Prospective Evaluation of a Method for Estimating Ascending Aortic Pressure From the Radial Artery Pressure Waveform

Hypertension - Tập 38 Số 4 - Trang 932-937 - 2001
Alfredo L. Pauca1, Michael F. O’Rourke1,2, Neal D. Kon1
1From the Anesthesiology Department (A.L.P) and Cardiothoracic Surgery (N.D.K.), Wake Forest University School of Medicine, Winston-Salem, NC; St. Vincent’s Clinic/University of New South Wales (M.F.O.), Sydney, Australia.
2Professor of Medicine, Medical Professorial Unit, St. Vincent's Hospital, Victoria St, Darlinghurst, NSW 2010, Australia.

Tóm tắt

Pressure wave reflection in the upper limb causes amplification of the arterial pulse so that radial systolic and pulse pressures are greater than in the ascending aorta. Wave transmission properties in the upper limbs (in contrast to the descending aorta and lower limbs) change little with age, disease, and drug therapy in adult humans. Such consistency has led to use of a generalized transfer function to synthesize the ascending aortic pressure pulse from the radial pulse. Validity of this approach was tested for estimation of aortic systolic, diastolic, pulse, and mean pressures from the radial pressure waveform. Ascending aortic and radial pressure waveforms were recorded simultaneously at cardiac surgery, before initiation of cardiopulmonary bypass, with matched, fluid-filled manometer systems in 62 patients under control conditions and during nitroglycerin infusion. Aortic pressure pulse waves, generated from the radial pulse, showed agreement with the measured aortic pulse waves with respect to systolic, diastolic, pulse, and mean pressures, with mean differences <1 mm Hg. Control differences in Bland-Altman plots for mean±SD in mm Hg were systolic, 0.0±4.4; diastolic, 0.6±1.7; pulse, −0.7±4.2; and mean pressure, −0.5±2.0. For nitroglycerin infusion, differences respectively were systolic, −0.2±4.3; diastolic, 0.6±1.7; pulse, −0.8±4.1; and mean pressure, −0.4±1.8. Differences were within specified limits of the Association for the Advancement of Medical Instrumentation SP10 criteria. In contrast, differences between recorded radial and aortic systolic and pulse pressures were well outside the criteria (respectively, 15.7±8.4 and 16.3±8.5 for control and 14.5±7.3 and 15.1±7.3 mm Hg for nitroglycerin). Use of a generalized transfer function to synthesize radial artery pressure waveforms can provide substantially equivalent values of aortic systolic, pulse, mean, and diastolic pressures.

Từ khóa


Tài liệu tham khảo

10.1161/hyp.32.3.560

10.1161/circ.100.4.354

10.1161/hyp.34.3.372

Breithaupt-Grogler K, Belz G. Epidemiology of arterial stiffness. Pathol Biol. 1999; 47: 604–613.

10.1161/circ.103.9.1245

Nichols WW O’Rourke MF. McDonald’s Blood Flow in Arteries. 4th ed. London UK: Arnold; 1997.

10.1378/chest.102.4.1193

10.1007/BF03011476

10.1161/hyp.25.5.1034

10.1161/circ.94.11.2923

O’Rourke MF Safar ME Dzau V eds. Arterial Vasodilation: Mechanisms and Therapy. Philadelphia: Lea and Febiger; 1993.

10.1093/eurheartj/14.2.160

10.1161/circ.95.7.1827

10.1161/circ.98.2.186

10.1109/10.764946

10.1097/00000542-198103000-00010

10.1097/00000542-199212000-00024

Karamanoglu M, Gallager DE, Avolio AP, O’Rourke MF. Pressure wave propagation in a multibranched model of the human upper limb. Am J Physiol. 1995; 269(pt 2): H1363–H1369.

10.1016/S0140-6736(86)90837-8

Wiggers C. The Pressure Pulse in the Cardiovascular System. New York NY: Longmans; 1928: 47–90 106–118.

10.1152/jappl.1956.8.5.483

10.1161/01.RES.3.6.623

10.1152/jappl.1956.9.3.433

10.1161/circ.37.6.954

10.1093/cvr/4.3.291

10.1093/oxfordjournals.eurheartj.a059669

10.1161/hyp.34.2.201

10.1161/circ.100.9.951

10.1161/circ.100.9.999

10.1056/NEJM200001203420301

10.1056/NEJM200007063430113

10.1152/jappl.1967.23.2.139

10.1161/circ.68.1.6851054

10.1097/00004872-199201000-00014

Gallagher D. Analysis of Pressure Wave Propagation in the Human Upper Limb: Physical Determinants and Clinical Applications [MD thesis]. New South Wales Australia: University of New South Wales; 1993.

Lasance HAJ Wesseling KH Ascoop CA. Peripheral pulse contour analysis in determining stroke volume. In: Progress Report 5. Utrecht Netherlands; Institute of Medical Physics; 1976: 59–62.

O’Rourke MF, Pauca A, Kon N, Tong W, Yung W, Qasem A, Avolio A. Calibrated ascending aortic pressure waves can be derived from the radial pulse using a generalized transfer function. Am J Hypertens. 1999; 12(pt 2): 166A.Abstract.

Soderstrom S, Nyberg J, Ponten J, Sellgren J, O’Rourke MF. Substantial equivalence between ascending aortic pressure waveforms and waveforms derived from the radial pulse using a generalised transfer function? FASEB J. 1998; 12: A712.

Takazawa K, O’Rourke MF, Fujita M, Tanaka N, Takeda K, Kurosu F, Ibukiyama C. Estimation of ascending aortic pressure from radial arterial pressure using a generalised transfer function. Z Kardiol. 1996; 85 (suppl 3): 137–139.

10.1161/01.hyp.21.4.504

10.1042/cs0970291