EFFECTS OF BALLOON LENGTH AND COMPLIANCE ON VASCULAR STENT EXPANSION

International Journal of Applied Mechanics - Tập 02 Số 03 - Trang 681-697 - 2010
Fangsen Cui1, H. P. LEE2, C. Lu3, Ping Chai4
1Institute of High Performance Computing, 1 Fusionopolis Way, #16-16 Connexis, Singapore 138632, Singapore
2Department of Mechanical Engineering, National University of Singapore, 9 Engineering Drive 1, Singapore 119260, Singapore
3King Abdullah University of Science and Technology, Saudi Arabia
4Department of Cardiology National University Hospital, 5 Lower Kent Ridge Road, Singapore 119074, Singapore

Tóm tắt

Vascular stents are used to dilate arteries that are narrowed or clogged by plaque. However, in-stent restenosis is still one of the major causes of the clinical failure. It is believed that vessel trauma imposed during stent deployment is closely correlated to restenosis. Clinical observations show that the longitudinal and axial geographic miss in pre/post-dilation are responsible for the vessel trauma. The interactions between the stent-strut and the artery are difficult to measure in vivo or clinically and reported results are very limited. A numerical approach that leverages on computing power can provide new insights into the stent implantation process. In this study, the effects of balloon length and compliance that play important roles during stent expansion were investigated. Areas in the vessels with high stress concentrations were identified as these were weaknesses that might have a high possibility of vascular injury. Two different types of numerical models were constructed: a simplified model that considered only the balloon and stent and a more comprehensive model that consisted of the balloon, stent, plaque, and artery. Virtual stent implantation trials were simulated and the phenomena of stent recoil, dogboning and foreshortening were observed and examined. It was found that balloons which were slightly longer than the stent and less compliance would be more likely to eliminate dogboning. Furthermore, a new parameter, namely the Ectropion angle, was introduced to describe the turning effect of the stent end in situations when dogboning could not adequately characterize this phenomenon. The present study could provide guidance for the placement of stents by clinical practitioners.

Từ khóa


Tài liệu tham khảo

10.1080/10255840108908007

10.1016/S0924-0136(01)01127-X

Chua S. N. D., J. Mater. Process. Technol., 143, 591

Chua S. N. D., J. Mater. Process. Technol., 156, 1772

10.1161/01.CIR.30.5.654

10.1016/S0021-9290(00)00098-1

Francesco S., Cardio Update, 14

Freed M., 1997, The New Manual of Interventional Cardiology

10.1016/j.jbiomech.2008.01.027

Gijsen F. J., BioMed. Eng. OnLine, 6, 7

10.1615/CritRevBiomedEng.v28.i12.90

10.1136/heart.88.4.401

10.3949/ccjm.71.10.815

Hibbeler R. C., 2000, Mechanics of Materials

10.1016/j.jmbbm.2007.07.002

10.1007/s10439-008-9606-9

10.1016/j.jbiomech.2004.07.022

10.1016/j.ijcard.2004.12.033

10.1016/0021-9290(94)90209-7

10.1016/S0735-1097(01)01742-9

10.1016/S0021-9290(02)00033-7

10.1161/01.RES.84.4.378

10.1016/0735-1097(92)90476-4

10.1161/01.CIR.0000069331.67148.2F

10.1161/01.CIR.102.4.399

10.1142/S1758825109000095

10.1111/j.1538-7836.2004.01092.x

Walke W., J. Mater. Process. Technol., 165, 1263

10.1016/j.jbiomech.2004.11.003

Wong A., Ann. Acad. Med. Singapore, 33, 423, 10.47102/annals-acadmedsg.V33N4p423

10.1093/oxfordjournals.bmb.a072448

10.1016/j.jbiomech.2006.11.009

10.1007/s11517-009-0432-5