Investigating split‐filter dual‐energy CT for improving liver tumor visibility for radiation therapy

Journal of Applied Clinical Medical Physics - Tập 21 Số 8 - Trang 249-255 - 2020
Lianna D. DiMaso1, J. Snow2, M Lawless2, M. Bassetti2, Larry A. DeWerd1, Jessie Y. Huang2
1Department of Medical Physics, University of Wisconsin-Madison, Madison, WI, USA
2Department of Human Oncology, University of Wisconsin-Madison, Madison, WI, USA

Tóm tắt

AbstractPurposeAccurate liver tumor delineation is crucial for radiation therapy, but liver tumor volumes are difficult to visualize with conventional single‐energy CT. This work investigates the use of split‐filter dual‐energy CT (DECT) for liver tumor visibility by quantifying contrast and contrast‐to‐noise ratio (CNR).MethodsSplit‐filter DECT contrast‐enhanced scans of 20 liver tumors including cholangiocarcinomas, hepatocellular carcinomas, and liver metastases were acquired. Analysis was performed on the arterial and venous phases of mixed 120 kVp‐equivalent images and VMIs at 57 keV and 40 keV gross target volume (GTV) contrast and CNR were calculated.ResultsFor the arterial phase, liver GTV contrast was 12.1 ± 10.0 HU and 43.1 ± 32.3 HU (P < 0.001) for the mixed images and 40 keV VMIs. Image noise increased on average by 116% for the 40 keV VMIs compared to the mixed images. The average CNR did not change significantly (1.6 ± 1.5, 1.7 ± 1.4, 2.4 ± 1.7 for the mixed, 57 keV and 40 keV VMIs (> 0.141)). For individual cases, however, CNR increases of up to 607% were measured for the 40 keV VMIs compared to the mixed image. Venous phase 40 keV VMIs demonstrated an average increase of 35.4 HU in GTV contrast and 121% increase in image noise. Average CNR values were also not statistically different, but for individual cases CNR increases of up to 554% were measured for the 40 keV VMIs compared to the mixed image.ConclusionsLiver tumor contrast was significantly improved using split‐filter DECT 40 keV VMIs compared to mixed images. On average, there was no statistical difference in CNR between the mixed images and VMIs, but for individual cases, CNR was greatly increased for the 57 keV and 40 keV VMIs. Therefore, although not universally successful for our patient cohort, split‐filter DECT VMIs may provide substantial gains in tumor visibility of certain liver cases for radiation therapy treatment planning.

Từ khóa


Tài liệu tham khảo

10.18632/oncotarget.9450

10.1148/rg.343135041

10.1007/s00330-010-1964-7

10.1148/radiol.2513081330

10.1148/radiol.11100978

10.1097/RLI.0b013e3181dfda78

10.1148/rg.293085729

10.1148/radiology.199.2.8668803

10.1159/000367764

10.1002/acm2.12435

Almeida IP, 2017, Dual‐energy CT quantitative imaging: a comparison study between twin‐beam and dual‐source CT scanners, MedicalPhysics, 44, 171