Optimal QRS detector

Medical & Biological Engineering & Computing - Tập 21 - Trang 343-350 - 1983
N. V. Thakor1, J. G. Webster1, W. J. Tompkins1
1Department of Electrical and Computer Engineering, University of Wisconsin-Madison, Madison

Tóm tắt

The problem of detecting the QRS complex in the presence of noise was analysed. Most QRS detectors contain a filter to improve the signal-to-noise ratio and compare the signal with a threshold. In an earlier paper we identified an optimal filter. Various techniques to generate threshold and detector designs were studied. Automatic gain-control circuits with a fixed threshold have a very slow response to different rhythms. Automatic threshold circuits based on simple peak-detection schemes have a fast response, but are very sensitive to sudden variations in QRS amplitudes and noise transients. None of the methods described to date present any optimisation criteria for detecting the signal (QRS complex) in the presence of noise. The probabilities of FPs (false positives) and FNs (false negatives) were investigated and an optimised threshold criterion based on FP/FN was developed. Presently, data are being collected to compare various techniques from their ROC (receiver operating characteristics).

Tài liệu tham khảo

Brydon, J. (1976) Automatic monitoring of cardiac arrhythmias. InIEE Medical Electronics Monographs 18–22,Hill, D. W. andWatson, B. W. (Eds.), Peter Peregrinus, 27–41. Fraden, J. andNeuman, M. R. (1980) QRS wave detection.Med. & Biol. Eng. & Comput.,18, 125–132. Gebben, V. D. andWebb, Jr., J. A. (1973) Cardiac R-wave detector with automatic sensitivity control. NASA Technical Note TN D-7152. Goovaerts, H. G., Ross, H. H., Van Den Akker, T. J. andSchneider, G. (1976) A digital QRS detector based on the principle of contour limiting.IEEE Trans. Biomed. Eng.,23, 154–160. Graeme, J. G. (1973)Applications of operational amplifiers. McGraw-Hill, New York. Hancock, J. C. andWintz, P. A. (1966)Signal detection theory. McGraw Hill, New York. Hewlett Packard Co. (1975)ECG Cardiotach Module 78209 A. Instruction manual, Waltham, MA. Law, H. F., Epstein, R. F. andEpstein, M. A. (1979) High-resolution determination of the R-R interval.Am. J. Physiol.,236, 894–898. Thakor, N. V. (1981) Design, implementation, and evaluation of a microcomputer-based portable arrhythmia monitor. Ph.D. Thesis, Dept. Elect. Comput. Eng., Univ. Wisconsin, Madison, WI. Thakor, N. V. andWebster, J. G. (1981) Design and evaluation of QRS and noise detectors for ambulatory ECG monitors.Med. & Biol. Eng. & Comput.,20, 709–714. Thakor, N. V., Webster, J. G. andTompkins, W. J. (1980) Optimal QRS filter.IEEE Front. Eng. Health Care, pp. 190–195. Whalen, A. D. (1971)Detection of signals in noise. Academic Press, London.