In vitro hydrodynamic properties of the Miethke proGAV hydrocephalus shunt

David Allin1, Zofia Czosnyka1, Marek Czosnyka1, Hugh K. Richards1, John D. Pickard1
1Academic Neurosurgical Unit, Addenbrooke's Hospital, Cambridge, UK

Tóm tắt

Abstract Background

Adjustable shunts are very popular in the management of hydrocephalus and are believed to help in minimizing the number of surgical revisions. The drawback with almost all constructions is that they may be accidentally readjusted in relatively weak magnetic fields (around 30–40 mTesla)

Materials and methods

The ProGav Miethke shunt is composed of an adjustable ballon-spring valve unit and an integrated over-drainage compensating gravitational device (known as the shunt assistant). A mechanical 'brake' is intended to prevent changes to the valve's performance level in a strong magnetic field. We evaluated the performance and hydrodynamic properties of a sample of three valves in the UK Shunt Evaluation Laboratory.

Results

All the shunts showed good mechanical durability over the three-month period of testing, and good stability of hydrodynamic performance over a one-month period

The pressure-flow performance curves, operating, opening and closing pressures fell within the limits specified by the manufacturer, and changed according to the programmed performance levels. The operating pressure increased when the shunt assistant was in the vertical position, as specified. The valve has a low hydrodynamic resistance (0.53 mm mmHg ml-1 min-1). External programming proved to be easy and reliable. Strong magnetic fields from a 3 Tesla MR scanner were not able to change the programming of the valve.

Conclusion

The ProGAV shunt is an adjustable, low resistance valve that is able to limit posture-related over-drainage. Unlike other adjustable valves, the ProGAV cannot be accidentally re-adjusted by external magnetic field such as a 3T MR scanner.

Từ khóa


Tài liệu tham khảo

Albeck MJ, Borgesen SE, Gjerris F, Schmidt JF, Sorensen PS: Intracranial pressure and cerebrospinal fluid outflow conductance in healthy subjects. J Neurosurg. 1991, 74 (4): 597-600.

Inoue T, Kuzu Y, Ogasawara K, Ogawa A: Effect of 3-tesla magnetic resonance imaging on various pressure programmable shunt valves. J Neurosurg. 2005, 163-165. Suppl 2

Nomura S, Fujisawa H, Suzuki M: Effect of cell phone magnetic fields on adjustable cerebrospinal fluid shunt valves. Surg Neurol. 2005, 63: 467-8. 10.1016/j.surneu.2004.06.022.

Anderson RC, Walker ML, Viner JM, Kestle JR: Adjustment and malfunction of a programmable valve after exposure to toy magnets. Case report. J Neurosurg. 2004, 222-5. Suppl 2

Schneider T, Knauff U, Nitsch J, Firsching R: Electromagnetic field hazards involving adjustable shunt valves in hydrocephalus. J Neurosurg. 2002, 96: 331-4.

Sprung C, Miethke C, Schlosser HG, Brock M: The enigma of underdrainage in shunting with hydrostatic valves and possible solutions. Acta Neurochir Suppl. 2005, 95: 229-35.

Lindner D, Preul C, Trantakis C, Moeller H, Meixensberger J: Effect of 3T MRI on the function of shunt valves – evaluation of Paedi GAV, Dual Switch and proGAV. Eur J Radiol. 2005, 56: 56-9. 10.1016/j.ejrad.2005.03.029.

Czosnyka Z, Czosnyka M, Richards HK, Pickard JD: Laboratory Testing of Hydrocephalus Shunts – Conclusion of the U.K. Shunt Evaluation Programme. Acta Neurochir. 2002, 144: 525-538. 10.1007/s00701-002-0922-9.

Czosnyka Z, Cieslicki K, Czosnyka M, Pickard JD: Hydrocephalus shunts and waves of intracranial pressure. Med Biol Eng Comput. 2005, 43: 71-77. 10.1007/BF02345125.

Bech-Azeddine R, Gjerris F, Waldemar G, Czosnyka M, Juhler M: Intraventricular or lumbar infusion test in adult communicating hydrocephalus? Practical consequences and clinical outcome of shunt operation. Acta Neurochir (Wien). 2005, 147: 1027-36. 10.1007/s00701-005-0589-0.

O'Kane MC, Richards H, Winfield P, Pickard JD: The United Kingdom Shunt Registry. Eur J Pediatr Surg. 1997, 1 (Suppl 7): 56.