Determination of manganese using the method of electrochemical enrichment
Tóm tắt
Manganese was determined by differential pulse voltammetry in pre‐enriched solution. The metal was first deposited on the surface of a hanging mercury drop electrode by electrolysis, and after a short oxidation interval, the reformed manganese(II) ions were determined by differential pulse cathodic polarization scan. The detection limit (with 120 seconds deposition) was 2 μg L−1. The only interfering element was cobalt (when present at tenfold concentration excess). Compared with conventional stripping voltammetry the described method has a lower detection limit value, and the analysis is performed during a shorter time interval. The proposed method was applied for the determination of manganese in natural waters.