On the implementation of a recently proposed dosimetric formalism to a robotic radiosurgery system

Medical Physics - Tập 37 Số 5 - Trang 2369-2379 - 2010
E. Pantelis1, A. Moutsatsos2, K. Zourari2, W Kilby3, Christos Antypas4, P. Papagiannis2, P. Karaiskos2, Evangelos Georgiou2, L. Sakelliou5
1Medical Physics Laboratory, Medical School, University of Athens, 75 Mikras Asias, 115 27 Athens, Greece and CyberKnife Center, Iatropolis-Magnitiki Tomografia, 54-56 Ethnikis-Antistaseos, 152 31, Athens, Greece
2Medical Physics Laboratory, Medical School, University of Athens, 75 Mikras Asias, 115 27 Athens (Greece)
3Accuray Inc., 1310 Chesapeake Terrace, Sunnyvale, California 94089
4First Department of Radiology, Aretaieion Hospital, Medical School, University of Athens, Athens, Greece and CyberKnife Center, Iatropolis‐Magnitiki Tomografia, 54‐56 Ethnikis‐Antistaseos, 152 31, Athens, Greece
5 Department of Physics, Nuclear and Particle Physics Section, University of Athens, Panepistimioupolis, Ilisia, 157 71 Athens (Greece)

Tóm tắt

Purpose:The aim of this work is to implement a recently proposed dosimetric formalism for nonstandard fields to the calibration and small field output factor measurement of a robotic stereotactic radiosurgery system.Methods:Reference dosimetry measurements were performed in the nonstandard, 60 mm diameter machine specific reference (msr) field using a Farmer ion chamber, five other cylindrical chambers with cavity lengths ranging from 16.25 down to 2.7 mm, and alanine dosimeters. Output factor measurements were performed for the 5, 7.5, 10, and 15 mm field sizes using microchambers, diode detectors, alanine dosimeters, TLD microcubes, and EBT Gafchromic films. Measurement correction factors as described in the proposed formalism were calculated for the ion chamber and diode detector output factor measurements based on published Monte Carlo data. Corresponding volume averaging correction factors were calculated for the alanine output factor measurements using 3D dose distributions, measured with polymer gel dosimeters.Results:Farmer chamber and alanine reference dosimetry results were found in close agreement, yielding a correction factor of for the chamber readings. These results were also found to be in agreement within experimental uncertainties with corresponding results obtained using the shorter cavity length ionization chambers. The mean measured dose values of the latter, however, were found to be consistently greater than that of the Farmer chamber. This finding, combined with an observed inverse relationship between the mean measured dose and chamber cavity length that follows the trend predicted by theoretical volume averaging calculations in the msr field, implies that the Farmer correction is greater than unity. Regarding the output factor results, deviations as large as 33% were observed between the different dosimeters used. These deviations were substantially decreased when appropriate correction factors were applied to the measured microchamber, diode, and alanine values. After correction, all diode and microchamber measured output factors agreed within 1.6% with the corresponding alanine measurements, and within 3.1% with the TLD measurements. The weighted mean output factors were , , , and for the 5, 7.5, 10, and 15 mm beams, respectively.Conclusions:The comparison of Farmer chamber measurements versus alanine reference dosimetry validates the use of the former for dosimetry in the msr field of this treatment delivery system. The corresponding results of this work obtained using chambers with different cavity lengths, combined with previous literature findings, suggest that a Farmer chamber dose response correction factor of 1.01 may improve calibration measurement accuracy when using the proposed dosimetric formalism. The correction factor is within 0.5% from unity for ion chambers with cavity lengths less than 10 mm. Substantial improvements in small field output factor measurement accuracy can be obtained when using microchambers and diodes by applying appropriately calculated correction factors to the detector measurements according to the proposed dosimetric formalism, and their routine use is therefore recommended.

Từ khóa


Tài liệu tham khảo

10.1118/1.2815356

10.1118/1.3232217

10.1118/1.2828195

10.1120/jacmp.v10i1.2939

10.1118/1.1544678

10.1118/1.2919099

10.1118/1.2734383

Francescon P., 2005, 271

10.1118/1.2219774

10.1118/1.2978228

10.1118/1.1855015

10.1118/1.2828187

10.1118/1.3005481

P.Andreo D. T.Burns K.Hohlfeld M. S.Huq T.Kanai F.Laitano V. G.Smyth andS.Vynckier “Absorbed dose determination in external beam radiotherapy ” International Atomic Energy Agency (IAEA) Technical Report Series No. 398 (Vienna 2000).

10.1118/1.3284208

10.1118/1.2975223

10.1088/0031-9155/45/3/308

10.1088/0031-9155/53/17/016

10.1118/1.3197062

BJR, 1996, Central axis depth dose data for use in radiotherapy, Br. J. Radiol., Suppl., 25, 62

Sharpe P., 2006

10.1088/0031-9155/53/10/020

10.1088/0031-9155/49/2/006

International Organization for Standardization “Guide to the expression of uncertainty in measurement (GUM) ”1995.

10.1120/jacmp.v9i1.2676

10.1118/1.3183500

10.1118/1.1929253

10.1118/1.1536161

10.1118/1.1916048

10.1118/1.1788911

10.1118/1.598915

10.1118/1.2905028

10.1088/0031-9155/54/15/006

10.1118/1.2940190

10.1088/0031-9155/54/9/024