Tóm tắt

The aim of this study was to evaluate flow from a new dispersive aortic cannula (Stealthflow) in the aortic arch using flow visualization methods. Particle image velocimetry was used to analyze flow dynamics in the mock aortic model. Flow patterns, velocity distribution, and streamlines with different shape cannulas were evaluated in a glass aortic arch model. We compared flow parameters in two different dispersive type cannulas: the Stealthflow and the Soft-flow cannula. A large vortex and regurgitant flow were observed in the aortic arch with both cannulas. With the Stealthflow cannula, a high-velocity area with a maximum velocity of 0.68 m/s appeared on the ostium of the cannula in the longitudinal plane. With the Soft-flow cannula, ‘multiple jet streams, each with a velocity less than 0.60 m/s, were observed at the cannula outlet. Regurgitant flow from the cannula to the brachiocephalic artery and to the ascending aorta on the greater curvature was specific to the Soft-flow cannula. The degree of regurgitation on the same site was lower with the Stealthflow cannula than with the Soft-flow cannula. The Stealthflow cannula has similar flow properties to those of the Soft-flow cannula according to glass aortic model analysis. It generates gentle flow in the aortic arch and slow flow around the ostia of the aortic arch vessels. The Stealthflow cannula is as effective as the Soft-flow cannula. Care must be taken when the patient has thick atheromatous plaque or frail atheroma on the lesser curvature of the aortic arch.

Tài liệu tham khảo

Minakawa M, Fukuda I, Yamazaki J, Fukui K, Yanaoka H, Inamura T. Effect of cannula shape on aortic wall and flow turbulence: hydrodynamic study during extracorporeal circulation in mock thoracic aorta. Artif Organs. 2007;31:880–6. Blauth CI, Cosgrove DM, Webb BW, Ratliff NB, Boylan M, Piedmonte MR, Lytle BW, Loop FD. Atheroembolism from the ascending aorta. An emerging problem in cardiac surgery. J Thorac Cardiovasc Surg. 1992;103:1104–12. Wareing TH, Davila-Roman VG, Barzilai B, Murphy SF, Kouchoukos NT. Management of the severely atherosclerotic ascending aorta during cardiac operations. J Thorac Cardiovasc Surg. 1992;103:453–62. Ura M, Sakata R, Nakayama Y, Goto T. Ultrasonographic demonstration of manipulation-related aortic injuries after cardiac surgery. J Am Coll Cardiol. 2000;35:1303–10. Fukuda I, Minakawa M, Fukui K, Taniguchi S, Daitoku K, Suzuki Y, et al. Breakdown of atheromatous plaque due to shear force from arterial perfusion cannula. Ann Thorac Surg. 2007;84:e17–8. Van der Linden J, Hadjinikolaou L, Bergman P, Lindblom D. Postoperative stroke in cardiac surgery is related to the location and extent of atherosclerotic disease in the ascending aorta. J Am Coll Cardiol. 2001;38:131–5. Weinstein GS. Left hemispheric strokes in coronary surgery: implications for end-hole aortic cannulas. Ann Thorac Surg. 2001;71:128–32. Muehrcke DD, Cornhill JF, Thomas JD, Cosgrove DM. Flow characteristics of aortic cannulae. J Card Surg. 1995;10:514–9.