Sensor fusion methods for reducing false alarms in heart rate monitoring

Journal of Clinical Monitoring and Computing - Tập 30 - Trang 859-867 - 2015
Gabriel Borges1, Valner Brusamarello1
1Electrical Engineering Department, Universidade Federal do Rio Grande do Sul, Porto Alegre, Brazil

Tóm tắt

Automatic patient monitoring is an essential resource in hospitals for good health care management. While alarms caused by abnormal physiological conditions are important for the delivery of fast treatment, they can be also a source of unnecessary noise because of false alarms caused by electromagnetic interference or motion artifacts. One significant source of false alarms is related to heart rate, which is triggered when the heart rhythm of the patient is too fast or too slow. In this work, the fusion of different physiological sensors is explored in order to create a robust heart rate estimation. A set of algorithms using heart rate variability index, Bayesian inference, neural networks, fuzzy logic and majority voting is proposed to fuse the information from the electrocardiogram, arterial blood pressure and photoplethysmogram. Three kinds of information are extracted from each source, namely, heart rate variability, the heart rate difference between sensors and the spectral analysis of low and high noise of each sensor. This information is used as input to the algorithms. Twenty recordings selected from the MIMIC database were used to validate the system. The results showed that neural networks fusion had the best false alarm reduction of 92.5 %, while the Bayesian technique had a reduction of 84.3 %, fuzzy logic 80.6 %, majority voter 72.5 % and the heart rate variability index 67.5 %. Therefore, the proposed algorithms showed good performance and could be useful in bedside monitors.

Tài liệu tham khảo

Phunchongharn P, Hossain E, Camorlinga S. Electromagnetic interference-aware transmission scheduling and power control for dynamic wireless access in hospital environments. IEEE Trans Inf Technol Biomed. 2011;15(6):890–9. Ogirala A, Stachel JR, Mickle MH. Electromagnetic interference of cardiac rhythmic monitoring devices to radio frequency identification: analytical analysis and mitigation methodology. IEEE Trans Inf Technol Biomed. 2011;15(6):848–53. Sweeney KT, Ward TE, McLoone SF. Artifact removal in physiological signals—practices and possibilities. IEEE Trans Inf Technol Biomed. 2012;16(3):488–500. Sweeney KT, Ayaz H, Ward TE, Izzetoglu M, McLoone SF, Onaral B. A methodology for validating artifact removal techniques for physiological signals. IEEE Trans Inf Technol Biomed. 2012;16(5):918–26. Oster J, Pietquin O, Kraemer M, Felblinger J. Nonlinear Bayesian filtering for denoising of electrocardiograms acquired in a magnetic resonance environment. IEEE Trans Biomed Eng. 2010;57(7):1628–38. Cvach M. Monitor alarm fatigue: an integrative review. Biomed Instrum Technol. 2012;46(4):268–77. Aboukhalil A, Nielsen L, Saeed M, Mark R, Clifford GD. Reducing false alarm rates for critical arrhythmias using the arterial blood pressure waveform. J Biomed Inform. 2008;41(3):442–51. Buxton OM, Ellenbogen JM, Wang W, Carballeira A, O’Connor S, Cooper D, Solet JM. Sleep disruption due to hospital noises: a prospective evaluation. Ann Intern Med. 2012;157(3):170–9. Blum JM, Kruger GH, Sanders KL, Gutierrez J, Rosenberg AL. Specificity improvement for network distributed physiologic alarms based on a simple deterministic reactive intelligent agent in the critical care environment. J Clin Monit Comput. 2009;23(1):21–30. Lipton JA, Van Ettinger MJB, Barendse RJ, Van Dam TB, van der Putten NHJJ, Nelwan SP. Alarms on the intensive cardiac care unit. In: Computers in cardiology, 2009. p. 253–256. Teichmann D, Foussier J, Jia J, Leonhardt S, Walter M. Noncontact monitoring of cardiorespiratory activity by electromagnetic coupling. IEEE Trans Biomed Eng. 2013;60(8):2142–52. Poh M-Z, Swenson NC, Picard RW. Motion-tolerant magnetic earring sensor and wireless earpiece for wearable photoplethysmography. IEEE Trans Inf Technol Biomed. 2010;14(3):786–94. Vullings R, De Vries B, Bergmans JWM. An adaptive Kalman filter for ECG signal enhancement. IEEE Trans Biomed Eng. 2011;58(4):1094–103. Silva I, Lee J, Mark RG. Signal quality estimation with multichannel adaptive filtering in intensive care settings. IEEE Trans Biomed Eng. 2012;59(9):2476–85. Behar J, Oster J, Qiao Li, Clifford GD. ECG signal quality during arrhythmia and its application to false alarm reduction. IEEE Trans Biomed Eng. 2013;60(6):1660–6. Krishnan R, Natarajan B, Warren S. Two-stage approach for detection and reduction of motion artifacts in photoplethysmographic data. IEEE Trans Biomed Eng. 2010;57(8):1867–76. Bruser C, Stadlthanner K, De Waele S, Leonhardt S. Adaptive beat-to-beat heart rate estimation in ballistocardiograms. IEEE Trans Inf Technol Biomed. 2011;15(5):778–86. Lipton JA, van Ettinger MJB, Barendse RJ, van Dam TB, van der Putten NHJJ, Nelwan SP. Frequency of alarms on the intensive cardiac care unit. In: Computers in cardiology, 2009. Qiao L, Clifford GD. Suppress false arrhythmia alarms of ICU monitors using heart rate estimation based on combined arterial blood pressure and Ecg analysis. In: Bioinformatics and biomedical engineering, 2008. Liu S, Gao RX, John D, Staudenmayer JW, Freedson PS. Multisensor data fusion for physical activity assessment. IEEE Trans Biomed Eng. 2012;59(3):687–96. Mo L, Liu S, Gao RX, John D, Staudenmayer JW, Freedson PS. Wireless design of a multisensor system for physical activity monitoring. IEEE Trans Biomed Eng. 2012;59(11):3230–7. Curone D, Tognetti A, Secco EL, Anania G, Carbonaro N, De-Rossi D, Magenes G. Heart rate and accelerometer data fusion for activity assessment of rescuers during emergency interventions. IEEE Trans Inf Technol Biomed. 2010;14(3):702–10. Mohseni HR, Woolrich MW, Kringelbach ML, Luckhoo H, Smith PP, Aziz TZ. Fusion of magnetometer and gradiometer sensors of MEG in the presence of multiplicative error. IEEE Trans Biomed Eng. 2012;59(7):1951–61. Sadjadi H, Hashtrudi-Zaad K, Fichtinger G. Fusion of electromagnetic trackers to improve needle deflection estimation: simulation study. IEEE Trans Biomed Eng. 2013;60(10):2706–15. Lanata A, Scilingo EP, De-Rossi D. A multimodal transducer for cardiopulmonary activity monitoring in emergency. IEEE Trans Inf Technol Biomed. 2010;14(3):817–25. Sarkar S, Koehler J. A dynamic risk score to identify increased risk for heart failure decompensation. IEEE Trans Biomed Eng. 2013;60(1):147–50. Ren H, Rank D, Merdes M, Stallkamp J, Kazanzides P. Multisensor data fusion in an integrated tracking system for endoscopic surgery. IEEE Trans Inf Technol Biomed. 2012;16(1):106–11. Zhang Y-T, Zheng Y-L, Lin W-H, Zhang H-Y, Zhou X-L. Challenges and opportunities in cardiovascular health informatics. IEEE Trans Biomed Eng. 2013;60(3):633–42. Sequeira J, Tsourdos A, Lazarus SB. Robust covariance estimation for data fusion from multiple sensors. IEEE Trans Instrum Meas. 2011;60(12):3833–44. Borges G, Brusamarello V. Bayesian fusion of multiple sensors for reliable heart rate detection. In: Instrumentation and Measurement Technology Conference (I2MTC), 2014. Baumgartner B, Rodel K, Knoll A. A data mining approach to reduce the false alarm rate of patient monitors. In: Engineering in Medicine and Biology Society (EMBC), 2012 Annual International Conference of the IEEE, p. 5935–5938, Aug 28 2012–Sept 1 2012. Li Q, Clifford GD. Signal quality and data fusion for false alarm reduction in the intensive care unit. J Electrocardiol 2012;45(6):596–603. Leite CR, Sizilio, GR, Neto AD, Valentim RA, Guerreiro AM. A fuzzy model for processing and monitoring vital signs in ICU patients. In: BioMedical Engineering Online (Online) 2011. Yang P, Dumont GA, Ansermino JM. Sensor fusion using a hybrid median filter for artifact removal in intraoperative heart rate monitoring. J Clin Monit Comput. 2009;23(2):75–83. Qiao L, Mark R, Clifford GD. Robust heart rate estimation from multiple asynchronous noisy sources using signal quality indices and a Kalman filter. Physiol Meas. 2008;29(1):15. Sayadi O, Shamsollahi MB. Life-threatening arrhythmia verification in ICU patients using the joint cardiovascular dynamical model and a Bayesian filter. IEEE Trans Biomed Eng. 2011;58(10):2748–57. Moody GB, Mark RG. A database to support development and evaluation of intelligent intensive care monitoring. In: Computers in cardiology, 1996. Mingqi C, Boric-Lubecke O, Lubecke VM. 0.5um CMOS implementation of analog heart-rate extraction with a robust peak detector. IEEE Trans Instrum Meas. 2008;57(4):690–8. Ravanshad N, Rezaee-Dehsorkh H, Lotfi R, Lian Y. A level-crossing based QRS-detection algorithm for wearable ECG sensors. IEEE J Biomed Health Inform. 2014;18(1):183–92. Chen B, Varshney PK. A Bayesian sampling approach to decision fusion using hierarchical models. IEEE Trans Signal Process. 2002;50(8):1809–18. Guerriero M, Svensson L, Willett P. Bayesian data fusion for distributed target detection in sensor networks. IEEE Trans Signal Process. 2010;58(6):3417–21. Zheng Y, Niu R, Varshney PK. Sequential Bayesian estimation with censored data for multi-sensor systems. IEEE Trans Signal Process. 2014;62(10):2626–41. Sun B-Y, Zhang X-M, Li J, Mao X-M. Feature fusion using locally linear embedding for classification. IEEE Trans Neural Netw. 2010;21(1):163–8. Zhang Y-Q, Fraser MD, Gagliano RA, Kandel A. Granular neural networks for numerical-linguistic data fusion and knowledge discovery. IEEE Trans Neural Netw. 2000;11(3):658–67. Xia Y, Leung H, Bosse E. Neural data fusion algorithms based on a linearly constrained least square method. IEEE Trans Neural Netw. 2002;13(2):320–9.