Effects of pre-contoured and in situ contoured rods on the mechanical strength and durability of posterior cervical instrumentation: a finite-element analysis and scanning electron microscopy investigation

Spine Deformity - Tập 8 - Trang 569-576 - 2020
Kee D. Kim1, Ripul Panchal2, Mark Moldavsky3, Wenhai Wang3, Brandon S. Bucklen3
1UC Davis Medical Center, UC Davis School of Medicine, Sacramento, USA
2American Neurospine Institute, Plano, USA
3Musculoskeletal Education and Research Center (MERC), A Division of Globus Medical, Inc, Audubon, USA

Tóm tắt

Finite-element analysis. Intraoperative contouring of rods is a common procedure for spine surgeons to match the native curvature of the spine, but it may lead to premature weakening of the rod. This study investigated the effect of different bending methods on rod fatigue performance. Rod failure in the cervical spine is of clinical concern, particularly when spanning the cervicothoracic region and when considering corrective osteotomies for deformity correction and global spinal alignment. Finite-element models were developed to simulate rod bending (3.5 mm D, 40 mm L) to achieve a 23° angle with 3 different bending methods: French single, multiple bending, and in situ bending. Simulations were conducted in 4 steps: rod bending, rod spring back, residual stress relaxation, and F1717 mechanical test simulation. French single bending resulted in the highest residual stress concentrations for both titanium (TiAlV) and cobalt chrome (CoCr) at 783 MPa and 507 MPa, respectively. During F1717 test simulation, the French single bent rod had its highest tensile stress in the middle, with 917 MPa and 623 MPa, respectively, for TiAlV and CoCr, compared to in situ (580 MPa and 586 MPa for TiAlV and CoCr) and the French multiple bent rod (765 MPa and 619 MPa for TiAlV and CoCr). The computational model found that CoCr rods made the construct least prone to deformation. French single bend with TiAlV rods put the construct at highest risk of failure. CoCr rods led to minimal physical changes in microstructure while showing evidence of flattening.

Tài liệu tham khảo

Asher M, Lai SM, Burton D et al (2004) Safety and efficacy of Isola instrumentation and arthrodesis for adolescent idiopathic scoliosis: two- to 12-year follow-up. Spine (Phila PA 1976) 29:2013–2023 Bridwell KH, Hanson DS, Rhee JM et al (2002) Correction of thoracic adolescent idiopathic scoliosis with segmental hooks, rods, and Wisconsin wires posteriorly: it’s bad and obsolete, correct? Spine (Phila PA 1976) 27:2059–2066 De Maio F, Dolan LA, De Luna V et al (2007) Posterior spine fusion with Moss-Miami instrumentation for adolescent idiopathic scoliosis: radiographic, clinical and patient-centered outcomes. Iowa Orthop J 27:28–39 Gorensek M, Kosak R, Travnik L et al (2013) Posterior instrumentation, anterior column reconstruction with single posterior approach for treatment of pyogenic osteomyelitis of thoracic and lumbar spine. Eur Spine J 22:633–641 Mikles MR, Stchur RP, Graziano GP (2004) Posterior instrumentation for thoracolumbar fractures. J Am Acad Orthop Surg 12:424–435 Remes V, Helenius I, Schlenzka D et al (2004) Cotrel-Dubousset (CD) or Universal Spine System (USS) instrumentation in adolescent idiopathic scoliosis (AIS): comparison of midterm clinical, functional, and radiologic outcomes. Spine (Phila PA 1976) 29:2024–2030 Resnick DK, Choudhri TF, Dailey AT et al (2005) Guidelines for the performance of fusion procedures for degenerative disease of the lumbar spine. Part 12: pedicle screw fixation as an adjunct to posterolateral fusion for low-back pain. J Neurosurg Spine 2:700–706 Rutherford EE, Tarplett LJ, Davies EM et al (2007) Lumbar spine fusion and stabilization: hardware, techniques, and imaging appearances. Radiographics 27:1737–1749 Vialle R, Mary P, Harding I et al (2008) Surgical treatment of severe thoracic scoliosis in skeletally mature patients. Orthopedics 31:218 Chapman MW (2001) Chapman’s orthopaedic surgery, 3rd edn. Lippincott Williams & Wilkins, Philadelphia Niinomi M (2008) Metallic biomaterials. J Artif Organs 11:105–110 Smith JS, Shaffrey CI, Ames CP et al (2012) Assessment of symptomatic rod fracture after posterior instrumented fusion for adult spinal deformity. Neurosurgery 71:862–867 Ames CP, Smith JS, Scheer JK et al (2013) A standardized nomenclature for cervical spine soft-tissue release and osteotomy for deformity correction: clinical article. J Neurosurg Spine 19:269–278 Kim HJ, Piyaskulkaew C, Riew KD (2015) Comparison of Smith-Petersen osteotomy versus pedicle subtraction osteotomy versus anterior-posterior osteotomy types for the correction of cervical spine deformities. Spine (Phila PA 1976) 40:143–146 Yamanaka K, Mori M, Yamazaki K et al (2015) Analysis of the fracture mechanism of Ti–6Al–4V alloy rods that failed clinically after spinal instrumentation surgery. Spine (Phila PA 1976) 40:E767–E773 Dick JC, Bourgeault CA (2001) Notch sensitivity of titanium alloy, commercially pure titanium, and stainless steel spinal implants. Spine (Phila Pa 1976) 26:1668–1672 Ashman RB, Birch JG, Bone LB et al (1988) Mechanical testing of spinal instrumentation. Clin Orthop Relat Res 227:113–125 ASTM International (2011) ASTM F1717-14: Standard test methods for spinal implant constructs in a vertebrectomy model. ASTM International, West Conshohocken La Barbera L, Galbusera F, Villa T et al (2014) ASTM F1717 standard for the preclinical evaluation of posterior spinal fixators: can we improve it? Proc Inst Mech Eng H 228:1014–1026 Villa T, La Barbera L, Galbusera F (2014) Comparative analysis of international standards for the fatigue testing of posterior spinal fixation systems. Spine J 14:695–704 Lindsey C, Deviren V, Xu Z et al (2006) The effects of rod contouring on spinal construct fatigue strength. Spine (Phila PA 1976) 31:1680–1687 Farahmand B, Saff C, Xie D, Abdi F (2007) Estimation of fatigue and fracture allowables for metallic materials under cyclic loading. AIAA-2007-2381. American Institute of Aeronautics and Astronautics, Honolulu Nicholas T (1980) Dynamic tensile testing of structural materials using a split Hopkinson bar apparatus. Air Force Wright Aeronautical Labs, Wright-Patterson AFB, Ohio Winegar CD, Lawrence JP, Friel BC et al (2010) A systematic review of occipital cervical fusion: techniques and outcomes. J Neurosurg Spine 13:5–16 Deen HG, Nottmeier EW, Reimer R (2006) Early complications of posterior rod-screw fixation of the cervical and upper thoracic spine. Neurosurgery 59:1062–1068 Nguyen TQ, Buckley JM, Ames C et al (2011) The fatigue life of contoured cobalt chrome posterior spinal fusion rods. Proc Inst Mech Eng H 225:194–198 Tang JA, Leasure JM, Smith JS et al (2013) Effect of severity of rod contour on posterior rod failure in the setting of lumbar pedicle subtraction osteotomy (PSO): a biomechanical study. Neurosurgery 72:276–282 (discussion 283) Hosseini S (2012) Fatigue of Ti–6Al–4V. InTech Open Access Publisher, Rijeka Stanford RE, Loefler AH, Stanford PM (2004) Multiaxial pedicle screw designs: static and dynamic mechanical testing. Spine (Phila PA 1976) 29(4):367–375 Slivka MA, Fan YK, Eck JC (2013) The effect of contouring on fatigue strength of spinal rods: is it okay to re-bend and which materials are best? Spine Deform 1:395–400