Retrieval and repositioning of an embolized atrial septal defect closure device using a gooseneck snare
Tóm tắt
Closure of all haemodynamically significant atrial septal defects (ASDs) is recommended irrespective of symptoms. Percutaneous device closure offers a favourable alternative to surgery with lower morbidity, shorter duration of hospital stays, and avoidance of a surgical scar. Though device closure is generally a safe procedure with high success rates, certain complications can arise including device embolization which poses a significant challenge for the treating team. We report one such case in which the ASD closure device got spontaneously released and embolized from the delivery cable into the left atrium prior to its deployment. We describe our approach for its retrieval and subsequently its successful deployment across the septal defect using a gooseneck snare. A 5-year-old asymptomatic child was found to have a murmur on a routine check-up. Evaluation revealed a haemodynamically significant, 18-mm ostium secundum ASD with normal pulmonary pressures and suitable margins for device closure. A 20-mm ASD closure device was traversed via an 8-Fr delivery system. While manipulating the left atrial (LA) disc from the right upper pulmonary vein (RUPV) approach, the device got spontaneously released. The right atrial (RA) disc was caught across the ASD, into the left atrium. This was confirmed by intraoperative transthoracic echocardiography and fluoroscopy. The haemodynamics and rhythm were stable. A 20-mm gooseneck snare was immediately passed through the delivery sheath and an attempt was made to catch the screw. With difficulty, the RA screw was caught with the snare and multiple attempts to retrieve the device into the sheath were unsuccessful. However, while negotiating, we were able to secure a favourable position of the device across the atrial septal defect, and after fluoroscopic and echocardiographic confirmation, the device was released. The child remained stable thereafter and was discharged 2 days later. Gooseneck snare is a valuable tool in the management of embolized ASD closure device. Occasionally, like in the index case, one may be successful in retrieving the embolized device and repositioning it across the ASD using a gooseneck snare, thus obviating the need for emergency surgery.
Tài liệu tham khảo
Wu M-H, Chen H-C, Lu C-W, Wang J-K, Huang S-C, Huang S-K (2010) Prevalence of congenital heart disease at live birth in Taiwan. J Pediatr 156(5):782–785. https://doi.org/10.1016/j.jpeds.2009.11.062
Lindsey JB, Hillis LD (2007) Clinical update: atrial septal defect in adults. Lancet 369(9569):1244–1246. https://doi.org/10.1016/S0140-6736(07)60576-5
Turner DR, Owada CY, Sang CJ Jr, Khan M, Lim DS (2017) Closure of secundum atrial septal defects with the AMPLATZER septal occluder: a prospective, multicenter, post-approval study. Circ Cardiovasc Interv 10(8):e004212
Feltes TF, Bacha E, Beekman RH III, Cheatham JP, Feinstein JA, Gomes AS et al (2011) Indications for cardiac catheterization and intervention in pediatric cardiac disease: a scientific statement from the American Heart Association. Circulation. 123(22):2607–2652. https://doi.org/10.1161/CIR.0b013e31821b1f10
Goel PK, Kapoor A, Batra A, Khanna R (2012) Transcatheter retrieval of embolized AMPLATZER Septal Occluder. Tex Heart Inst J 39(5):653–656
Parag B, Sanjeev N, Pruthvi CR, Krishna P, Kumar SB (2020) TCT CONNECT-200 Transthoracic Echocardiography–Guided Device Closure of Atrial Septal Defect (ASD) in Adult and Pediatric Population. J Am Coll Cardiol 76(17 Supplement S):B85
Yang M-C, Wu J-R (2018) Recent review of transcatheter closure of atrial septal defect. Kaohsiung J Med Sci 34(7):363–369. https://doi.org/10.1016/j.kjms.2018.05.001
Wilkinson JL, Goh TH (1998) Early clinical experience with use of the ‘Amplatzer Septal Occluder’device for atrial septal defect. Cardiol Young 8(3):295–302. https://doi.org/10.1017/S104795110000679X
Peuster M, Boekenkamp R, Kaulitz R, Fink C, Hausdorf G (2000) Transcatheter retrieval and repositioning of an Amplatzer device embolized into the left atrium. Catheter Cardiovasc Interv 51(3):297–300. https://doi.org/10.1002/1522-726X(200011)51:3<297::AID-CCD11>3.0.CO;2-O