Intratester and intertester reliability and criterion validity of the parallelogram and universal goniometers for active knee flexion in healthy subjects
Tóm tắt
A new parallelogram goniometer was designed by the Rehabilitation Centre of the Royal Ottawa Health Care Group in 1983. The advantage of using such a goniometer is that the clinician is not required to estimate the joint axis of rotation when taking a measurement. The parallelogram goniometer has obtained a good intratester and intertester reliability when measuring active range of motion of hip abduction on eight individuals with hip pathologies. However, the validity of the parallelogram goniometer has not been examined. The purposes of this study were to examine the intratester and intertester reliability and the criterion validity of the parallelogram and universal goniometers for active knee flexion on healthy individuals.
Sixty healthy university students (44 females and 16 males; mean age of 20.6 yrs.) participated to this study.
Measurements with the universal and parallelogram goniometers were taken in two different positions, the smaller and larger angles of active knee flexion. All measurements were taken by two trained testers. A radiograph was taken in both positions to serve as the ‘gold standard’. The sequence of the measurements and radiographs were randomly selected. The intra and intertester reliability of both goniometers were established by calculating the intraclass correlation coefficients (ICCs) using the repeated‐measures ANOVA. The criterion validity was examined by calculating Pearson product‐moment correlation coefficients (r) between each goniometric and radiologic measurements. A 0.05 level of significance was chosen for each statistical test.
Intratester reliability ranged from good to excellent for the small angles (ICC=0.85 and 0.87) and the large angles (ICC=0.91 and 0.96) when using the parallelogram goniometer. Intertester reliability was fair for the small angles of flexion (ICC=0.43 to 0.52) and good to excellent for the large angles of flexion (ICC=0.82 to 0.88). The parallelogram goniometer was found to have greater validity when measuring the large angles of knee flexion (r=0.73 and 0.77) compared to the small angles of knee flexion (r=0.33 and 0.41). Similar results of reliability and validity were obtained with the universal goniometer.
The results of this study have clinical importance. The use of the parallelogram goniometer was found to be as reliable and valid as the universal goniometer when measuring active knee flexion. However, the parallelogram goniometer offered clinicians the advantages of obtaining precise angular measurements with fewer adjustments, and a faster application technique. Further studies on the parallelogram goniometer are necessary among individuals presenting with altered range of motion at different joints. Copyright © 1997 Whurr Publishers Ltd.
Từ khóa
Tài liệu tham khảo
BalmerS.Clinical testing of the new parallelogram goniometer. Proceedings of the 10th Canadian Medical and Biological Engineering Conference: Ottawa 1984.
Basmajian JV, 1978, Muscles Alive, 257
Enwemeka CS, 1986, Radiographic verification of knee goniometry, Scandinavian Journal of Rehabilitation Medicine, 18, 47, 10.2340/165019771986184749
Fleiss JL, 1986, The Design and Analysis of Clinical Experiments
Lehmkuhl LD, 1983, Brunnstrom's Clinical Kinesiology, 303
Norkin CC, 1995, Measurement of Joint Motion: A Guide to Goniometry
O'Riain MD, 1983, A new goniometer for physiotherapy, Physiotherapy Canada, 35, 280
SPSS: Statistical Package for Social Sciences Inc., 1994, SPSS\PC+: Version 4.0.