Evaluation of an Oscillometric Blood Pressure Monitor on Anesthetized Cats and the Effect of Cuff Placement and Fur on Accuracy

Veterinary Surgery - Tập 26 Số 4 - Trang 347-353 - 1997
Keith Branson1,2, Colette Wagner‐Mann1,2, F. A. Mann1,2
1Department of Veterinary Medicine and Surgery, College of Veterinary Medicine
2the Division of Cardiothoracic Surgery, School of Medicine, University of Missouri, Columbia, MO.

Tóm tắt

Objectives— (1) To determine the usefulness of one specific oscillometric monitor for making indirect measurements of arterial pressure in cats. (2) To determine the difference between two specific cuff placement sites. (3) To determine if clipping the hair beneath the cuff has an affect on the accuracy of oscillometric blood pressure determination.Study Design— Prospective study comparing the accuracy of the Datascope Passport (Data‐scope Corp, Paramus, NJ) with concurrent invasive measurements.Animals— Six anesthetized domestic felines weighing 4.5 to 5 kg.Methods— The direct arterial pressure was measured using a cannula placed in the right common carotid artery. Oscillometric cuffs of appropriate size were placed on both thoracic limbs distal to the elbow and both pelvic limbs distal to the stifle. The hair in the areas of cuff placement on the right limbs was clipped circumferentially. Measurements of systolic, diastolic, and mean arterial pressure (MAP) were taken for each site during normotension, hypotension, and hypertension. Comparisons between indirect and direct measurements were made using a parametric analysis of method comparison.Results— No significant differences were noted when die clipped limbs were compared with the corresponding limbs which were left undipped (P>.378) or when the thoracic limb measurements were compared widi those of the pelvic limb (P>.088). There were significant differences (P<.002) between the two pressure measurement methods for the systolic, diastolic, and mean arterial pressures over all three pressure ranges.Conclusions— The Datascope Passport did not accurately estimate the invasively measured arterial pressure.Clinical Relevance— Use of noninvasive blood pressure monitoring equipment is increasing in use in veterinary medicine, and the accuracy of one specific monitor is reported.

Từ khóa


Tài liệu tham khảo

Gorback MS, 1988, Complications in Critical Care Medicine, 296

Hunter JS, 1990, Adaptation of human oscillometric blood pressure monitors for use in dogs, Am J Vet Res, 51, 1439, 10.2460/ajvr.1990.51.09.1439

Coulter DB, 1984, Blood pressures obtained by indirect measurement in conscious dogs, J Am Vet Med Assoc, 184, 1375

Hamlin RL, 1982, Noninvasive measurement of systemic arterial pressure in dogs by automatic sphygmomanometry, Am J Vet Res, 43, 1271

Sawyer DC, 1991, Comparison of direct and indirect blood pressure measurement in anesthetized dogs, Lab Anim Sci, 41, 134

10.1016/0034-5288(94)90116-3

10.1111/j.1939-1676.1995.tb03301.x

SawyerDC StrilerE. BohartG.:Indirect blood pressure measurements in dogs cats and horses: Correlation with direct arterial pressures and site of measurement. Proceedings of 5th International Congress of Vet Anesthesia 209 1994(abstr).

Hansen B., 1995, Kirk's Current Veterinary Therapy XII, 110

10.1016/S0140-6736(86)90837-8

10.2307/2987937

Bruner JM, 1981, Comparison of direct and indirect methods of measuring arterial blood pressure, part H, Med Instrum, 15, 97

Bruner JM, 1981, Comparison of direct and indirect methods of measuring arterial blood pressure, Med Instrum, 15, 11

Bruner JM, 1981, Comparison of direct and indirect methods of measuring arterial blood pressure, part HI, Med Instrum, 15, 182

10.1007/BF01617900