Fused deposition modeling of patient‐specific polymethylmethacrylate implants

Emerald - 2010
DavidEspalin1, KarinaArcaute1, DavidRodriguez1, FranciscoMedina1, MatthewPosner2, RyanWicker1
1College of Engineering, W.M. Keck Center for 3D Innovation, The University of Texas at El Paso, El Paso, Texas, USA and Department of Mechanical Engineering, The University of Texas at El Paso, El Paso, Texas, USA
2Department of Orthopedic Surgery, William Beaumont Army Medical Center, El Paso, Texas, USA

Tóm tắt

PurposeThe purpose of this paper is to investigate the use of medical‐grade polymethylmethacrylate (PMMA) in fused deposition modeling (FDM) to fabricate porous customized freeform structures for several applications including craniofacial reconstruction and orthopaedic spacers. It also aims to examine the effects of different fabrication conditions on porosity and mechanical properties of PMMA samples.Design/methodology/approachThe building parameters and procedures to properly and consistently extrude PMMA filament in FDM for building 3D structures were determined. Two experiments were performed that examined the effects of different fabrication conditions, including tip wipe frequency, layer orientation, and air gap (AG) (or distance between filament edges) on the mechanical properties and porosity of the fabricated structures. The samples were characterized through optical micrographs, and measurements of weight and dimensions of the samples were used to calculate porosity. The yield strength, strain, and modulus of elasticity of the samples were determined through compressive testing.FindingsResults show that both the tip wipe frequency (one wipe every layer or one wipe every ten layers) and layer orientation (transverse or axial with respect to the applied compressive load) used to fabricate the scaffolds have effects on the mechanical properties and resulting porosity. The samples fabricate in the transverse orientation with the high tip wipe frequency have a larger compressive strength and modulus than the lower tip wipe frequency samples (compressive strength: 16±0.97 vs 13±0.71 MPa, modulus: 370±14 vs 313±29 MPa, for the high vs low tip wipe frequency, respectively). Also, the samples fabricated in the transverse orientation have a larger compressive strength and modulus than the ones fabricated in the axial orientation (compressive strength: 16±0.97 vs 13±0.83 MPa, modulus: 370±14 vs 281±22 MPa; for samples fabricated with one tip wipe per layer in the transverse and axial orientations, respectively). In general, the stiffness and yield strength decreased when the porosity increased (compressive strength: 12±0.71 to 7±0.95 MPa, Modulus: 248±10 to 165±16 MPa, for samples with a porosity ranging from 55 to 70 percent). As a demonstration, FDM is successfully used to fabricate patient‐specific, 3D PMMA implants with varying densities, including cranial defect repair and femur models.Originality/valueThis paper demonstrates that customized, 3D, biocompatible PMMA structures with varying porosities can be designed and directly fabricated using FDM. By enabling the use of PMMA in FDM, medical implants such as custom craniofacial implants can be directly fabricated from medical imaging data improving the current state of PMMA use in medicine.

Từ khóa


Tài liệu tham khảo

Blecha, L.D., Rakotomanana, L., Razafinmahery, F., Terrier, A. and Pioletti, D.P. (2009), “Targeted mechanical properties for optimal fluid motion inside artificial bone substitutes”, Journal of Orthopedic Research, Vol. 27, pp. 1082‐7.

Eppley, B.L. (2005), “Biomechanical testing of alloplastic PMMA cranioplasty materials”, Journal of Craniofacial Surgery, Vol. 16 No. 1, pp. 140‐3.

Eufinger, H., Wehmoller, M., Harders, A. and Heuser, L. (1995), “Prefabricated prostheses for the reconstruction of skull defects”, International Journal of Oral and Maxillofacial Surgery, Vol. 24, pp. 104‐10.

Frutos Cabanillas, P., Diez Peña, E., Barrales‐Rienda, J.M. and Frutos, G. (2000), “Validation and in‐vitro characterization of antibiotic‐loaded bone cement release”, International Journal of Pharmaceutics, Vol. 209, pp. 15‐26.

Gopakumar, S. (2004), “RP in medicine: a case study in cranial reconstructive surgery”, Rapid Prototyping Journal, Vol. 10 No. 3, pp. 207‐11.

Gronet, P.M., Waskewicz, G.A. and Richardson, C. (2003), “Preformed acrylic cranial implants using fused deposition modeling: a clinical report”, Journal of Prosthetic Dentistry, Vol. 90, pp. 429‐33.

Heinl, P., Muller, L., Korner, C., Singer, R.F. and Muller, F.A. (2008), “Cellular Ti‐6Al‐4V structures with interconnected macro porosity for bone implants fabricated by selective electron beam melting”, Acta Biomaterialia, Vol. 4, pp. 1536‐44.

Kujala, S., Ryhanen, J., Danilov, A. and Tuukkanen, J. (2003), “Effect of porosity on the osteointegration and bone ingrowth of a weight‐bearing nickel‐titanium bone graft substitute”, Biomaterials, Vol. 24, pp. 4691‐97.

Lee, S.C., Wu, C.T., Lee, S.T. and Chen, P.J. (2009), “Cranioplasty using polymethyl methacrylate prostheses”, Journal of Clinical Neuroscience, Vol. 16, pp. 56‐63.

Lewis, G. (1997), “Properties of acrylic bone cement: state of the art review”, Journal of Biomedical Materials Research, Vol. 38 No. 2, pp. 155‐82.

Materialise (2009), Medical Solutions: Cranioplasty Medical Models, available at: www.materialise.com/materialise/view/en/421133‐Cranioplasty.html.

Osswald, T., Baur, E., Brinkmann, S., Oberbach, K. and Schmachtenberg, E. (2006), International Plastics Handbook: the Resource for Plastics Engineers, Hanser Gardner, Cincinnati, OH.

Pelletier, M.H., Malisano, L., Smitham, P.J., Okamoto, K. and Walsh, W.R. (2009), “The compressive properties of bone cements containing large doses of antibiotics”, The Journal of Arthoplasty, Vol. 24 No. 3, pp. 454‐61.

Ruhé, Q.P., Wolke, J.G.C., Spauwen, P.H.M. and Jansen, J.A. (2006), “Calcium phosphate ceramics for bone tissue engineering”, in Bronzino, J. (Ed.), Tissue Engineering and Artificial Organs, The Biomedical Engineering Handbook, 3rd ed., CRC Taylor & Francis, Boca Raton, FL, pp. 38‐1‐38‐18.

Wang, T.M., Xi, J.T. and Jin, Y. (2007), “A model research for prototype warp deformation in the FDM process”, International Journal of Advanced Manufacturing Technology, Vol. 33 Nos 11/12, pp. 1087‐96.

Zein, I., Hutmacher, D.W., Tan, K.C. and Teoh, S.H. (2002), “Fused deposition modeling of novel scaffold architectures for tissue engineering applications”, Biomaterials, Vol. 23 No. 4, pp. 1169‐85.

Cai, S. and Juntong, X. (2008), “A control approach for pore size distribution in the bone scaffold based on the hexahedral mesh refinement”, Computer‐Aided Design, Vol. 40 Nos 10/11, pp. 1040‐50.