Emergency ventilation with the Ventrain® through an airway exchange catheter in a porcine model of complete upper airway obstruction

Canadian Journal of Anaesthesia - Tập 64 - Trang 37-44 - 2016
Michiel W. P. de Wolf1, Reiner Gottschall2, Niels P. Preussler2, Markus Paxian2, Dietmar Enk1
1Department of Anesthesiology and Pain Therapy, Maastricht University Medical Center, Maastricht, The Netherlands
2Department of Anesthesiology and Intensive Care Medicine, Jena University Hospital, Jena, Germany

Tóm tắt

During difficult airway management, oxygen insufflation through airway-exchange and intubating catheters (AEC/IC) can lead to life-threatening hyperinflation. Ventrain® was originally designed to facilitate emergency ventilation using active expiration through short, small-bore cannulas. Herein, we studied its efficacy (oxygenation and ventilation) and safety (avoidance of hyperinflation) in a long, small-bore AEC. In six anesthetized pigs, the upper airway was obstructed, except for a 100 cm long, 3 mm internal diameter AEC. After apneic desaturation to a peripheral oxygen saturation (SpO2) of < 70%, ventilation through the AEC was started with Ventrain at an oxygen flow of 15 L·min−1, a frequency of 30 breaths·min−1, and an inspiration/expiration ratio of approximately 1:1. It was continued for ten minutes. Within one minute, severe hypoxia was reversed from a median [interquartile range] arterial saturation (SaO2) of 48 [34-56] % before initiation of Ventrain ventilation to 100 [99-100] % afterward (median difference 54%; 95% confidence interval [CI] 44 to 67; P = 0.028). In addition, hypercarbia was reversed from PaCO2 of 59 [53-61] mmHg to 40 [38-42] mmHg (median difference of −18 mmHg; 95% CI −21 to −15; P = 0.028). After ten minutes of Ventrain use, peak inspiratory and end-expiratory pressures were lower than during baseline pressure-controlled ventilation (8 [7-9] mmHg vs 12 [10-14] mmHg and −2 [−3 to +1] mmHg vs 4 [2 to 4] mmHg, respectively; P = 0.027 for both). No hemodynamic deterioration occurred. Ventrain provides rapid reoxygenation and effective ventilation through a small-bore AEC in pigs with an obstructed airway. In clinical emergency situations of obstructed airways, this device may be able to overcome problems of unintentional hyperinflation and high intrapulmonary pressures when ventilating through long, small-bore catheters and could therefore minimize the risks of barotrauma and hemodynamic instability.

Tài liệu tham khảo

Popat M, Mitchell V, Dravid R, Patel A, Swampillai C, Higgs A. Difficult Airway Society Guidelines for the management of tracheal extubation. Anaesthesia 2012; 67: 318-40. Cavallone LF, Vannucci A. Extubation of the difficult airway and extubation failure. Anesth Analg 2013; 116: 368-83. Dworkin R, Benumof JL, Benumof R, Karagianes TG. The effective tracheal diameter that causes air trapping during jet ventilation. J Cardiothorac Anesth 1990; 4: 731-6. Cooper RM. The use of an endotracheal ventilation catheter in the management of difficult extubations. Can J Anesth 1996; 43: 90-3. Duggan LV, Law JA, Murphy MF. Brief review: Supplementing oxygen through an airway exchange catheter: efficacy, complications, and recommendations. Can J Anesth 2011; 58: 560-8. Hamaekers AE, Borg PA, Enk D. Ventrain: an ejector ventilator for emergency use. Br J Anaesth 2012; 108: 1017-21. Enk D. Patent US 8950400 B2. Gas flow reversing element, United States Patent and Trademark Office, February 10, 2015. Available from URL: https://www.google.si/patents/US8950400 (accessed September 2016). Lang SA. Emergency airway management: What are the roles for surgical cricothyroidotomy and the Ventrain® device? Can J Anesth 2016; 63: 997-8. Berry M, Tzeng Y, Marsland C. Percutaneous transtracheal ventilation in an obstructed airway model in post-apnoeic sheep. Br J Anaesth 2014; 113: 1039-45. Hamaekers AE, van der Beek T, Theunissen M, Enk D. Rescue ventilation through a small-bore transtracheal cannula in severe hypoxic pigs using expiratory ventilation assistance. Anesth Analg 2015; 120: 890-4. Paxian M, Preussler NP, Reinz T, Schlueter A, Gottschall R. Transtracheal ventilation with a novel ejector-based device (Ventrain) in open, partly obstructed, or totally closed upper airways in pigs. Br J Anaesth 2015; 115: 308-16. Dias EM, Hamaekers A, Borg PA, Enk D. Adequate minute volume ventilation through a 100 cm long, 3 mm inner diameter airway exchange catheter by expiratory ventilation assistance (EVA): 19AP5-6. Eur J Anaesthesiol 2012; 29: 239-40. McLean S, Lanam CR, Benedict W, Kirkpatrick N, Kheterpal S, Ramachandran SK. Airway exchange failure and complications with the use of the Cook Airway Exchange Catheter®: a single center cohort study of 1177 patients. Anesth Analg 2013; 117: 1325-7. Mort TC. Continuous airway access for the difficult extubation: the efficacy of the airway exchange catheter. Anesth Analg 2007; 105: 1357-62. Harris K, Chalhoub M, Maroun R, Elsayegh D. Endotracheal tube exchangers: should we look for safer alternatives? Heart Lung 2012; 41: 67-9. Willemsen MG, Noppens R, Mulder AL, Enk D. Ventilation with the Ventrain through a small lumen catheter in the failed paediatric airway: two case reports. Br J Anaesth 2014; 112: 946-7. Segal N, Parquette B, Ziehr J, Yannopoulos D, Lindstrom D. Intrathoracic pressure regulation during cardiopulmonary resuscitation: a feasibility case-series. Resuscitation 2013; 84: 450-3. Buczkowski PW, Fombon FN, Lin ES, Russell WC, Thompson JP. Air entrainment during high-frequency jet ventilation in a model of upper tracheal stenosis. Br J Anaesth 2007; 99: 891-7.