Biaxial Mechanical Properties of the Natural and Glutaraldehyde Treated Aortic Valve Cusp—Part I: Experimental Results

Journal of Biomechanical Engineering - Tập 122 Số 1 - Trang 23-30 - 2000
Kristen L. Billiar1, Michael S. Sacks1
1Department of Biomedical Engineering, University of Miami, Coral Gables, FL 33124

Tóm tắt

To date, there are no constitutive models for either the natural or bioprosthetic aortic valve (AV), in part due to experimental complications related to the AV’s small size and heterogeneous fibrous structure. In this study, we developed specialized biaxial testing techniques for the AV cusp, including a method to determine the local structure–strain relationship to assess the effects of boundary tethering forces. Natural and glutaraldehyde (GL) treated cusps were subjected to an extensive biaxial testing protocol in which the ratios of the axial tensions were held at constant values. Results indicated that the local fiber architecture clearly dominated cuspal deformation, and that the tethering effects at the specimen boundaries were negligible. Due to unique aspects of cuspal fiber architecture, the most uniform region of deformation was found at the lower portion as opposed to the center of the cuspal specimen. In general, the circumferential strains were much smaller than the radial strains, indicating a profound degree of mechanical anisotropy, and that natural cusps were significantly more extensible than the GL treated cusps. Strong mechanical coupling between biaxial stretch axes produced negative circumferential strains under equibiaxial tension. Further, the large radial strains observed could not be explained by uncrimping of the collagen fibers, but may be due to large rotations of the highly aligned, circumferential-oriented collagen fibers in the fibrosa. In conclusion, this study provides new insights into the AV cusp’s structure–function relationship in addition to requisite data for constitutive modeling. [S0148-0731(00)00901-8]

Từ khóa


Tài liệu tham khảo

Thubrikar, M., 1990, The Aortic Valve, CRC, Boca Raton.

Lee, J. M., Boughner, D. R., and Courtman, D. W., 1984, “The Glutaraldehyde-Stabilized Porcine Aortic Valve Xenograft. II. Effect of Fixation With or Without Pressure on the Tensile Viscoelastic Properties of the Leaflet Material,” J. Biomed. Mater. Res., 18, pp. 79–98.

Lee, J. M., Courtman, D. W., and Boughner, D. R., 1984, “The Glutaraldehyde-Stabilized Porcine Aortic Valve Xenograft. I. Tensile Viscoelastic Properties of the Fresh Leaflet Material,” J. Biomed. Mater. Res., 18, pp. 61–77.

Vesely, I., and Noseworthy, R., 1992, “Micromechanics of the Fibrosa and the Ventricularis in Aortic Valve Leaflets,” J. Biomech., 25, pp. 101–113.

Mayne, A. S. D., Christie, G. W., Smaill, B. H., Hunter, P. J., and Barratt-Boyes, B. G., 1989, “An Assessment of the Mechanical Properties of Leaflets From Four Second-Generation Porcine Bioprosthesis With Biaxial Testing Techniques,” J. Thorac. Cardiovasc. Surg., 98, pp. 170–180.

Christie, G. W., and Barratt-Boyes, B. G., 1995, “Age-Dependent Changes in the Radial Stretch of Human Aortic Valve Leaflets Determined by Biaxial Stretching,” Ann. Thorac. Surg., 60, pp. 156–159.

Brossollet, L. J., and Vito, R. P., 1996, “A New Approach to Mechanical Testing and Modeling of Biological Tissues, With Application to Blood Vessels,” ASME J. Biomech. Eng., 118, pp. 433–439.

Lanir, Y., and Fung, Y. C., 1974, “Two-Dimensional Mechanical Properties of Rabbit Skin—I. Experimental System,” J. Biomech., 7, pp. 29–34.

Lanir, Y., and Fung, Y. C., 1974, “Two-Dimensional Mechanical Properties of Rabbit Skin—II. Experimental Results,” J. Biomech., 7, pp. 171–182.

Hunter, P. J., Smaill, B. H., and Nielson, P. M. F., 1986, “Biaxial Mechanical Testing of Biological Tissue,” Biophys. J., 49, p. 9090.

Sacks, M. S., and Chuong, C. J., 1993, “Biaxial Mechanical Properties of Passive Right Ventricular Free Wall Myocardium,” ASME J. Biomech. Eng., 115, pp. 202–205.

Sacks, M. S., and Chuong, C. J., 1998, “Orthotropic Mechanical Properties of Chemically Treated Bovine Pericardium,” Ann. Biomed. Eng., 26, pp. 892–902.

Sacks, M. S., Smith, D. B., and Hiester, E. D., 1998, “The Aortic Valve Microstructure: Effects of Trans-Valvular Pressure,” J. Biomed. Mater. Res., 41, pp. 131–141.

Sacks, M. S. , 1999, “A Method for Planar Biaxial Testing That Includes In-Plane Shear,” ASME J. Biomech. Eng., 121, pp. 551–555.

Nielsen, P. M. F., Hunter, P. J., and Smaill, B. H., 1991, “Biaxial Testing of Membrane Biomaterials: Testing Equipment and Procedures,” ASME J. Biomech. Eng., 113, pp. 295–300.

Spencer, A. J. M., 1980, Continuum Mechanics, Longman Scientific & Technical, New York.

Bathe, K. J., 1982, Finite Elements Procedures in Engineering Analysis, Prentice-Hall, Englewood Cliffs, NJ.

Press, W. H., Flannery, B. P., Teukolsky, S. A., and Vetterling, W. T., 1988, Numerical Recipes in C, Cambridge University Press, Cambridge.

Chew, P. H., Yin, F. C. P., and Zeger, S. L., 1986, “Biaxial Stress–Strain Properties of Canine Pericardium,” J. Mol. Cell. Cardiol., 18, pp. 567–578.

Humphrey, J. D., and Yin, F. C. P., 1988, “Biaxial Mechanical Behavior of Excised Epicardium,” ASME J. Biomech. Eng., 110, pp. 349–351.

Choi, H. S., and Vito, R. P., 1990, “Two Dimensional Stress–Strain Relationship for Canine Pericardium,” ASME J. Biomech. Eng., 112, pp. 153–159.

Oomens, C. W. J., Ratingen, M. R. V., Janssen, J. D., Kok, J. J., and Hendriks, M. A. N., 1993, “A Numerical–Experimental Method for a Mechanical Characterization of Biological Materials,” J. Biomech., 26, pp. 617–621.

Billiar, K. L., and Sacks, M. S., 1997, “A Method to Quantify the Fiber Kinematics of Planar Tissues Under Biaxial Stretch,” J. Biomech., 30, pp. 753–756.

Viidik, A., 1980, “Interdependence between Structure and Function in Collagenous Tissues,” Biology of Collagen, Viidik, A., and Vuust, J., eds., Academic Press, London.

Christie, G., and Stephenson, R., 1989, “Modelling the Mechanical Role of the Fibrosa and Ventricularis in the Porcine Bioprosthesis,” International Symposium on Surgery for Heart Valve Disease, ICR Publishers, London, pp. 815–824.

Hilbert, S., Sword, L., Batchelder, K., Barrick, M., and Ferrans, V., 1996, “Simultaneous Assessment of Bioprosthetic Heart Valve Biomechanical Properties and Collagen Crimp Length,” J. Biomed. Mater. Res., 31, pp. 503–509.

Sacks, M. S., Smith, D. B., and Hiester, E. D., 1997, “A SALS Device for Planar Connective Tissue Microstructural Analysis,” Ann. Biomed. Eng., 25, pp. 678–689.

Rousseau, E. P. M., Sauren, A. A. H. J., Van Hout, M. C., and Van Steenhoven, A. A., 1983, “Elastic and Viscoelastic Material Behaviour of Fresh and Glutaraldehyde-Treated Porcine Aortic Valve Tissue,” J. Biomech., 16, pp. 339–348.

Sauren, A., van Hout, M., van Steenhoven, A., Veldpaus, F., and Janssen, J., 1983, “The Mechanical Properties of Porcine Aortic Valve Tissues,” J. Biomech., 16, pp. 327–337.

Vesely, I. , 1991, “Analysis of the Medtronic Intact Bioprosthetic Valve,” J. Thorac. Cardiovasc. Surg., 101, pp. 90–99.

Vesely, I., Lozon, A., and Talman, E., 1993, “Is Zero-Pressure Fixation of Bioprosthetic Valves Truly Stress Free?” J. Thorac. Cardiovasc. Surg., 106, pp. 288–298.

Christie, G. W. , 1992, “Anatomy of Aortic Heart Valve Leaflets: The Influence of Glutaraldehyde Fixation on Function,” Eur. J. Cardio-thoracic Surg., 6, pp. S25–S33S25–S33.

Schoen, F. , 1997, “Aortic Valve Structure-Function Correlations: Role of Elastic Fibers No Longer a Stretch of the Imagination,” J. Heart Valve Disease, 6, pp. 1–6.

Lee, M. C., LeWinter, X. X., Freeman, G., Shabetai, R., and Fung, Y. C., 1985, “Biaxial Mechanical Properties of the Pericardium in Normal and Volume Overload Dogs,” Am. J. Physiol, 249, pp. H222–H230H222–H230.

Thubrikar, M., Aouad, J., and Nolan, S. P., 1986, “Comparison of the In-Vivo and In-Vitro Mechanical Properties of Aortic Valve Leaflets,” J. Thorac. Cardiovasc. Surg., 92, pp. 29–36.

May-Newman, K., and Yin, F. C. P., 1995, “Biaxial Mechanical Behavior of Excised Porcine Mitral Valve Leaflets,” Am. J. Physiol, 269, pp. H1319–H1327H1319–H1327.

Lo, D., and Vesely, I., 1995, “Biaxial Strain Analysis of the Porcine Aortic Valve,” Ann. Thorac. Surg., 60, pp. 374–378.

Billiar, K., 1998, “A Structurally Guided Constitutive Model for Aortic Valve Bioprostheses: Effects of Glutaraldehyde Treatment and Mechanical Fatigue,” Bioengineering, Philadelphia, University of Pennsylvania.

Clark, R. E., and Finke, E. C., 1974, “Scanning and Light Microscopy of Human Aortic Leaflets in Stressed and Relaxed States,” J. Thorac. Cardiovasc. Surg., 67, pp. 792–804.

Christie, G. W. , 1992, “Computer Modelling of Bioprosthetic Heart Valves,” Eur. J. Cardio-thoracic Surg., 6, pp. S95–S101S95–S101.

Cataloglu, A., Clark, R. E., and Gould, P. L., 1977, “Stress Analysis of Aortic Valve Leaflets With Smoothed Geometrical Data,” J. Biomech., 10, pp. 153–158.

Hamid, M., Sabbah, H., and Stein, P., 1985, “Finite Element Evaluation of Stresses on Closed Leaflets of Bioprosthetic Heart Valves With Flexible Stents,” Finite Elem. Anal. Design, 1, pp. 213–225.

Rousseau, E., van Steenhoven, A., and Janssen, J., 1988, “A Mechanical Analysis of the Closed Hancock Heart Valve Prosthesis,” J. Biomech., 21, pp. 545–562.

Vesely, I., and Boughner, D., 1989, “Analysis of the Bending Behaviour of Porcine Xenograft Leaflets and of Natural Aortic Valve Material: Bending Stiffness, Neutral Axis and Shear Measurements,” J. Biomech., 22, pp. 655–671.

Thubrikar, M., Skinner, J., Eppink, R., and Nolan, S., 1982, “Stress Analysis of Porcine Bioprosthetic Heart Valves In Vivo,” J. Biomed. Mater. Res., 16, p. 811811.

Talman, E. A., and Boughner, D. R., 1995, “Glutaraldehyde Fixation Alters the Internal Shear Properties of Porcine Aortic Heart Valve Tissue,” Ann. Thorac. Surg., 60, pp. S369–S373S369–S373.

Gloeckner, D., Billiar, K., and Sacks, M., 1999, “Effects of Mechanical Fatigue on the Bending Properties of the Porcine Bioprosthetic Heart Valve,” ASAIO J., 45, pp. 59–63.