Extraction of electron beam dose parameters from EBT2 film data scored in a mini phantom

Springer Science and Business Media LLC - Tập 36 - Trang 339-346 - 2013
Dedri O’Reilly1, Cobus J. L. Smit1, Freek C. P. du Plessis1
1Department of Medical Physics, Faculty of Health Sciences, University of the Free State, Bloemfontein, South Africa

Tóm tắt

Quality assurance of medical linear accelerators includes dosimetric parameter measurement of therapeutic electron beams e.g. relative dose at a depth of 80 % (R80). This parameter must be within a tolerance of 0.2 cm of the declared value. Cumbersome water tank measurements can be regarded as a benchmark to measure electron depth dose curves. A mini-phantom was designed and built, in which a strip of GAFCHROMIC® EBT2 film could be encased tightly for electron beam depth dose measurement. Depth dose data were measured for an ELEKTA Sl25 MLC, ELEKTA Precise, and ELEKTA Synergy (Elekta Oncology Systems, Crawley, UK) machines. The electron beam energy range was between 4 and 22 MeV among the machines. A 10 × 10 cm2 electron applicator with 95 cm source-surface-distance was used on all the machines. 24 h after irradiation, the EBT2 film strips were scanned on Canon CanoScan N670U scanner. Afterwards, the data were analysed with in-house developed software that entailed optical density to dose conversion, and optimal fitting of the PDD data to de-noise the raw data. From the PDD data R80 values were solved for and compared with acceptance values. A series of tests were also carried out to validate the use of the scanner for film Dosimetry. These tests are presented in this study. It was found that this method of R80 evaluation was reliable with good agreement with benchmark water tank measurements using a commercial parallel plate ionization chamber as the radiation detector. The EBT2 film data yielded R80 values that were on average 0.06 cm different from benchmark water tank measured R80 values.

Tài liệu tham khảo

Khan FM (2003) The physics of radiation therapy. Williams and Wilkins, Baltimore Podgorsak EB (2004) Radiation physics for medical physicists. McHill University Health Center, Montreal Hogstrom KR, Almond PR (2006) Review of electron beam therapy physics. Phys Med Biol 51(13):R455–R489 International Commision on Radiation Units and Measurements: Radiation Dosimetry (1984) Electrons with initial energies between 1 and 50 MeV, ICRU Report 35, Washington, DC Kutcher GJ, Coia L, Gillin M, Hanson WF, Leibel S, Morton RJ et al (1994) Comprehensive QA for radiation oncology: report of AAPM radiation therapy committee task group 40. Med Phys 21(4):581–681 Kirkby C, Ghasroddashti E, Angers C. (2013) Technical quality control guidelines for Canadian radiation treatment centres: medical linear accelerators and multileaf collimators. Rogers DWO (1998) Accuracy of the Burns equation for stopping-power ratio as a function of depth and R50. Med Phys 25(3):2961–2963 Richley L, John AC, Coomber H, Fletcher S (2010) Evaluation and optimization of the new EBT2 radio-chromic film dosimetry system. Phys Med Biol 55:2601–2617 Devic S, Seuntjens J, Hegyi G, Podgorsak EB, Soares CG, Kirov AS et al (2004) Dosimetric properties of improved Gafchromic films for seven different digitizers. Med Phys 31(9):2392–2401 Andres C, del Castillo A, Tortosa R, Alonso D, Barquero R (2010) A comprehensive study of the Gafchromic EBT2 radio-chromic film. A comparison with EBT. Med Phys 37(12):6271–6278 Aland T, Kairn T, Kenny J (2011) Evaluation of Gafchromic EBT2 dosimetry system for radiotherapy quality assurance. Australas Phys Eng Sci Med 34:251–260 Aldelaijan S, Devic S, Mohammed HT, Liang L, DeBlois F et al (2010) Evaluation of EBT2 model Gafchromic™ film performance in water. Med Phys 37(7):3687–3693 Arjomandy B, Tailor R, Anand A, Shaoo N, Gillin M, Prado K et al (2010) Energy dependence and dose response of Gafchromic EBT2 film over a wide range of photon, electron, and proton beam energies. Med Phys 37(5):1942–1947 Hartmann B, Matišíková M, Jäkel O (2010) Technical note: homogeneity of Gafchromic EBT2 film. Med Phys 37(4):1753–1756 Kairn T, Aland T, Kenny J (2010) Local heterogeneities in early batches of EBT2 film: a suggested solution. Phys Med Biol 55:L37–L42 Mayer RR, Ma F, Chen Y, Miller RI, Belard A, McDonough J et al (2012) Enhanced dosimetry procedures and assessment for EBT2 radiochromic film. Med Phys 39(4):2147–2156 Devic S, Wang Y, Tomic N, Podgorsak EB (2006) Sensitivity of linear CCD array based film scanners used for film dosimetry. Med Phys 33:3393–3396 Smith L, Hill R, Nakano M, Kin J, Kuncic Z (2011) The measurement of backscatter factors of kilovoltage X-ray beams using Gafchromic™ EBT2 film. Australas Phys Eng Sci Med 34:261–266 Lynch BD, Kozelka J, Ranade MK, Li JG, Simon WE, Dempsey JF (2006) Important considerations for radiochromic film dosimetry with flatbed CCD scanners and EBT Gafchromic film. Med Phys 33(12):4551–4556 Park S, Kang S, Cheong K, Hwang T, Kim H, Han T et al (2012) Variations in dose distribution and optical properties of Gafchromic™ EBT2 film according to scanning mode. Med Phys 39(5):2524–2535 Devic S, Aldelaijan S, Mohammed H et al (2010) Absorption spectra time evolution of EBT2 model GAFCHROMIC™ film. Med Phys 37(5):2207–2214 IAEA (2001) Absorbed dose determination in external beam radiotherapy: an international code of practice for dosimetry based on standards of absorbed dose to water. Technical report series, vol 398. IAEA, Vienna Almond P, Biggs P, Coursey B, Hanson W, Huq M, Nath R, Rogers D (1999) Protocol for Clinical reference dosimetry of high-energy photon and electron beams: recommendations of AAPM radiation therapy committee task group 51. Med Phys 26(11):1847–1870 Fanning DW (1999) IDL Programming techniques. Fort Collins, USA Strydom WJ (1991) Central axis depth dose curve for electron beams. Med Phys 18(6):1254–1255 Desroches J, Bouchard H, Lacroix F (2010) Technical note: potential errors in optical density measurements due to scanning side in EBT and EBT2 Gafchromic film dosimetry. Med Phys 37(4):1565–1570 Khan FM, Doppke KP, Hogstrom KR et al (1991) Clinical Electron beam dosimetry: recommendations of AAPM radiation therapy committee task group 25. Med Phys 18(1):73–109 Niroomand-Rad A, Blackwell CR, Coursey BM et al (1998) Radiochromic film dosimetry: recommendations of AAPM radiation therapy committee task group 55. Med Phys 25(11):2093–2115 Niroomand-Rad A, Gillin MT, Kline RW, Grimm DF (1986) Film dosimetry of small electron beams for routine radiotherapy planning. Med Phys 13(3):416–421 Gerbi BJ, Dimitroyannis DA (2003) The response of Kodak EDR2 film in high-energy electron beams. Med Phys 30(10):2703–2705 Arjomandy B, Tailor R, Zhao L, Devic S (2012) EBT2 film as a depth-dose measurement tool for radiotherapy beams over a wide range of energies and modalities. Med Phys 39(2):912–921 Kairn T, Hardcastle N, Kenny J, Meldrum R, Tomé WA, Aland T (2011) EBT2 radiochromic film for quality assurance of complex IMRT treatments of the prostate: micro-collimated IMRT, RapidArc and TomoTherapy. Australas Phys Eng Sci Med 34:333–343