Electrogram voltage and pacing threshold before ablation, measured by mini-electrodes, predict parameters indicative of transmural lesions in the human atrium

Springer Science and Business Media LLC - Tập 57 - Trang 443-452 - 2019
Carla Lázaro1, Teresa Barrio-López1, Eduardo Castellanos1, Mercedes Ortiz1, Martín Arceluz1, Jesús Almendral1
1Electrophysiology Laboratory and Arrhythmia Unit, Hospital Madrid Monteprincipe, Grupo HM Hospitales, University CEU-San Pablo, Madrid, Spain

Tóm tắt

An important attenuation of the atrial signal recorded with mini-electrodes (ME) embedded in an 8-mm tip was associated with a transmural radiofrequency lesion. Our aim was to assess if parameters obtained from ME or conventional bipoles before applications predict successful atrial lesions. We prospectively included 33 consecutive patients undergoing cavotricuspid isthmus (CTI) ablation. Electrogram voltages and pacing thresholds were measured with ME and conventional bipoles before and after radiofrequency (RF) applications. The time before the loss of capture during applications was recorded. Lesions were considered successful, in accordance with preclinical data, if ME voltage decreased > 54%. Of 207 applications, 107 could be analyzed. During applications, voltages decreased more in the ME than in the conventional bipoles (66.8 ± 26.1% vs 37.5 ± 42.5%, P = 0.001). Likewise, pacing threshold increased significantly more using the ME (86.3 ± 22.9% ME, 52.6 ± 35.6% conventional, P = 0.001). ME pre-ablation voltages were significantly higher and pacing thresholds significantly lower in successful lesions (voltage 0.88 ± 0.71 vs 0.26 ± 0.18 mV, P = 0.0001; threshold 1.6 ± 1.7 vs 2.8 ± 3.0, P = 0.04). Neither of these parameters with conventional bipoles nor time to loss of capture showed differences. A ME voltage > 0.33 mV and a pacing threshold < 1.5 mA predicted a successful lesion with 0.78 and 0.6 sensitivity and 0.78 and 0.59 specificity. Certain pre-ablation parameters derived from ME such as electrogram voltage and pacing threshold differ from those obtained by a conventional configuration and can predict a successful atrial lesion.

Tài liệu tham khảo

Avitall B, Helms RW, Koblish JB, Sieben W, Kotov AV, Gupta GN. The creation of linear contiguous lesions in the atria with an expandable loop catheter. J Am Coll Cardiol. 1999;33(4):972–84. Sapp JL, Soejima K, Cooper JM, Epstein LM, Stevenson WG. Ablation lesion size correlates with pacing threshold: a physiological basis for use of pacing to assess ablation lesions. Pacing Clin Electrophysiol. 2004;27(7):933–7. Avitall B, Horbal P, Vance D, Koblish J, Kalinski A. Maximal electrogram attenuation recorded from mini electrodes embedded on 4.5-mm irrigated and 8-mm nonirrigated catheters signifies lesion maturation. J Cardiovasc Electrophysiol. 2015;26(2):192–202. Gepstein L, Hayam G, Shpun S, Cohen D, Ben-Haim SA. Atrial linear ablations in pigs. Chronic effects on atrial electrophysiology and pathology. Circulation. 1999;100(4):419–26. Sanchez JE, Kay GN, Benser ME, Hall JA, Walcott GP, Smith WM, et al. Identification of transmural necrosis along a linear catheter ablation lesion during atrial fibrillation and sinus rhythm. J Interv Card Electrophysiol. 2003;8(1):9–17. Jumrussirikul P, Atiga WL, Lardo AC, Berger RD, Halperin H, Hutchins GM, et al. Prospective comparison of lesions created using a multipolar microcatheter ablation system with those created using a pullback approach with standard radiofrequency ablation in the canine atrium. Pacing Clin Electrophysiol. 2000;23(2):203–13. Eitel C, Hindricks G, Sommer P, Gaspar T, Kircher S, Wetzel U, et al. Circumferential pulmonary vein isolation and linear left atrial ablation as a single-catheter technique to achieve bidirectional conduction block: the pace-and-ablate approach. Heart Rhythm. 2010;7(2):157–64. Avitall B, Mughal K, Hare J, Helms R, Krum D. The effects of electrode-tissue contact on radiofrequency lesion generation. Pacing Clin Electrophysiol. 1997;20(12 Pt 1):2899–910. Lloyd MS, Hoskins MH, Shah AD, Langberg JJ. Electrogram characteristics of ablated and non-ablated myocardium in humans: a comparison of miniaturized embedded electrodes and conventional ablation electrodes. J Cardiovasc Electrophysiol. 2016;27(7):820–4. Yokoyama K, Nakagawa H, Shah DC, Lambert H, Leo G, Aeby N, et al. Novel contact force sensor incorporated in irrigated radiofrequency ablation catheter predicts lesion size and incidence of steam pop and thrombus. Circ Arrhythm Electrophysiol. 2008;1(5):354–62. Barrio-Lopez T, Ortiz M, Castellanos E, Lazaro C, Salas J, Madero S, et al. Relationship between left atrium catheter contact force and pacing threshold. J Interv Card Electrophysiol. 2017;49(2):147–55. Palzhjsiapkja B. Novel ablation catheter technology that improves mapping resolution and monitoring of lesion maturation. J Innovations Card Rhythm Manage. 2012;3:599–609. Levy MR, Merchant FM, Langberg JJ, Delurgio DB. Use of microelectrode near-field signals to determine catheter contact. J Arrhythm. 2018;34(1):23–9. Sulkin MS, Laughner JI, Hilbert S, Kapa S, Kosiuk J, Younan P, et al. Novel measure of local impedance predicts catheter-tissue contact and lesion formation. Circ Arrhythm Electrophysiol. 2018;11(4):e005831. Khoury DS, Rao L, Ding C, Sun H, Youker KA, Panescu D, et al. Localizing and quantifying ablation lesions in the left ventricle by myocardial contrast echocardiography. J Cardiovasc Electrophysiol. 2004;15(9):1078–87. Iwasawa J, Miyazaki S, Takagi T, Taniguchi H, Nakamura H, Hachiya H, et al. Cavotricuspid isthmus ablation using a catheter equipped with mini electrodes on the 8 mm tip: a prospective comparison with an 8 mm dumbbell-shaped tip catheter and 8 mm tip cryothermal catheter. Europace. 2016;18(6):868–72.