Signal strength versus cuff length in nerve cuff electrode recordings

IEEE Transactions on Biomedical Engineering - Tập 49 Số 9 - Trang 1045-1050 - 2002
L.N.S. Andreasen1, J.J. Struijk1
1Center for Sensory Motor Interaction, University of Aalborg, Aalborg, Denmark

Tóm tắt

When a nerve cuff electrode is used for the recording of signals from peripheral nerves, cuff dimensions have to be chosen. Traditionally, the peak-to-peak amplitude of the single-fiber action potential (SFAP) is optimized through the choice of cuff diameter and cuff length. In this paper, the dependency of the root-mean-square (RMS) value of the nerve signal on the cuff dimensions was studied and compared with the peak-to-peak value of the SFAP. A simple approximation for signal optimization by cuff dimensioning is suggested. The results were obtained from modeled SFAPs and from the electroneurogram (ENG) created by superimposed SFAPs, obtained from an inhomogeneous volume conductor model. The results show that the RMS value of the nerve signal is considerably more sensitive to the cuff length than the SFAP peak-to-peak amplitude, and that the RMS of the ENG is a linear function of the fiber diameter.

Từ khóa

#Noise level #Extracellular #Noise reduction #Biomedical electrodes #Conductors #Petroleum #Animals #Humans #Neurofeedback

Tài liệu tham khảo

10.1016/S0006-3495(76)85719-0 10.1016/S0006-3495(97)78890-8 10.1007/BF02345015 10.1385/0-89603-185-3:65 10.1007/BF00236210 10.1113/jphysiol.1978.sp012588 10.1109/86.481970 lickel, 1994, restoration of lateral hand grasp using natural sensors, Proc Topical Worskhop of the Concerted Action RAFT thomsen, 1998, Characterization and optimization of whole nerve recording cuff electrodes stein, 1975, principles underlying new methods for chronic neural recording, Le Journal Canadien des Sciences Neurologiques, 235 10.1109/10.867963 riso, 1996, performance of alternative amplifier configurations for tripolar nerve cuff recorded eng, Proc IEEE-EMBS Annu Meeting 10.1016/0022-5193(71)90155-X andreasen, 2000, configuration for improved snr in nerve cuff electrode recordings, Proc 5th Annu Conf IFESS Soc hoffer, 1974, nerve fiber activity during normal movements, Soc Neurosci, 4, 300