Issues encountered with kilovoltage x-ray reference dosimetry when changing codes of practice from TRS 277 to TRS 398

Springer Science and Business Media LLC - Tập 32 - Trang 11-15 - 2009
E. Jhala1, B. Steer1, J. Laban2, L. Greig1
1Wellington Blood and Cancer Centre, Wellington Hospital, Wellington, New Zealand
2National Radiation Laboratory, Ministry of Health, Christchurch, New Zealand

Tóm tắt

A difference of 12 % has been observed in the output of an 80 kV, 2.2 mm Al HVL x-ray beam when comparing measurements made according to the TRS 398 medium energy protocol with measurements made according to the TRS 277 low energy protocol. Absorbed dose to water chamber calibration factors used for the TRS 398 measurements were derived from standards of air kerma with the use of the TRS 277 medium energy protocol, and given this fact the discrepancy observed can be considered in terms of a difference between the TRS 277 low and medium energy protocols for this beam. Repeat measurements using the TRS 277 low and medium energy protocols have been made to confirm this. The most likely origins for the discrepancy observed are the chamber perturbation correction,pu, obtained from TRS 277, and the value of the measured percentage depth dose at a depth of 2 g.cm−2 for this beam. Given these findings, reference dosimetry for this beam will be performed according to the TRS 398 low energy protocol.

Tài liệu tham khảo

Code of Safe Practice for the use of Irradiating Apparatus in Medical Therapy. National Radiation Laboratory, NRL C12, 1992 (amended 1998, 2000). Technical report series 277.Absorbed dose determination in photon and electron beams. An international code of practice. Second edition. International Atomic Energy Agency, Vienna, 1997. Technical report series 398.Absorbed Dose Determination in External Beam Radiotherapy: An International Code of Practice for Dosimetry based on Standards of Absorbed Dose to Water. International Atomic Energy Agency, Vienna, 2004. Klevenhagen, S. C., Aukett, R. J., Harrison, R. M., Moretti, C., Nahum, A. E., and Rosser, K. E.,The IPEMB code of practice for the determination of absorbed dose for x-rays below 300 kV generating potential (0.035 mm Al – 4mm Cu HVL; 10–300 kV generating potential). Phys Med Biol 41: 2605–2625, 1996. Ma, C.-M., Coffey, C. W., DeWerd, L. A. and Nath, R., Seltzer S.M., and Seuntjens J.P.,AAPM protocol for 40–300 kV x-ray beam dosimetry in radiotherapy and radiobiology AAPM TG 61, Med Phys 28 (6) June 2001. Grosswendt, B.,Backscatter factors for X-rays generated at voltages between 10 and 100 kV, Phys. Med. Biol. 29: 579–591, 1984. British Journal of Radiology Supplement No. 25Central Axis Depth Dose Data for Use in Radiotherapy: 1996. British Institute of Radiology 1996. Aukett, R. J., Burns, J. E., Greener, A. G., Harrison, R. M., Moretti, C., Nahum, A. E., and Rosser, K. E.,Addendum to the IPEMB code of practice for the determination of absorbed dose for x-rays below 300 kV generating potential (0.035 mm Al – 4mm Cu HVL. Phys Med Biol. 50: 2739–2748, May 2005. Rosenschold, P. M., Nilsson, P. and Knoos, T.,Kilovoltage xray dosimetry — an experimental comparison between different dosimetry protocols. Phys. Med. Biol. 53: 4431–4442, 2008. IAEA-Tecdoc-1455. Implementation of the International Code of Practice on Dosimetry in Radiotherapy (TRS 398): Review of testing results. International Atomic Energy Agency, Vienna, 2005.