Coronary calcium mass scores measured by identical 64-slice MDCT scanners are comparable: a cardiac phantom study

Springer Science and Business Media LLC - Tập 26 - Trang 89-98 - 2009
Hildebrand Dijkstra1, Marcel J. W. Greuter1, Jaap M. Groen1, Rozemarijn Vliegenthart-Proença1, KlaasJan W. K. Renema2, Frank de Lange2, Matthijs Oudkerk1
1Department of Radiology, University Medical Center Groningen, Groningen, The Netherlands
2Department of Radiology, University Medical Center St. Radboud, Nijmegen, The Netherlands

Tóm tắt

To assess whether absolute mass scores are comparable or differ between identical 64-slice MDCT scanners of the same manufacturer and to compare absolute mass scores to the physical mass and between scan modes using a calcified phantom. A non-moving anthropomorphic phantom with nine calcifications of three sizes and three densities was scanned 30 times on three 64-slice MDCT scanners of manufacturer A and on three 64-slice MDCT scanners of manufacturer B in both sequential and spiral scan mode. The mean mass scores and mass score variabilities of seven calcifications were determined for all scanners; two non-detectable calcifications were omitted. It was analyzed whether identical scanners yielded similar or significantly different mass scores. Furthermore mass scores were compared to the physical mass and mass scores were compared between scan modes. The mass score calibration factor was determined for all scanners. Mass scores obtained on identical scanners were similar for almost all calcifications. Overall, mass score differences between the scanners were small ranging from 1.5 to 3.4% for the total mass scores, and most differences between scanners were observed for high density calcifications. Mass scores were significantly different from the physical mass for almost all calcifications and all scanners. In sequential mode the total physical mass (167.8 mg) was significantly overestimated (+2.3%) for 4 out of 6 scanners. In spiral mode a significant overestimation (+2.5%) was found for system B and a significant underestimation (−1.8%) for two scanners of system A. Mass scores were dependent on the scan mode, for manufacturer A scores were higher in sequential mode and for manufacturer B in spiral mode. For system A using spiral scan mode no differences were found between identical scanners, whereas a few differences were found using sequential mode. For system B the scan mode did not affect the number of different mass scores between identical scanners. Mass scores obtained in the same scan mode are comparable between identical 64-slice CT scanners and identical 64-slice CT scanners on different sites can be used in follow-up studies. Furthermore, for all systems significant differences were found between mass scores and the physical calcium mass; however, the differences were relatively small and consistent.

Tài liệu tham khảo

Agatston AS, Janowitz WR, Detrano R et al (1990) Quantification of coronary artery calcium using ultrafast computed tomography. J Am Coll Cardiol 15:827–832 Vliegenthart R, Oudkerk M, Witteman JC (2005) Coronary calcification improves cardiovascular risk prediction in the elderly. Circulation 112(4):572–577 Detrano R, Guerci AD, Kronmal RA et al (2008) Coronary calcium as a predictor of coronary events in four racial or ethnic groups. N Engl J Med 358(13):1336–1345 Geluk CA, Dikkers R, Zijlstra F et al (2008) Measurement of coronary calcium scores by electron beam computed tomography or exercise testing as initial diagnostic tool in low-risk patients with suspected coronary artery disease. Eur Radiol 18(2):244–252 Arad Y, Spadaro LA, Goodman K et al (2000) Prediction of coronary events with electron beam computed tomography. J Am Coll Cardiol 36(4):1253–1260 Callister TQ, Cooil B, Raggi P et al (1998) Coronary artery disease: improved reproducibility of calcium scoring with an electron-beam CT volumetric method. Radiology 208:807–814 Detrano R, Kang X, Nickerson S et al (1994) Accuracy of quantifying coronary hydroxyapatite with electron beam tomography. Investig Radiol 29(8):733–738 Ulzheimer S, Kalender WA (2003) Assessment of calcium scoring performance in cardiac computed tomography. Eur Radiol 13(3):484–497 McCollough CH, Ulzheimer S, Kalender WA et al (2007) Coronary artery calcium: a multi-institutional, multimanufacturer international standard for quantification at cardiac CT. Radiology 243(2):527–538 Hoffmann U, Siebert U, O’Donnell CJ et al (2006) Evidence for lower variability of coronary artery calcium mineral mass measurements by multi-detector computed tomography in a community-based cohort—consequences for progression studies. Eur J Radiol 57(3):396-402 Horiguchi J, Shen Y, Ito K et al (2005) Electron beam CT versus 16-MDCT on the variability of repeated coronary artery calcium measurements in a variable heart rate phantom. AJR Am J Roentgenol 185(4):995–1000 Hong C, Bae KT, Zhu F et al (2003) Coronary artery calcium quantification at multi-detector row CT: influence of heart rate and measurement methods on interacquisition variability initial experience. Radiology 228(1):95–100 Hong C, Bae KT, Pilgram TK (2003) Coronary artery calcium: accuracy and reproducibility of measurements with multi-detector row CT—assessment of effects of different thresholds and quantification methods. Radiology 227(3):795–801 Detrano RC, Anderson M, Bild DE (2005) Coronary calcium measurements: effect of CT scanner type and calcium measure on rescan reproducibility—MESA study. Radiology 236(2):477–484 Chung H, McClelland RL, Budoff MJ (2008) Repeatability limits for measurement of coronary artery calcified plaque with cardiac CT in the multi-ethnic study of atherosclerosis. AJR Am J Roentgenol 190(2):W87–W92 Sevrukov AB, Bland JM, Kondos GT (2005) Serial electron beam CT measurements of coronary artery calcium: has your patient’s calcium score actually changed? AJR Am J Roentgenol 185(6):1546–1553 Rutten A, Isgum I, Prokop M (2008) Coronary calcification: effect of small variation of scan starting position on Agatston, volume, and mass scores. Radiology 246(1):90–98 Greuter MJW, Dijkstra H, Groen JM et al (2007) 64 Slice MDCT generally underestimates coronary calcium scores as compared to EBT: a phantom study. Med Phys 34(9):3510–3519 Hong C, Pilgram TK, Bae KT (2003) Improving mass measurement of coronary artery calcification using threshold correction and thin collimation in multi-detector row computed tomography: in vitro experiment. Acad Radiol 10(9):969–977 Raggi P, Callister TQ, Cooil B (2002) Calcium scoring of the coronary artery by electron beam CT: how to apply an individual attenuation threshold. AJR Am J Roentgenol 178(2):497–502 Groen JM, Dijkstra H, Greuter MJ, Oudkerk M (2009) Threshold adjusted calcium scoring using CT is less susceptible to cardiac motion and more accurate. Med Phys 36(2):438–446 Birnbaum BA, Hindman N, Lee J, Babb JS (2007) Multi-detector row CT attenuation measurements: assessment of intra- and interscanner variability with an anthropomorphic body CT phantom. Radiology 242(1):109–119 Rosol M, Sachdev K, Hoffmann U et al (2006) A novel model to test accuracy and reproducibility of MDCT scan protocols for coronary calcium in vivo. Int J Cardiovasc Imaging 22(1):111–118 Horiguchi J, Kiguchi M, Fujioka C, Arie R, Shen Y, Sunasaka K, Kitagawa T, Yamamoto H, Ito K (2008) Variability of repeated coronary artery calcium scoring and radiation dose on 64- and 16-slice computed tomography by prospective electrocardiographically-triggered axial and retrospective electrocardiographically-gated spiral computed tomography: a phantom study. Acad Radiol 15(8):958–965 Thomas CK, Muhlenbruch G, Mahnken AH et al (2006) Coronary artery calcium scoring with multislice computed tomography: in vitro assessment of a low tube voltage protocol. Investig Radiol 41(9):668–673 Groen JM, Greuter MJ, Schmidt B, Suess C, Vliegenthart R, Oudkerk M (2007) The influence of heart rate, slice thickness, and calcification density on calcium scores using 64-slice multidetector computed tomography: a systematic phantom study. Investig Radiol 42(12):848–855 Groen JM, Greuter MJ, Vliegenthart R, Suess C, Schmidt B, Zijlstra F, Oudkerk M (2008) Calcium scoring using 64-slice MDCT, dual source CT and EBT: a comparative phantom study. Int J Cardiovasc Imaging 24(5):547–556 Achenbach S, Ropers D, Holle J, Muschiol G, Daniel WG, Moshage W (2000) In-plane coronary arterial motion velocity: measurement with electron-beam CT. Radiology 216(2):457–463