Glucose Biosensors: An Overview of Use in Clinical Practice

Sensors - Tập 10 Số 5 - Trang 4558-4576
Eun‐Hyung Yoo1, Soo‐Youn Lee2
1Department of Laboratory Medicine, Konyang University Hospital, College of Medical Science Konyang University, Daejon, Korea
2 SungKyunKwan University

Tóm tắt

Blood glucose monitoring has been established as a valuable tool in the management of diabetes. Since maintaining normal blood glucose levels is recommended, a series of suitable glucose biosensors have been developed. During the last 50 years, glucose biosensor technology including point-of-care devices, continuous glucose monitoring systems and noninvasive glucose monitoring systems has been significantly improved. However, there continues to be several challenges related to the achievement of accurate and reliable glucose monitoring. Further technical improvements in glucose biosensors, standardization of the analytical goals for their performance, and continuously assessing and training lay users are required. This article reviews the brief history, basic principles, analytical performance, and the present status of glucose biosensors in the clinical practice.

Từ khóa


Tài liệu tham khảo

Cowie, 2010, Prevalence of diabetes and high risk for diabetes using hemoglobin A1c criteria in the U.S. population in 1988–2006, Diabetes Care, 33, 562, 10.2337/dc09-1524

Narayan, 2006, Impact of recent increase in incidence on future diabetes burden: U.S., 2005–2050, Diabetes Care, 29, 2114, 10.2337/dc06-1136

Wild, 2004, Global prevalence of diabetes: estimates for the year 2000 and projections for 2030, Diabetes Care, 27, 1047, 10.2337/diacare.27.5.1047

Shaw, 2010, Global estimates of the prevalence of diabetes for 2010 and 2030, Diabetes Res. Clin. Pract, 87, 4, 10.1016/j.diabres.2009.10.007

American Diabetes Association (2010). Diagnosis and classification of diabetes mellitus. Diabetes Care, 33, S62–69.

Poolsup, 2009, Meta-analysis of the benefits of self-monitoring of blood glucose on glycemic control in type 2 diabetes patients: an update, Diabetes Technol. Ther, 11, 775, 10.1089/dia.2009.0091

Murata, 2003, Intensified blood glucose monitoring improves glycemic control in stable, insulin-treated veterans with type 2 diabetes: the Diabetes Outcomes in Veterans Study (DOVES), Diabetes Care, 26, 1759, 10.2337/diacare.26.6.1759

Skeie, 2009, Self-Monitoring of Blood Glucose in Type 1 Diabetes Patients with Insufficient Metabolic Control: Focused Self-Monitoring of Blood Glucose Intervention Can Lower Glycated Hemoglobin A1C, J. Diabetes Sci. Technol, 3, 83, 10.1177/193229680900300109

Tunis, 2010, Self-monitoring of blood glucose (SMBG) for type 2 diabetes patients treated with oral anti-diabetes drugs and with a recent history of monitoring: cost-effectiveness in the US, Curr. Med. Res. Opin, 26, 151, 10.1185/03007990903400071

Boutati, 2009, Self-monitoring of blood glucose as part of the integral care of type 2 diabetes, Diabetes Care, 32, S205, 10.2337/dc09-S312

Jovanovic, 2008, Using meal-based self-monitoring of blood glucose as a tool to improve outcomes in pregnancy complicated by diabetes, Endocr. Pract, 14, 239, 10.4158/EP.14.2.239

Pickup, 2009, Self-monitoring of blood glucose in diabetes: is it worth it?, Ann. Clin. Biochem, 46, 273, 10.1258/acb.2009.009011

The Diabetes Control and Complications Trial Research Group (1993). The effect of intensive treatment of diabetes on the development and progression of long-term complications in insulin-dependent diabetes mellitus. N. Engl. J. Med, 329, 977–986.

Holman, 2008, 10-year follow-up of intensive glucose control in type 2 diabetes, N. Engl. J. Med, 359, 1577, 10.1056/NEJMoa0806470

Stratton, 2000, Association of glycaemia with macrovascular and microvascular complications of type 2 diabetes (UKPDS 35): prospective observational study, BMJ, 321, 405, 10.1136/bmj.321.7258.405

American Diabetes Association (2010). Standards of medical care in diabetes--2010. Diabetes Care, 33, S11–61.

Newman, 2005, Home blood glucose biosensors: a commercial perspective, Biosens. Bioelectron, 20, 2435, 10.1016/j.bios.2004.11.012

Turner, 2000, Biosensors--sense and sensitivity, Science, 290, 1315, 10.1126/science.290.5495.1315

Clark, 1962, Electrode systems for continuous monitoring in cardiovascular surgery, Ann. N. Y. Acad. Sci, 102, 29, 10.1111/j.1749-6632.1962.tb13623.x

Updike, 1967, The enzyme electrode, Nature, 214, 986, 10.1038/214986a0

Hiratsuka, 2008, Amperometric biosensor based on glucose dehydrogenase and plasma-polymerized thin films, Anal. Sci, 24, 483, 10.2116/analsci.24.483

Chambers, 2008, Biosensor recognition elements, Curr. Issues Mol. Biol, 10, 1

Iqbal, 2000, A review of molecular recognition technologies for detection of biological threat agents, Biosens. Bioelectron, 15, 549, 10.1016/S0956-5663(00)00108-1

Newman, J.D., and Turner, A.P. (1992). Biosensors: Principles and practice, Portland Press.

Habermuller, 2000, Electron-transfer mechanisms in amperometric biosensors, Fresenius J. Anal. Chem, 366, 560, 10.1007/s002160051551

Pearson, 2000, Analytical aspects of biosensors, Ann. Clin. Biochem, 37, 119, 10.1258/0004563001899131

Thevenot, 2001, Electrochemical biosensors: recommended definitions and classification, Biosens. Bioelectron, 16, 121

Turner, 1999, In vitro diagnostics in diabetes: meeting the challenge, Clin. Chem, 45, 1596, 10.1093/clinchem/45.9.1596

Heller, 1996, Amperometric biosensors, Curr. Opin. Biotechnol, 7, 50, 10.1016/S0958-1669(96)80094-2

Price, 2003, Point-of-care testing in diabetes mellitus, Clin. Chem. Lab. Med, 41, 1213, 10.1515/CCLM.2003.185

Higgins, 1986, Quinoprotein glucose dehydrogenase and its application in an amperometric glucose sensor, Biosensors, 2, 71, 10.1016/0265-928X(86)80011-6

Slein, M.W. (1963). D-glucose: Determination with hexokinase and glucose-6-phosphate dehydrogenase, Academic Press.

Heller, 2008, Electrochemical glucose sensors and their applications in diabetes management, Chem. Rev, 108, 2482, 10.1021/cr068069y

Bankar, 2009, Glucose oxidase--an overview, Biotechnol. Adv, 27, 489, 10.1016/j.biotechadv.2009.04.003

Weibel, 1971, The glucose oxidase mechanism. Interpretation of the pH dependence, J. Biol. Chem, 246, 2734, 10.1016/S0021-9258(18)62246-X

Guilbault, 1973, An enzyme electrode for the amperometric determination of glucose, Anal. Chim. Acta, 64, 439, 10.1016/S0003-2670(01)82476-4

Jin, 1998, PQQ as redox shuttle for quinoprotein glucose dehydrogenase, Biol. Chem, 379, 1207

Zayats, 2005, Reconstitution of apo-glucose dehydrogenase on pyrroloquinoline quinone-functionalized au nanoparticles yields an electrically contacted biocatalyst, J. Am. Chem. Soc, 127, 12400, 10.1021/ja052841h

Raitman, O.A., Patolsky, F., Katz, E., and Willner, I. (2002). Electrical contacting of glucose dehydrogenase by the reconstitution of a pyrroloquinoline quinone-functionalized polyaniline film associated with an Au-electrode: an in situ electrochemical SPR study. Chem. Commun. (Camb.), 1936–1937.

Bartlett, 1987, Strategies for the development of amperometric enzyme electrodes, Biosensors, 3, 359, 10.1016/0265-928X(87)80018-4

Bartlett, 2002, Modified electrodes for NADH oxidation and dehydrogenase-based biosensors, Bioelectrochemistry, 56, 117, 10.1016/S1567-5394(02)00047-6

Gorton, 2002, Electrocatalytic oxidation of NAD(P) H at mediator-modified electrodes, J. Biotechnol, 82, 371

Drury, 1965, DEXTROSTIX--A RAPID METHOD OF ESTIMATING BLOOD GLUCOSE LEVELS, J. Ir. Med. Assoc, 56, 52

Jensen, 1965, Clinical tests with the dextrostix. A new method for rapid blood sugar determination, Ugeskr. Laeger, 127, 709

Korp, 1965, A NEW RAPID BLOOD-SUGAR DETERMINATION AT BEDSIDE (DEXTROSTIX), Wien. Med. Wochenschr, 115, 435

Schmidt, 1965, Blood sugar determination using the dextrostix, Ugesk. Laeger, 127, 706

Updike, 1967, Reagentless substrate analysis with immobilized enzymes, Science, 158, 270, 10.1126/science.158.3798.270

Wang, 2008, Electrochemical glucose biosensors, Chem. Rev, 108, 814, 10.1021/cr068123a

Liu, 2001, Improved design for the glucose biosensor, Food technology and biotechnology, 39, 55

Cass, 1984, Ferrocene-mediated enzyme electrode for amperometric determination of glucose, Anal. Chem, 56, 667, 10.1021/ac00268a018

Frew, 1987, Electrochemical biosensors, Anal. Chem, 59, 933A, 10.1021/ac00142a720

Shichiri, 1982, Wearable artificial endocrine pancrease with needle-type glucose sensor, Lancet, 2, 1129, 10.1016/S0140-6736(82)92788-X

Chaubey, 2002, Mediated biosensors, Biosens. Bioelectron, 17, 441, 10.1016/S0956-5663(01)00313-X

Frew, 1987, Electron-transfer biosensors, Philos. Trans. R. Soc. Lond. B. Biol. Sci, 316, 95, 10.1098/rstb.1987.0020

Ballarin, 2007, Enzyme electrodes based on sono-gel containing ferrocenyl compounds, Biosens. Bioelectron, 22, 1317, 10.1016/j.bios.2006.05.034

Hill, 1986, Homogeneous ferrocene-mediated amperometric immunoassay, Anal. Chem, 58, 1203, 10.1021/ac00297a050

Williams, 1970, Electrochemical-enzymatic analysis of blood glucose and lactate, Anal. Chem, 42, 118, 10.1021/ac60283a032

Hu, 2009, The evolution of commercialized glucose sensors in China, Biosens. Bioelectron, 24, 1083, 10.1016/j.bios.2008.08.051

Hilditch, 1991, Disposable electrochemical biosensors, Analyst, 116, 1217, 10.1039/an9911601217

Matthews, 1987, Pen-sized digital 30-second blood glucose meter, Lancet, 1, 778, 10.1016/S0140-6736(87)92802-9

Murray, 1987, Chemically modified electrodes. Molecular design for electroanalysis, Anal. Chem, 59, 379A

Zhang, 2004, Third-generation biosensors based on the direct electron transfer of proteins, Anal. Sci, 20, 603, 10.2116/analsci.20.603

Gregg, 1990, Cross-linked redox gels containing glucose oxidase for amperometric biosensor applications, Anal. Chem, 62, 258, 10.1021/ac00202a007

Lin, 2005, Carbon nanotubes (CNTs) for the development of electrochemical biosensors, Front. Biosci, 10, 492, 10.2741/1545

Riklin, 1995, Improving enzyme-electrode contacts by redox modification of cofactors, Nature, 376, 672, 10.1038/376672a0

Khan, 1996, Design of a stable charge transfer complex electrode for a third-generation amperometric glucose sensor, Anal. Chem, 68, 2939, 10.1021/ac9510393

Palmisano, 2002, A disposable, reagentless, third-generation glucose biosensor based on overoxidized poly(pyrrole)/tetrathiafulvalene-tetracyanoquinodimethane composite, Anal. Chem, 74, 5913, 10.1021/ac0258608

Azevedo, 2003, Horseradish peroxidase: a valuable tool in biotechnology, Biotechnol. Annu. Rev, 9, 199, 10.1016/S1387-2656(03)09003-3

Mecerreyes, 2004, Design of an amperometric biosensor using polypyrrole-microgel composites containing glucose oxidase, Biosens. Bioelectron, 20, 1111, 10.1016/j.bios.2004.05.018

Vidal, 1998, Electropolymerization of pyrrole and immobilization of glucose oxidase in a flow system: influence of the operating conditions on analytical performance, Biosens. Bioelectron, 13, 371, 10.1016/S0956-5663(97)00116-4

Wu, 2006, Mediator-free amperometric determination of glucose based on direct electron transfer between glucose oxidase and an oxidized boron-doped diamond electrode, Anal. Bioanal. Chem, 385, 1330, 10.1007/s00216-006-0534-y

Albisser, 1974, Clinical control of diabetes by the artificial pancreas, Diabetes, 23, 397, 10.2337/diab.23.5.397

Bindra, 1991, Design and in vitro studies of a needle-type glucose sensor for subcutaneous monitoring, Anal. Chem, 63, 1692, 10.1021/ac00017a008

Csoregi, 1995, Design and optimization of a selective subcutaneously implantable glucose electrode based on “wired” glucose oxidase, Anal. Chem, 67, 1240, 10.1021/ac00103a015

Henry, 1998, Getting under the skin: implantable glucose sensors, Anal. Chem, 70, 594A, 10.1021/ac981937+

Schmidtke, 1998, Measurement and modeling of the transient difference between blood and subcutaneous glucose concentrations in the rat after injection of insulin, Proc. Natl. Acad. Sci. USA, 95, 294, 10.1073/pnas.95.1.294

Rebrin, 2000, Can interstitial glucose assessment replace blood glucose measurements?, Diabetes Technol. Ther, 2, 461, 10.1089/15209150050194332

Gross, 2000, Performance evaluation of the MiniMed continuous glucose monitoring system during patient home use, Diabetes Technol. Ther, 2, 49, 10.1089/152091500316737

Cox, 2009, An overview of continuous glucose monitoring systems, J. Pediatr. Health Care, 23, 344, 10.1016/j.pedhc.2009.06.002

Hashiguchi, 1994, Development of a miniaturized glucose monitoring system by combining a needle-type glucose sensor with microdialysis sampling method. Long-term subcutaneous tissue glucose monitoring in ambulatory diabetic patients, Diabetes Care, 17, 387, 10.2337/diacare.17.5.387

Poscia, 2003, A microdialysis technique for continuous subcutaneous glucose monitoring in diabetic patients (part 1), Biosens. Bioelectron, 18, 891, 10.1016/S0956-5663(02)00216-6

Wentholt, 2005, Comparison of a needle-type and a microdialysis continuous glucose monitor in type 1 diabetic patients, Diabetes Care, 28, 2871, 10.2337/diacare.28.12.2871

Nielsen, 2009, Glucose monitoring by microdialysis: performance in a multicentre study, Diabet. Med, 26, 714, 10.1111/j.1464-5491.2009.02750.x

Tamborlane, 2008, Continuous glucose monitoring and intensive treatment of type 1 diabetes, N. Engl. J. Med, 359, 1464, 10.1056/NEJMoa0805017

Bode, 2009, Sustained benefit of continuous glucose monitoring on A1C, glucose profiles, and hypoglycemia in adults with type 1 diabetes, Diabetes Care, 32, 2047, 10.2337/dc09-0846

Juvenile Diabetes Research Foundation Continuous Glucose Monitoring Study Group (2009). The effect of continuous glucose monitoring in well-controlled type 1 diabetes. Diabetes Care, 32, 1378–1383.

Montagnana, 2009, Continuous glucose monitoring and type 1 diabetes, N. Engl. J. Med, 360, 190, 10.1056/NEJMc082243

Klonoff, 1997, Noninvasive blood glucose monitoring, Diabetes Care, 20, 433, 10.2337/diacare.20.3.433

Oliver, 2009, Glucose sensors: a review of current and emerging technology, Diabet. Med, 26, 197, 10.1111/j.1464-5491.2008.02642.x

Rabinovitch, 1982, Noninvasive glucose monitoring of the aqueous humor of the eye: Part I. Measurement of very small optical rotations, Diabetes Care, 5, 254, 10.2337/diacare.5.3.254

Goetz, 1995, Application of a multivariate technique to Raman spectra for quantification of body chemicals, IEEE Trans. Biomed. Eng, 42, 728, 10.1109/10.391172

Gabriely, 1999, Transcutaneous glucose measurement using near-infrared spectroscopy during hypoglycemia, Diabetes Care, 22, 2026, 10.2337/diacare.22.12.2026

MacKenzie, 1999, Advances in photoacoustic noninvasive glucose testing, Clin. Chem, 45, 1587, 10.1093/clinchem/45.9.1587

Larin, 2002, Noninvasive blood glucose monitoring with optical coherence tomography: a pilot study in human subjects, Diabetes Care, 25, 2263, 10.2337/diacare.25.12.2263

Burnett, 2006, Approved IFCC recommendation on reporting results for blood glucose: International Federation of Clinical Chemistry and Laboratory Medicine Scientific Division, Working Group on Selective Electrodes and Point-of-Care Testing (IFCC-SD-WG-SEPOCT), Clin. Chem. Lab. Med, 44, 1486

Montagnana, 2009, Overview on self-monitoring of blood glucose, Clin. Chim. Acta, 402, 7, 10.1016/j.cca.2009.01.002

Goldstein, 2004, Tests of glycemia in diabetes, Diabetes Care, 27, 1761, 10.2337/diacare.27.7.1761

American Diabetes Association (1996). American Diabetes Association: clinical practice recommendations 1996. Diabetes Care, 19, S1–118.

International Organization for Standardization (2003). In vitro diagnostic test systems-Requirements for blood-glucose monitoring systems for self-testing in managing diabetes mellitus, International Organization for Standardization.

Clinical and Laboratory Standards Institute (2003). Evaluation of the linearity of quantitative measurement procedures; A statistical approach; Approved guideline, Clinical and Laboratory Standards Institute.

Clinical and Laboratory Standards Institute (2005). Interference testing in clinical chemistry; Approved guideline, Clinical and Laboratory Standard Institute.

Heinemann, 2007, Measurement quality of blood glucose meters: is there a need for an institution with an unbiased view?, J. Diabetes Sci. Technol, 1, 178, 10.1177/193229680700100207

Barreau, 1988, Effect of hematocrit concentration on blood glucose value determined on Glucometer II, Diabetes Care, 11, 116, 10.2337/diacare.11.2.116

Rao, 2005, Accuracy evaluation of a new glucometer with automated hematocrit measurement and correction, Clin. Chim. Acta, 356, 178, 10.1016/j.cccn.2005.01.027

Tang, 2000, Effects of different hematocrit levels on glucose measurements with handheld meters for point-of-care testing, Arch. Pathol. Lab. Med, 124, 1135, 10.5858/2000-124-1135-EODHLO

Palmisano, 1993, Ascorbic acid interferences in hydrogen peroxide detecting biosensors based on electrochemically immobilized enzymes, Analytical chemistry, 65, 2690, 10.1021/ac00067a024

Vaidya, 2005, Effect of interference on amperometric glucose biosensors with cellulose acetate membrane, Electroanalysis, 6, 677, 10.1002/elan.1140060811

Schleis, 2007, Interference of maltose, icodextrin, galactose, or xylose with some blood glucose monitoring systems, Pharmacotherapy, 27, 1313, 10.1592/phco.27.9.1313

Flore, 2009, Analytical interferences in point-of-care testing glucometers by icodextrin and its metabolites: an overview, Perit. Dial. Int, 29, 377, 10.1177/089686080902900403

Janssen, 1998, Positive interference of icodextrin metabolites in some enzymatic glucose methods, Clin. Chem, 44, 2379, 10.1093/clinchem/44.11.2379

Oyibo, 2002, Blood glucose overestimation in diabetic patients on continuous ambulatory peritoneal dialysis for end-stage renal disease, Diabet. Med, 19, 693, 10.1046/j.1464-5491.2002.00753.x

Tang, 2000, Effects of drugs on glucose measurements with handheld glucose meters and a portable glucose analyzer, Am. J. Clin. Pathol, 113, 75, 10.1309/QAW1-X5XW-BVRQ-5LKQ

Cartier, 1998, Toxic levels of acetaminophen produce a major positive interference on Glucometer Elite and Accu-chek Advantage glucose meters, Clin. Chem, 44, 893, 10.1093/clinchem/44.4.893

Solnica, 2003, Analytical performance of glucometers used for routine glucose self-monitoring of diabetic patients, Clin. Chim. Acta, 331, 29, 10.1016/S0009-8981(03)00079-2

Vesper, 2007, Approaches for improving glucose monitor measurements for self-monitoring of blood glucose: from measurement harmonization to external quality assessment programs, J. Diabetes Sci. Technol, 1, 153, 10.1177/193229680700100202