Low Frequency of Early Complications With Dual-mobility Acetabular Cups in Cementless Primary THA
Tóm tắt
Dislocation complicates 1% to 5% of primary total hip arthroplasties (THAs). As a result, some surgeons consider dual-mobility articulations, which are usually used in the revision setting to decrease the likelihood of dislocation, as an option for primary THA. However, few studies have evaluated their use in this setting. (1) What is the cup survivorship when the dual-mobility articulation is used in the setting of primary THA? (2) What are the clinical outcomes with this approach? (3) What are the radiographic outcomes? (4) What are the complications of dual-mobility articulations in primary THA? Between 2011 and 2013, the five participating surgeons performed 495 cementless primary THAs. During that time, one of the five surgeons used dual-mobility articulations for all THAs, and the other four used it whenever the acetabular cup size was 52 mm or greater to enable a 28-mm head. Of the 495 patients, 453 (92%) were performed using this device. Smaller patients were treated with a standard THA. Of the 453 patients, a total of 43 patients (10%) were lost to followup before the 2-year minimum. The resulting 410 patients who were included in the analysis (164 men, 246 women) had a mean age of 64 years (SD, 12 years). The mean followup was 3 years (SD, 0.7 years). We performed Kaplan-Meier analyses to assess survivorship to aseptic failure and all-cause acetabular component survivorship. Clinical outcomes were evaluated using the Harris hip score (HHS); radiographs were assessed for cup migration, progressive radiolucencies, and positional changes of the components; and any surgery-related complications were recorded. The survivorship to aseptic failure and all-cause acetabular component survivorship was 99.8% (failures, n = 1) (95% confidence interval [CI], 4.517-4.547) and 99.3% (failures, aseptic, n = 1; septic, n = 2) (95% CI, 4.494–4.543); one hip had trunnion notching caused by impingement of a malpositioned cup, which was treated with revision of the cup and stem; and two patients had periprosthetic infections that were treated with two-stage revisions. There were no dislocations. Patients had a mean HHS of 94 (SD, 6) at final followup. On radiographic evaluation, no progressive radiolucencies or positional changes of the components were identified. Surgical complications included one traumatic avulsion of the abductors, one traumatic avulsion of the greater trochanter, which was repaired without revision of any of the components, and one loose femoral stem, which required revision of the femoral component only. Dual-mobility cups in primary THA yield seemingly comparable survivorship and complications to conventional THA bearings at short-term followup. Because serious complications have occasionally been reported with the use of these bearings, larger, longer term, comparative–and ideally, randomized–trials will be needed to establish the superiority of one approach over the other. Until or unless such studies show the superiority of dual-mobility designs for primary THA, we recommend that in the setting of uncomplicated primary THA, dual-mobility articulations be used only in centers that track their results carefully or in research protocols. Level IV, therapeutic study.
Tài liệu tham khảo
Banka TR, Ast MP, Parks ML. Early intraprosthetic dislocation in a revision dual-mobility hip prosthesis. Orthopedics. 2014;37:e395–397.
Banzhof JA, Robbins CE, Ven A, Talmo CT, Bono JV. Femoral head dislodgement complicating use of a dual mobility prosthesis for recurrent instability. J Arthroplasty. 2013;28:543.e541–543.
Bauchu P, Bonnard O, Cyprès A, Fiquet A, Girardin P, Noyer D. The dual-mobility POLARCUP: first results from a multicenter study. Orthopedics. 2008;31(Suppl 2).
Berry DJ, von Knoch M, Schleck CD, Harmsen WS. Effect of femoral head diameter and operative approach on risk of dislocation after primary total hip arthroplasty. J Bone Joint Surg Am. 2005;87:2456–2463.
Bouchet R, Mercier N, Saragaglia D. Posterior approach and dislocation rate: a 213 total hip replacements case-control study comparing the dual mobility cup with a conventional 28-mm metal head/polyethylene prosthesis. Orthop Traumatol Surg Res. 2011;97:2–7.
Boyer B, Philippot R, Geringer J, Farizon F. Primary total hip arthroplasty with dual mobility socket to prevent dislocation: a 22-year follow-up of 240 hips. Int Orthop. 2012;36:511–518.
Bystrom S, Espehaug B, Furnes O, Havelin LI, Norwegian Arthroplasty R. Femoral head size is a risk factor for total hip luxation: a study of 42,987 primary hip arthroplasties from the Norwegian Arthroplasty Register. Acta Orthop Scand. 2003;74:514–524.
Caton JH, Prudhon JL, Ferreira A, Aslanian T, Verdier R. A comparative and retrospective study of three hundred and twenty primary Charnley type hip replacements with a minimum follow up of ten years to assess whether a dual mobility cup has a decreased dislocation risk. Int Orthop. 2014;38:1125–1129.
Cooper HJ, Della Valle CJ. Large diameter femoral heads: is bigger always better? Bone Joint J. 2014;96:23–26.
Epinette JA. Clinical outcomes, survivorship and adverse events with mobile-bearings versus fixed-bearings in hip arthroplasty-a prospective comparative cohort study of 143 ADM versus 130 trident cups at 2 to 6-year follow-up. J Arthroplasty. 2015;30:241–248.
Epinette JA, Beracassat R, Tracol P, Pagazani G, Vandenbussche E. Are modern dual mobility cups a valuable option in reducing instability after primary hip arthroplasty, even in younger patients? J Arthroplasty. 2014;29:1323–1328.
Etienne G, Ragland PS, Mont MA. Use of constrained acetabular liners in total hip arthroplasty. Orthopedics. 2005;28:463–469; quiz 470–471.
Farizon F, de Lavison R, Azoulai JJ, Bousquet G. Results with a cementless alumina-coated cup with dual mobility. A twelve-year follow-up study. Int Orthop. 1998;22:219–224.
Fehring KA, Berry DJ. Dissociation and Intrapelvic entrapment of a dual-mobility polyethylene component. Clin Orthop Relat Res. 2016;474:1072–1076.
Guyen O, Pibarot V, Vaz G, Chevillotte C, Carret JP, Bejui-Hugues J. Unconstrained tripolar implants for primary total hip arthroplasty in patients at risk for dislocation. J Arthroplasty. 2007;22:849–858.
Hamadouche M, Arnould H, Bouxin B. Is a cementless dual mobility socket in primary THA: a reasonable option? Clin Orthop Relat Res. 2012;470:3048–3053.
Jauregui JJ, Pierce TP, Elmallah RK, Cherian JJ, Delanois RE, Mont MA. Dual mobility cups: an effective prosthesis in revision total hip arthroplasties for preventing dislocations. Hip Int. 2015:0.
Langlois J, El Hage S, Hamadouche M. Intraprosthetic dislocation: a potentially serious complication of dual mobility acetabular cups. Skeletal Radiol. 2014;43:1013–1016.
Mohammed R, Cnudde P. Severe metallosis owing to intraprosthetic dislocation in a failed dual-mobility cup primary total hip arthroplasty. J Arthroplasty. 2012;27:493.e491–493.
Mohammed R, Hayward K, Mulay S, Bindi F, Wallace M. Outcomes of dual-mobility acetabular cup for instability in primary and revision total hip arthroplasty. J Orthop Traumatol. 2015;16:9–13.
Odland AN, Sierra RJ. Intraprosthetic dislocation of a contemporary dual-mobility design used during conversion THA. Orthopedics. 2014;37:e1124–1128.
Philippot R, Adam P, Reckhaus M, Delangle F, Verdot F, Curvale G, Farizon F. Prevention of dislocation in total hip revision surgery using a dual mobility design. Orthop Traumatol Surg Res. 2009;95:407–413.
Philippot R, Boyer B, Farizon F. Intraprosthetic dislocation: a specific complication of the dual-mobility system. Clin Orthop Relat Res. 2013;471:965–970.
Philippot R, Camilleri JP, Boyer B, Adam P, Farizon F. The use of a dual-articulation acetabular cup system to prevent dislocation after primary total hip arthroplasty: analysis of 384 cases at a mean follow-up of 15 years. Int Orthop. 2009;33:927–932.
Phillips CB, Barrett JA, Losina E, Mahomed NN, Lingard EA, Guadagnoli E, Baron JA, Harris WH, Poss R, Katz JN. Incidence rates of dislocation, pulmonary embolism, and deep infection during the first six months after elective total hip replacement. J Bone Joint Surg Am. 2003;85:20–26.
Prudhon JL, Ferreira A, Verdier R. Dual mobility cup: dislocation rate and survivorship at ten years of follow-up. Int Orthop. 2013;37:2345–2350.
Sanchez-Sotelo J, Berry DJ. Epidemiology of instability after total hip replacement. Orthop Clin North Am. 2001;32:543–552, vii.
Singh JA. Epidemiology of knee and hip arthroplasty: a systematic review. Open Orthop J. 2011;5:80–85.
Tian JL, Sun L, Hu RY, Tian XB. Cementless total hip arthroplasty with structural allograft for massive acetabular defect in hip revision. Chin J Traumatol. 2014;17:331–334.
Williams JT Jr, Ragland PS, Clarke S. Constrained components for the unstable hip following total hip arthroplasty: a literature review. Int Orthop. 2007;31:273–277.