Evaluation of the dosimetric properties of a diode detector for small field proton radiosurgery

Journal of Applied Clinical Medical Physics - Tập 16 Số 6 - Trang 51-64 - 2015
Grant McAuley1, Anthony V. Teran2, Jerry D. Slater2, James M. Slater2, Andrew Wroe2
1Department of Radiation Medicine, Loma Linda University, Loma Linda, CA
2Department of Radiation Medicine, Loma Linda University Medical Center, Loma Linda, CA USA

Tóm tắt

The small fields and sharp gradients typically encountered in proton radiosurgery require high spatial resolution dosimetric measurements, especially below 1–2 cm diameters. Radiochromic film provides high resolution, but requires postprocessing and special handling. Promising alternatives are diode detectors with small sensitive volumes (SV) that are capable of high resolution and real‐time dose acquisition. In this study we evaluated the PTW PR60020 proton dosimetry diode using radiation fields and beam energies relevant to radiosurgery applications. Energies of 127 and 157 MeV (9.7 to 15 cm range) and initial diameters of 8, 10, 12, and 20 mm were delivered using single‐stage scattering and four modulations (0, 15, 30, and 60 mm) to a water tank in our treatment room. Depth dose and beam profile data were compared with PTW Markus N23343 ionization chamber, EBT2 Gafchromic film, and Monte Carlo simulations. Transverse dose profiles were measured using the diode in "edge‐on" orientation or EBT2 film. Diode response was linear with respect to dose, uniform with dose rate, and showed an orientation‐dependent (i.e., beam parallel to, or perpendicular to, detector axis) response of less than 1%. Diode vs. Markus depth‐dose profiles, as well as Markus relative dose ratio vs. simulated dose‐weighted average lineal energy plots, suggest that any LET‐dependent diode response is negligible from particle entrance up to the very distal portion of the SOBP for the energies tested. Finally, while not possible with the ionization chamber due to partial volume effects, accurate diode depth‐dose measurements of 8, 10, and 12 mm diameter beams were obtained compared to Monte Carlo simulations. Because of the small SV that allows measurements without partial volume effects and the capability of submillimeter resolution (in edge‐on orientation) that is crucial for small fields and high‐dose gradients (e.g., penumbra, distal edge), as well as negligible LET dependence over nearly the full the SOBP, the PTW proton diode proved to be a useful high‐resolution, real‐time metrology device for small proton field radiation measurements such as would be encountered in radiosurgery applications.PACS numbers: 87.56.‐v, 87.56.jf, 87.56.Fc

Từ khóa


Tài liệu tham khảo

10.1118/1.598407

10.1088/0031-9155/52/1/015

10.1118/1.2065467

10.1088/0031-9155/55/2/006

10.1118/1.4737890

10.1016/j.radmeas.2011.07.041

10.1093/oxfordjournals.rpd.a006026

10.1088/0031-9155/45/10/320

10.1088/0031-9155/11/3/302

10.2307/3583700

10.1088/0031-9155/45/9/310

10.1007/s00411-010-0299-8

10.1088/0031-9155/44/11/308

10.1118/1.596617

10.1088/0031-9155/32/9/004

10.2307/3575020

Cuttone G, 1999, First dosimetry intercomparison results for the CATANA project, Phys Med., 15, 121

10.1177/15330346070060S407

10.1088/0031-9155/55/21/R01

10.1118/1.3576105

10.1118/1.4754797

Agostinelli S, 2003, GEANT4‐a simulation toolkit, Nucl Instrum Meth Phys Res A., 506, 250, 10.1016/S0168-9002(03)01368-8

10.1109/TNS.2006.869826

Physics Reference Manual version 9.6.0. 2012. Available from:http://geant4.web.cern.ch/geant4/UserDocumentation/UsersGuides/PhysicsReferenceManual/BackupVersions/V9.6/fo/PhysicsReferenceManual.pdf

10.1118/1.4793262

10.1177/1533034614557409

1995, Rossum Gv. Python tutorial: Technical Report CS‐R9526

10.1007/978-3-642-85184-1

10.1088/0031-9155/59/20/R303

10.1118/1.595963

10.1118/1.2989089

10.1118/1.2936334

Paelinck L, 2007, Radiochromic EBT film dosimetry: effect of film orientation and batch on the lateral correction of the scanner, Radiother Oncol., 84, S194

10.1118/1.4736800

10.1118/1.3566067

10.1118/1.3549759

10.1118/1.3264176