Choosing the best supraglottic airway for ophthalmic general anaesthesia: a manikin study
Tóm tắt
General anaesthesia is sometimes favoured over regional anaesthesia in ophthalmic surgery. The use of supraglottic airway (SGA) or laryngeal mask airway (LMA) as the primary airway device is increasing due to numerous advantages over tracheal intubation. Compared with 1st generation SGAs, 2nd generation SGAs have an added benefit of isolating the airway from the alimentary tract. However, the vertical profile of SGAs may encroach into the surgical field and hence interfere with surgery. We investigated the vertical projections of 1st generation SGAs (LMA Classic, Ambu AuraFlex) and commonly used 2nd generation SGAs in our institution (LMA ProSeal, LMA Supreme, LMA Protector, Ambu AuraGain and I-gel) in a manikin model. Each device was connected to a corrugated catheter mount or angled connector following insertion as per usual clinical practice in our institutions. Vertical projections of all devices were measured from the chin using a centimetre ruler. Securing of airway device to the chin with an adhesive tape was possible for the LMA Classic and Ambu AuraFlex with straight corrugated connector, whereas the stiffer 2nd generations SGAs required the addition of an angled connector or straight corrugated tubing to direct the airway tube caudally, away from the surgical field. The LMA ProSeal had the lowest vertical projection amongst the 2nd generation SGAs and may be the suitable choice for ophthalmic surgery. We also describe a novel technique of utilising a 1st generation SGA with placement of an orogastric tube, although with some reservations. This study has several limitations and transferability of our findings into clinical practice is questionable as the use of a manikin may not fully imitate the real condition of the patient. Our study is the first study comparing vertical projected height of different SGAs in manikin, but future studies should investigate the use of SGA in the clinical setting during ophthalmic surgery.
Tài liệu tham khảo
Kumar CM, Eke T, Dodds C, Deane JS, El-Hindy N, Johnston RL, et al. Local anaesthesia for ophthalmic surgery—new guidelines from the Royal College of Anaesthetists and the Royal College of Ophthalmologists. Eye (Lond). 2012;26:897–8.
Kelly DJ, Farrell SM. Physiology and role of intraocular pressure in contemporary anesthesia. Anesth Analg. 2018;126:1551–62.
Brimacombe J. The advantages of the LMA over the tracheal tube or facemask: a meta-analysis. Can J Anaesth. 1995;42:1017–23.
Yu SH, Beirne OR. Laryngeal mask airways have a lower risk of airway complications compared with endotracheal intubation: a systematic review. J Oral Maxillofac Surg. 2010;68:2359–76.
Holloway KB. Control of the eye during general anaesthesia for intraocular surgery. Br J Anaesth. 1980;52:671–9.
Joshi GP, Inagaki Y, White PF, Taylor-Kennedy L, Wat LI, Gevirtz C, et al. Use of the laryngeal mask airway as an alternative to the tracheal tube during ambulatory anesthesia. Anesth Analg. 1997;85:573–7.
Qamarul Hoda M, Samad K, Ullah H. ProSeal versus Classic laryngeal mask airway (LMA) for positive pressure ventilation in adults undergoing elective surgery. Cochrane Database Syst Rev. 2017;7:CD009026.
Timmermann A, Bergner UA, Russo SG. Laryngeal mask airway indications: new frontiers for second-generation supraglottic airways. Curr Opin Anaesthesiol. 2015;28:717–26.
Frerk C, Mitchell VS, McNarry AF, Mendonca C, Bhagrath R, Patel A, et al. Difficult Airway Society 2015 guidelines for management of unanticipated difficult intubation in adults. Br J Anaesth. 2015;115:827–48.