The distribution of the neutron spectra in the thermal column hole of Xi’an pulse reactor was measured with the time-of-flight method. Compared with the thermal Maxwellian theory neutron spectra, the thermal neutron spectra measured is a little softer, and the average neutron energy of the experimental spectra is about 0.042±0.01 eV. The thermal neutron fluence rate at the front end of thermal column hole, measured with gold foil activation techniques, is about 1.18×105 cm−2 s−1. The standard uncertainty of the measured thermal neutron fluence is about 3%. The spectra-averaged cross section of 197Au(n, γ) determined by the experimental thermal neutron spectra is (92.8±0.93) ×10−24 cm2.