This paper describes the various aspects of safety and performance of a specific commercial infant incubator and its compliance with the set requirements and regulations according to IEC 601-1 and IEC 601-2-19. The qualitative and the quantitative tests of electrical safety have been carried out. In the electrical test, the grounding resistance, leakage current, line voltage, temperature control, and the safety of thermostat were tested for their compliance with the IEC 601/BS5724. The methodology and equipment employed for the compliance tests are also discussed.
Từ khóa
#Circuit testing #Pediatrics #IEC standards #Fasteners #Inspection #Biomedical engineering #Electrical safety #Thermostats #Safety devices #GroundingTài liệu tham khảo
0, British Standard – Safety of Medical Electrical Equipment IEC 601/BS 5724
0, British Standard – Medical Electrical Equipment Part 2 Particular Requirements for Safety section 2 119 Specification For Baby Incubators British Standards Institution (BSI) BS 5724 section 2 119 1991 IEC 601–2–19 1990
0, Ohio® Care™ Plus Service Manual No 6600–0017–000 09/22/92
ding, 2001, Report on Safety of Infant Incubator in Biomedical Engineering