC Med Res Methodol, 10, 22, 10.1186/1471-2288-10-22 Terwee, 2007, Quality criteria were proposed for measurement properties of health status questionnaires, J Clin Epidemiol, 60, 34, 10.1016/j.jclinepi.2006.03.012 Mokkink, 2010, The COSMIN checklist for assessing the methodological quality of studies on measurement properties of health status measurement instruments: an international Delphi study, Qual Life Res, 19, 539, 10.1007/s11136-010-9606-8 Terwee, 2012, Rating the methodological quality in systematic reviews of studies on measurement properties: a scoring system for the COSMIN checklist, Qual Life Res, 21, 651, 10.1007/s11136-011-9960-1 Mokkink, 2010, Inter-rater reliability of the COSMIN (COnsensus-based Standards for the selection of health status Measurement Instruments) Checklist, BMC Med Res Methodol, 10, 82, 10.1186/1471-2288-10-82 Dobson, 2012, Measurement properties of performance-based measures to assess physical function in hip and knee osteoarthritis: a systematic review, Osteoarthritis Cartilage, 20, 1548, 10.1016/j.joca.2012.08.015 Bear-Lehman, 1989, Evaluating the hand: issues in reliability and validity, Phys Ther, 69, 1025, 10.1093/ptj/69.12.1025 Brown, 2000, Validity and reliability of the dexter hand evaluation and therapy system in hand-injured patients, J Hand Ther, 13, 37, 10.1016/S0894-1130(00)80051-4 Cook, 2007, Measurements of wrist and finger postures: A comparison of goniometric and motion capture techniques, J Appl Biomech, 23, 70, 10.1123/jab.23.1.70 Edgar, 2009, Goniometry and linear assessments to monitor movement outcomes: Are they reliable tools in burn survivors?, Burns, 1, 58, 10.1016/j.burns.2008.06.010 Hartzell, 2013, Range of motion measurements in hand surgery: should they be used for assessing outcomes?, Ann Plast Surg, 71, 355, 10.1097/SAP.0b013e31824ca676 Harvey, 1996, Computer-assisted evaluation of hand and arm function after thermal injury, J Burn Care Res, 17, 176, 10.1097/00004630-199603000-00015 Lefèvre-Colau, 2001, Reliability of two goniometers in assessing rheumatoid finger mobility: relationship between mobility and disability, Eura Medicophys, 37, 3 Macionis, 2013, Is diagrammatic goniometry feasible for finger ROM evaluation and self-evaluation?, Clin Orthop Relat Res, 471, 1894, 10.1007/s11999-012-2777-6 Marx, 1999, What do we know about the reliability and validity of physical examination tests used to examine the upper extremity?, J Hand Surg Am, 24, 185, 10.1053/jhsu.1999.jhsu24a0185 Smith, 2009, Visual and computer software-aided estimates of Dupuytren's contractures: correlation with clinical goniometric measurements, Ann R Coll Surg Engl, 91, 296, 10.1308/003588409X359259 Tajali, 2011, Intrarater and interrater reliability of electrogoniometric measurements in patients with hand and wrist limitations, J Hand Ther, 24, 383, 10.1016/j.jht.2011.07.015 de Carvalho, 2012, Analysis of the reliability and reproducibility of goniometry compared to hand photogrammetry, Acta Ortop Bras, 20, 139 Carpinella, 2006, Experimental protocol for the kinematic analysis of the hand: definition and repeatability, Gait Posture, 23, 445, 10.1016/j.gaitpost.2005.05.001 Cerveri, 2005, Derivation of centers and axes of rotation for wrist and fingers in a hand kinematic model: methods and reliability results, Ann Biomed Eng, 33, 402, 10.1007/s10439-005-1743-9 Chiu, 1998, The motion analysis system and goniometry of the finger joints, J Hand Surg Br, 23, 788, 10.1016/S0266-7681(98)80098-3 Dipietro, 2003, Evaluation of an instrumented glove for hand-movement acquisition, J Rehabil Res Dev, 40, 179, 10.1682/JRRD.2003.03.0181 Georgeu, 2002, Lateral digital photography with computer-aided goniometry versus standard goniometry for recording finger joint angles, J Hand Surg, 27B, 184, 10.1054/JHSB.2001.0692 Goodson, 2007, Direct, quantitative clinical assessment of hand function: usefulness and reproducibility, Man Ther, 12, 144, 10.1016/j.math.2006.06.002 Gutierrez-Martinez, 2014, System to measure the range of motion of the joints of the human hand, Rev Invest Clin, 66, S122 Harris, 2011, Intrarater and interrater reliability of electro-goniometric measurements in patients with hand and wrist limitations, J Hand Ther, 24, 383, 10.1016/j.jht.2011.07.016 Juul-Kristensen, 2007, Inter-examiner reproducibility of tests and criteria for generalized joint hypermobility and benign joint hypermobility syndrome, Rheumatology (Oxford), 46, 1835, 10.1093/rheumatology/kem290 Kortier, 2014, Assessment of hand kinematics using inertial and magnetic sensors, J Neuroeng Rehabil, 11, 10.1186/1743-0003-11-70 Kulkarni-Lambore, 2000, Reliability and validity of 3Space Isotrak in the measurement of metacarpophalangeal joint range, Physiother Theory Pract, 16, 219, 10.1080/095939800458677 Maury, 2002, Visual estimation of finger angles: Do we need goniometers?, J Hand Surg, 27B, 382 Oess, 2012, Design and evaluation of a low-cost instrumented glove for hand function assessment, J Neuroeng Rehabil, 9, 2, 10.1186/1743-0003-9-2 Torok, 2010, Reliability and validity of the delta finger-to-palm (FTP), a new measure of finger range of motion in systemic sclerosis, Clin Exp Rheumatol, 28, S28 Wang, 2011, A durable, low-cost electrogoniometer for dynamic measurement of joint trajectories, Med Eng Phys, 33, 546, 10.1016/j.medengphy.2010.12.008 Williams, 2000, A goniometric glove for clinical hand assessment. Construction, calibration and validation, J Hand Surg, 25B, 200, 10.1054/jhsb.1999.0360 Wise, 1990, Evaluation of a fiber optic glove for semi-automated goniometric measurements, J Rehabil Res Dev, 27, 411, 10.1682/JRRD.1990.10.0411 Goldsmith, 1998, Inter-rater reliability of two trained raters using a goniometer for the measurement of finger joints, Br J Hand Ther, 3, 11, 10.1177/175899839800300206 Weiss, 1994, Using the Exos Handmaster to measure digital range of motion: reliability and validity, Med Eng Phys, 16, 323, 10.1016/1350-4533(94)90059-0 Stam, 2006, The compangle: A new goniometer for joint angle measurements of the hand: A technical note, Eura Medicophys, 42, 37 Macionis, 2013, Reliability of the standard goniometry and diagrammatic recording of finger joint angles: a comparative study with healthy subjects and non-professional raters, BMC Musculoskelet Disord, 14, 17, 10.1186/1471-2474-14-17 Lewis, 2010, Interrater and intrarater reliability of finger goniometric measurements, Am J Occup Ther, 64, 555, 10.5014/ajot.2010.09028 Kato, 2007, The accuracy of goniometric measurements of proximal interphalangeal joints in fresh cadavers: comparison between methods of measurement, types of goniometers, and fingers, J Hand Ther, 20, 12, 10.1197/j.jht.2006.11.015 Hamilton, 1969, Reliability of goniometers in assessing finger joint angle, Phys Ther, 49, 465, 10.1093/ptj/49.5.465 Engstrand, 2012, Interrater reliability in finger joint goniometer measurement in Dupuytren's disease, Am J Occup Ther, 66, 98, 10.5014/ajot.2012.001925 Armstrong, 1987, Finger flexion function in rheumatoid arthritis: The reliability of eight simple tests, Br J Rheumatol, 26, 118, 10.1093/rheumatology/26.2.118 Burr, 2003, Inter-rater and intra-rater reliability when measuring interphalangeal joints: comparison between three hand-held goniometers, Physiotherapy, 89, 641, 10.1016/S0031-9406(05)60097-1 Lucan, 2013, The reliability of a quality appraisal tool for studies of diagnostic reliability (QAREL), BMC Med Res Methodol, 13, 111, 10.1186/1471-2288-13-111 Armstrong, 1998, Reliability of range-of-motion measurement in the elbow and forearm, J Shoulder Elbow Surg, 7, 573, 10.1016/S1058-2746(98)90003-9 Dent, 1993, Which tests to choose when assessing hand function, J R Coll Surg Edinb, 38, 315 McCulley, 1999, A new multiangle goniometer, Ann Plast Surg, 42, 221 Mentzel, 2001, Reliability of measurement of range of motion in the fingers using a sensor glove, Handchir Mikrochir Plast Chir, 33, 59 Metcalf, 2008, Validation and application of a computational model for wrist and hand movements using surface markers, IEEE Trans Biomed Eng, 55, 1199, 10.1109/TBME.2007.908087 Metcalf, 2011, Modified kinematic technique for measuring pathological hyperextension and hypermobility of the interphalangeal joints, IEEE Trans Biomed Eng, 58, 1224, 10.1109/TBME.2011.2106126 Rippstein, 1977, New aids for appraising and measuring, Orthopade, 6, 81 Okeke, 1999, Comparison of goniometric measurement reliability of hip joint flexion and proximal interphalangeal joint flexion in healthy adults, S Afr J Physiother, 55, 24 Dewey, 2012, Effect of burn bandages on hand goniometry measurements, J Burn Care Res, 1, S153 Vaz, 2013, The case for using the repeatability coefficient when calculating test-retest reliability, PLoS One, 8, e73990, 10.1371/journal.pone.0073990 Portney, 2009, Statistical measures of reliability, Vol. 3, 608 de Vet, 2006, When to use agreement versus reliability measures, J Clin Epidemiol, 59, 1033, 10.1016/j.jclinepi.2005.10.015 Adams, 1999, Clinical Assessment Recommendations, Vol. 2, 56 Hayashi, 2013, Essential motion of metacarpophalangeal joints during activities of daily living, J Hand Ther, 26, 69, 10.1016/j.jht.2012.10.004