Flow-compensated diffusion encoding in MRI for improved liver metastasis detection

PLoS ONE - Tập 17 Số 5 - Trang e0268843
Frederik B. Laun1, Tobit Führes1, Hannes Seuß2,1, Astrid Müller1, Sebastian Bickelhaupt1, Alto Stemmer3, Thomas Benkert3, Michael Uder1, Marc Saake1
1Institute of Radiology, University Hospital Erlangen, Friedrich-Alexander-Universität Erlangen-Nürnberg, Erlangen, Germany
2Department of Radiology, Klinikum Forchheim—Fränkische Schweiz gGmbH, Forchheim, Germany
3Siemens Healthcare GmbH, Erlangen, Germany

Tóm tắt

Magnetic resonance (MR) diffusion-weighted imaging (DWI) is often used to detect focal liver lesions (FLLs), though DWI image quality can be limited in the left liver lobe owing to the pulsatile motion of the nearby heart. Flow-compensated (FloCo) diffusion encoding has been shown to reduce this pulsation artifact. The purpose of this prospective study was to intra-individually compare DWI of the liver acquired with conventional monopolar and FloCo diffusion encoding for assessing metastatic FLLs in non-cirrhotic patients. Forty patients with known or suspected multiple metastatic FLLs were included and measured at 1.5 T field strength with a conventional (monopolar) and a FloCo diffusion encoding EPI sequence (single refocused; b-values, 50 and 800 s/mm2). Two board-certified radiologists analyzed the DWI images independently. They issued Likert-scale ratings (1 = worst, 5 = best) for pulsation artifact severity and counted the difference of lesions visible at b = 800 s/mm² separately for small and large FLLs (i.e., < 1 cm or > 1 cm) and separately for left and right liver lobe. Differences between the two diffusion encodings were assessed with the Wilcoxon signed-rank test. Both readers found a reduction in pulsation artifact in the liver with FloCo encoding (p < 0.001 for both liver lobes). More small lesions were detected with FloCo diffusion encoding in both liver lobes (left lobe: six and seven additional lesions by readers 1 and 2, respectively; right lobe: five and seven additional lesions for readers 1 and 2, respectively). Both readers found one additional large lesion in the left liver lobe. Thus, flow-compensated diffusion encoding appears more effective than monopolar diffusion encoding for the detection of liver metastases.

Từ khóa


Tài liệu tham khảo

H Donato, 2017, Liver MRI: From basic protocol to advanced techniques, Eur J Radiol, 93, 30, 10.1016/j.ejrad.2017.05.028

M Cives, 2018, Gastroenteropancreatic Neuroendocrine Tumors, Ca-Cancer J Clin, 68, 471, 10.3322/caac.21493

EO Stejskal, 1965, Spin Diffusion Measurements: Spin Echoes in the Presence of a Time‐Dependent Field Gradient, The Journal of Chemical Physics, 42, 288, 10.1063/1.1695690

DS Grebenkov, 2007, NMR survey of reflected Brownian motion, Rev Mod Phys, 79, 1077, 10.1103/RevModPhys.79.1077

FB Laun, 2015, Symmetry of the gradient profile as second experimental dimension in the short-time expansion of the apparent diffusion coefficient as measured with NMR diffusometry, J Magn Reson, 259, 10, 10.1016/j.jmr.2015.07.003

E Aliotta, 2017, Convex optimized diffusion encoding (CODE) gradient waveforms for minimum echo time and bulk motion-compensated diffusion-weighted MRI, Magn Reson Med, 77, 717, 10.1002/mrm.26166

O Pena-Nogales, 2019, Optimized Diffusion-Weighting Gradient Waveform Design (ODGD) formulation for motion compensation and concomitant gradient nulling, Magn Reson Med, 81, 989, 10.1002/mrm.27462

E Aliotta, 2018, Eddy current-nulled convex optimized diffusion encoding (EN-CODE) for distortion-free diffusion tensor imaging with short echo times, Magn Reson Med, 79, 663, 10.1002/mrm.26709

MJ Middione, 2020, Optimization methods for magnetic resonance imaging gradient waveform design, NMR in biomedicine, 33, 10.1002/nbm.4308

TG Reese, 2003, Reduction of eddy-current-induced distortion in diffusion MRI using a twice-refocused spin echo, Magn Reson Med, 49, 177, 10.1002/mrm.10308

O Reynaud, 2016, Surface-to-Volume Ratio Mapping of Tumor Microstructure Using Oscillating Gradient Diffusion Weighted Imaging, Magn Reson Med, 76, 237, 10.1002/mrm.25865

J. Stepisnik, 1981, Analysis of NMR Self-Diffusion Measurements by a Density Matrix Calculation, Physica B, 104, 350, 10.1016/0378-4363(81)90182-0

N Shemesh, 2016, Conventions and Nomenclature for Double Diffusion Encoding NMR and MRI, Magn Reson Med, 75, 82, 10.1002/mrm.25901

FB Laun, 2011, Determination of the Defining Boundary in Nuclear Magnetic Resonance Diffusion Experiments, Phys Rev Lett, 107

JH Maki, 1991, The use of gradient flow compensation to separate diffusion and microcirculatory flow in MRI, Magn Reson Med, 17, 95, 10.1002/mrm.1910170114

N Fujita, 1992, Separation of Diffusion and Slow Flow Effects by Use of Flow Rephasing and Dephasing, Magnetic Resonance in Medicine, 24, 109, 10.1002/mrm.1910240112

A Ahlgren, 2016, Quantification of microcirculatory parameters by joint analysis of flow-compensated and non-flow-compensated intravoxel incoherent motion (IVIM) data, NMR in biomedicine, 29, 640, 10.1002/nbm.3505

A Wetscherek, 2015, Flow-compensated intravoxel incoherent motion diffusion imaging, Magn Reson Med, 74, 410, 10.1002/mrm.25410

S Brockstedt, 1995, Use of an Enhanced Gradient System for Diffusion Mr-Imaging with Motion-Artifact Reduction, Acta Radiol, 36, 662, 10.3109/02841859509176768

S McTavish, 2021, Gradient nonlinearity correction in liver DWI using motion‑compensated diffusion encoding waveforms, Magnetic Resonance Materials in Physics, Biology and Medicine

F Szczepankiewicz, 2021, Motion-compensated gradient waveforms for tensor-valued diffusion encoding by constrained numerical optimization, Magn Reson Med, 85, 2117, 10.1002/mrm.28551

TC Kwee, 2009, Influence of cardiac motion on diffusion-weighted magnetic resonance imaging of the liver, MAGMA, 22, 319, 10.1007/s10334-009-0183-1

A Riexinger, 2020, On the dependence of the cardiac motion artifact on the breathing cycle in liver diffusion-weighted imaging, PLoS One, 15, e0239743, 10.1371/journal.pone.0239743

M Ozaki, 2013, Motion artifact reduction of diffusion-weighted MRI of the liver: use of velocity-compensated diffusion gradients combined with tetrahedral gradients, J Magn Reson Imaging, 37, 172, 10.1002/jmri.23796

SS Rauh, 2020, A mixed waveform protocol for reduction of the cardiac motion artifact in black-blood diffusion-weighted imaging of the liver, Magnetic Resonance Imaging, 67, 59, 10.1016/j.mri.2019.12.011

YX Zhang, 2019, Motion-robust and blood-suppressed M1-optimized diffusion MR imaging of the liver, Magn Reson Med, 82, 302, 10.1002/mrm.27735

EA Eisenhauer, 2009, New response evaluation criteria in solid tumours: Revised RECIST guideline (version 1.1), European Journal of Cancer, 45, 228, 10.1016/j.ejca.2008.10.026

H Kandpal, 2009, Respiratory-triggered versus breath-hold diffusion-weighted MRI of liver lesions: comparison of image quality and apparent diffusion coefficient values, AJR Am J Roentgenol, 192, 915, 10.2214/AJR.08.1260

JS Choi, 2013, Comparison of breathhold, navigator-triggered, and free-breathing diffusion-weighted MRI for focal hepatic lesions, Journal of Magnetic Resonance Imaging, 38, 109, 10.1002/jmri.23949

A Furuta, 2014, Comparison of monopolar and bipolar diffusion weighted imaging sequences for detection of small hepatic metastases, Eur J Radiol, 83, 1626, 10.1016/j.ejrad.2014.06.003

M Saake, 2021, Image Quality and Detection of Small Focal Liver Lesions in Diffusion-Weighted Imaging: Comparison of Navigator Tracking and Free-Breathing Acquisition, Invest Radiol, 56, 579, 10.1097/RLI.0000000000000776

B Taouli, 2009, Diffusion-Weighted Imaging of the Liver: Comparison of Navigator Triggered and Breathhold Acquisitions, Journal of Magnetic Resonance Imaging, 30, 561, 10.1002/jmri.21876

A Tavakoli, 2019, Improved Liver Diffusion-Weighted Imaging at 3 T Using Respiratory Triggering in Combination With Simultaneous Multislice Acceleration, Invest Radiol, 54, 744, 10.1097/RLI.0000000000000594

C Dreher, 2020, Advanced Diffusion-Weighted Abdominal Imaging Qualitative and Quantitative Comparison of High and Ultra-High b-Values for Lesion Detection and Image Quality, Invest Radiol, 55, 285, 10.1097/RLI.0000000000000639

JR Landis, 1977, The measurement of observer agreement for categorical data, Biometrics, 33, 159, 10.2307/2529310

T Metens, 2016, Liver apparent diffusion coefficient repeatability with individually predetermined optimal cardiac timing and artifact elimination by signal filtering. Journal of magnetic resonance imaging, JMRI, 43, 1100, 10.1002/jmri.25089

Z Xiang, 2018, Evaluation of Regional Variability and Measurement Reproducibility of Intravoxel Incoherent Motion Diffusion Weighted Imaging Using a Cardiac Stationary Phase Based ECG Trigger Method, Biomed Res Int, 2018, 4604218, 10.1155/2018/4604218

S Ichikawa, 2019, Improving the Quality of Diffusion-weighted Imaging of the Left Hepatic Lobe Using Weighted Averaging of Signals from Multiple Excitations, Magnetic Resonance in Medical Sciences, 18, 225, 10.2463/mrms.mp.2018-0085

J Liau, 2012, Cardiac motion in diffusion-weighted MRI of the liver: artifact and a method of correction, J Magn Reson Imaging, 35, 318, 10.1002/jmri.22816

K Moulin, 2019, Effect of flow-encoding strength on intravoxel incoherent motion in the liver, Magn Reson Med, 81, 1521, 10.1002/mrm.27490

FB Laun, 2017, On the Vanishing of the t-term in the Short-Time Expansion of the Diffusion Coefficient for Oscillating Gradients in Diffusion NMR, Front Phys, 5, 56, 10.3389/fphy.2017.00056

M Bruegel, 2008, Diagnosis of hepatic metastasis: comparison of respiration-triggered diffusion-weighted echo-planar MRI and five t2-weighted turbo spin-echo sequences, AJR American journal of roentgenology, 191, 1421, 10.2214/AJR.07.3279

K Coenegrachts, 2007, Improved focal liver lesion detection: comparison of single-shot diffusion-weighted echoplanar and single-shot T2 weighted turbo spin echo techniques, Br J Radiol, 80, 524, 10.1259/bjr/33156643

DM Koh, 2008, Detection of colorectal hepatic metastases using MnDPDP MR imaging and diffusion-weighted imaging (DWI) alone and in combination, European radiology, 18, 903, 10.1007/s00330-007-0847-z

K Nasu, 2006, Hepatic metastases: diffusion-weighted sensitivity-encoding versus SPIO-enhanced MR imaging, Radiology, 239, 122, 10.1148/radiol.2383041384

T Parikh, 2008, Focal liver lesion detection and characterization with diffusion-weighted MR imaging: comparison with standard breath-hold T2-weighted imaging, Radiology, 246, 812, 10.1148/radiol.2463070432

B Taouli, 2010, Diffusion-weighted MR imaging of the liver, Radiology, 254, 47, 10.1148/radiol.09090021