Monte Carlo modeling of the transverse‐axis dose distribution of the Model 200 interstitial brachytherapy source

Medical Physics - Tập 27 Số 4 - Trang 643-654 - 2000
J.F. Williamson1
1Radiation Oncology Center, Mallinckrodt Institute of Radiology, Washington University School of Medicine, St. Louis, Missouri 63110, USA. [email protected]

Tóm tắt

This study presents the first theoretical analysis of the absolute dose‐rate distribution about the Model 200 interstitial brachytherapy source. Monte Carlo photon‐transport (MCPT) simulation techniques have been used to evaluate the transverse‐axis dose‐rate distribution of the Model 200 source as a function of thickness of the Pd metal coating (containing the plated onto the surfaces of right cylindrical graphite pellets contained within the seed. The dose‐rate constant, Λ, was realistically estimated by simulating the wide‐angle, free‐air chamber (WAFAC) calibration geometry. The WAFAC is the experimental realization of NIST's (National Institute of Standards and Technology) recently implemented primary standard of air‐kerma strength Our results show that polar angle‐ and distance‐dependent oblique filtration and shielding effects induce significant and unexpected photon fluence anisotropy near the transverse‐axis and inverse square law deviations at typical calibration distances. Any source consisting of radioactivity deposited on a highly attenuating surface with sharp edges may exhibit such effects. In the case of the Model 200 seed, the Pd metal thickness does not significantly influence the relative dose distribution in water at distances less than 5 cm, but does make Λ sensitive to the measurement geometry. Fortunately, the WAFAC averages fluence over a sufficiently large aperture that the resultant Λ, 0.68±0.02 cGy⋅h−1⋅U−1, is almost independent of Pd metal layer thickness and in close agreement with recent measurements and calculations. This value is 20% higher than that of the renormalized Task Group 43 Λ value.

Từ khóa


Tài liệu tham khảo

10.1118/1.597458

10.1118/1.596692

Meigooni A. S., 1990, Dosimetry of palladium‐103 brachytherapy sources for permanent implants, Endocurietherapy/Hyperthermia Oncology, 6, 107

10.6028/jres.089.017

Seltzer S. M., 1998, New NIST air‐kerma strength standards for I‐125 and Pd‐103 brachytherapy seeds (abstract), Med. Phys., 25, 170

Loevinger R., 1993, Wide‐angle free‐air chamber for calibration of low‐energy brachytherapy sources, Med. Phys., 20, 907

10.1118/1.598456

J. L. Carden “X‐ray emitting interstitial implants ” U.S.A. patent no. 5405309 Theragenics Corporation April 11 1995.

10.1016/S0360-3016(96)00417-8

10.1118/1.596631

10.1118/1.596181

10.1118/1.596069

J. F. Williamson “Monte Carlo simulation of photon transport phenomena ” inMonte Carlo Simulation in the Radiological Sciences edited by R. L. Morin (CRC Press Boca Raton FL 1988) pp. 53–102.

10.1118/1.597088

E. F. Plechaty D. E. Cullen and R. J. Howerton Tables and Graphs of Photo‐Interaction Cross Sections from 0.1 keV to 100 MeV Derived from the LLL Evaluated‐Nuclear‐Data Library UCRL‐50400 Vol. 6 Rev. 2 Lawrence Livermore Laboratory 1978.

R. S. I. C. RSIC HUGO‐VI: Photon Interaction Data in ENDF/B‐V1 Format RSIC Data Library DLC‐146 Oak Ridge National Laboratory 1989.

J. H. Hubbell and S. M. Seltzer Tables of X‐Ray Mass Attenuation Coefficients and Mass Energy‐Absorption Coefficients 1 keV to 20 MeV for Elements Z=1 to 92 and 48 Additional Substances of Dosimetric Interest NISTIR 5632 National Institute of Standards and Technology 1995.

10.1063/1.555523

10.1063/1.555593

10.1063/1.555672

NCRP A Handbook of Radioactivity Measurements NCRP Report No. 58(National Council on Radiation Protection and Measurements Bethseda MD 1985).

10.1118/1.596810

10.1118/1.597484

10.1016/0360-3016(86)90360-3

10.1088/0031-9155/32/2/005

S. M. Seltzer and P. J. Lamperti Status of NIST Primary Standards forandTherapy Seeds Based on the Wide‐Angle Free‐Air Chamber (WAFAC): Proceedings of a CIRMS Workshop National Institute of Standards and Technology 1999.

10.1118/1.598923

B. M. Coursey Model 200 Air‐Kerma Calibration Personal Communication November 1999 National Institutes of Standards and Technology Ionizing Radiation Division Gaithersburg MD.

10.1118/1.598789

10.1118/1.598925

10.1088/0031-9155/44/1/001

10.1118/1.598458

T. Robin and J. Rodgers Theragenics Cd‐109 Source Change History—Theraseed® Model 200 Personal Communication August 30 1999 Theragenics Corporation Buford GA.

10.1118/1.598708

10.1118/1.598570